Evaluation of two automatic systems for measurement of the erythrocyte sedimentation rate

N/A
N/A
Protected

Academic year: 2021

Condividi "Evaluation of two automatic systems for measurement of the erythrocyte sedimentation rate"

Copied!
1
0
0

Testo completo

(1)

Evaluation of two automatic systems for measurement of the erythrocyte sedimentation rate Zenobi R°, Viola FG°, Ballerini S°, Bellincampi L°, Federici G°

°Department of Laboratory Medicine, University-Hospital of “Tor Vergata” Rome, Italy Our aim was to compare two different automated devices for the measurement of the length of erythrocyte sedimentation rate (LSRB) in order to decrease the amount of blood sample collected. We compared VES-MATIC 60 (DIESSE s.p.a.,Grosseto, Italy) with VES-CUBE 200 (DIESSE s.p.a.,Grosseto, Italy), that requires K2-EDTA and sodium-citrate blood samples, respectively. A total of 160 whole blood samples from patients of the Laboratory Medicine of University Hospital “Tor Vergata” Rome were included in this study; LSRB was determined following ICSH

recommendations. Linear regression analysis was used for accuracy study; Reproducibility was assessed using an internal quality control for at least one month (mean = 6.93, SD = 0.775 CV=10.5).

Our data showed statistically significant correlation between the two different systems, when the data of VES-CUBE 200 were adjusted to 18°C by Manley’s normogramm. Moreover, both systems presented a good reproducibility and similar CV value.

In conclusion, both systems can be used for LSRB determination with the same range of reference; in addition, our previous study showed that LSRB values on plasma-citrate or on K2-EDTA were the same on VES-MATIC 60, therefore the automatic system for LSRB measurement, VES-CUBE 200, shows important advantages on one hand it decreases the amount of whole blood collected being the sample undiluted, on the other hand the test can be performed on the same tube collected for hemochrome analysis, using K2-EDTA as anticoagulant.

1)Caswell M, Stuart J. Assessment of Diesse Ves-matic automated system for measuring erythrocyte sedimentation rate. J Clin Pathol 1992 Feb;45(2):184

Riferimenti

Documenti correlati

We characterized the distribution of secondary structures in silk fibers for a large set of South American historical textiles, coupling FTIR with multivariate statistical analysis

Il sistema delle cure nell’ambito della medicina del territorio in una logica di rete ha specificamen- te l’obiettivo di garantire una presa in carico integrata dei bisogni

For this matter of fact, the concept of GOG (Grade of Generation) protocols has been introduced in order to describe different Levels of Geometry, in function of

system of adiponectin (ADN), an insulin-sensitizing and anti- inflammatory adipokine, in tissues known to be the target of chronic diseases associated to obesity

A differenza di CONSTRUCT però, invece di specificare un template di costruzione del nuovo grafo, DESCRIBE recupererà tutte le informazioni connesse alle risorse unificate

quando si vuole rappresentare il comportamento sintattico delle lexical entry e la corrispondenza tra argomenti sintattici e semantici. • Le descrizioni delle lexical entry

In this study, we describe the results of a retrospective analy- sis, carried out during 2016 and 2017, concerning the measles antibody prevalence in patients and

MATH@TOV is offering a unique opportunity to improve substantially the Master Programme in Pure and Applied Mathematics and make it internationally competitive. 1) MATH@TOV will