MAY/JUNE 2016 | GLAUCOMA TODAY 39

COVER FOCUS

Optical coherence tomography (OCT) is widely used in the diagnosis and detec- tion of glaucomatous progression. Many manufacturers make spectral-domain OCT (SD-OCT) devices, the most widely used type of OCT. In contrast to time-domain OCT, with SD-OCT, a greater area of the retina can be measured without inter- polation of data. Additionally, the anterior segment can be imaged using the same machine. Most OCT machines have built-in proprietary software that can measure the peripapillary area as well as the macular region. Most units also have the capability of segmentation analysis, providing data regarding substructures of the retina. These data are presented in the form of a report, however, in which the patient’s measurements are compared to a normative data- base (also proprietary to the manufacturer). The overall as well as sectoral data are categorized as normal, borderline, or abnormal along with color-coding of green, yellow, and red.

Clinicians need to confirm a few rules of accurate imaging before accepting the results. One is the signal strength. Poor signal strength leads to a faint image, with the software unable to accurately identify the boundaries of the layers that it is trying to measure. Second is motion artifact. The eye typically has five microsaccades per second, so unless there is image tracking, motion artifact is very likely even with the most cooperative patient. The third rule is accu- rate acquisition of the image without cutoff at the edges and with good centration of the optic nerve and the fovea.

PATHOLOGY

After confirming that the aforementioned rules have been met, the physician must rule out the presence of coexisting

PITFALLS IN OPTICAL

COHERENCE TOMOGRAPHY IMAGING

Essentials for correct interpretation.

BY SANJAY ASRANI, MD

Figure 1. Misidentification artifact. Vitreous-retina interface is misidentified (arrow) as the outer surface of the nerve fiber layer.

Figure 2. False progression artifact. Release of epiretinal membrane (noted in April visit; yellow arrow) creates a false appearance of progression (blue arrow) that the clinician may miss if he or she does not carefully examine the raw images.

(2)

40 GLAUCOMA TODAY | MAY/JUNE 2016

COVER FOCUS

pathology that may confound the results and thus the interpretation.1 For example, vitreous traction on the retinal nerve fiber layer (RNFL) may result in artifactually increased RNFL thickness (Figure 1).

Epiretinal membranes in the peripapillary and macu- lar region are a common cause of increased RNFL or macular thickness, respectively. Next, the physician must confirm that the software has correctly identi- fied the layers it is trying to measure (eg, RNFL, total retinal thickness, ganglion cell layer thickness, etc.).

The physician then needs to determine if the patient’s ocular condition merits comparison to a normative database. For example, high myopes are not included in normative databases, and thus their measurements typically read as abnormal.

It is imperative for the clinician to have ready access to the raw images acquired before looking at the interpretation provided in the report. The tech- nician acquiring the images needs to confirm that the patient blinks just before image acquisition or instills artificial tears to improve the image quality.

The technician must also confirm the quality and accuracy of the data acquired before the patient returns to the physician.

PROGRESSION

Assessing progression presents different chal- lenges. The software must be capable of aligning to the baseline or previous image acquired in terms of location as well as angulation, because a small but different tilt of the head can result in erroneous interpretation. The physician must also determine if the change detected by the machine is clinically significant (greater than 2 standard deviations of the measurement reproducibility2). When review- ing macular measurements, clinicians must consider whether a change is a result of glaucoma, release of the posterior vitreous attachments to the RNFL, or improvement of overall blood sugar control (Figure 2). Moreover, in the presence of uveitis, coex- isting RNFL edema and subclinical macular edema can result in erroneous interpretation of normal- appearing thickness.3,4 Upon subsequent resolution of such edema (with uveitis treatment), the thickness of these layers can dramatically decrease once again, giving an erroneous interpretation of glaucomatous progression.

Cystic spaces in the RNFL or subretinal space in the peripapillary region in myopia as well as cys- tic changes in the outer plexiform layer in rapidly advancing glaucoma can also confound interpreta- tion.4 Masqueraders of glaucoma such as neurosyphi- lis, optic disc drusen, ischemic optic neuropathy, Figure 3. Masquerader of glaucoma. Macular thickness asymmetry

with loss in the superotemporal quadrant of the left eye (arrow; A).

RNFL loss in the same quadrant of the left eye (arrow; B). The loss of both RNFL and macular thickness, however, were from a retinal scar caused by toxoplasmosis (C).

A

B

C

(3)

MAY/JUNE 2016 | GLAUCOMA TODAY 41

COVER FOCUS

hemianopia, and optic neuritis present another set of chal- lenges to the assessment of OCT results (Figure 3).5

Clinicians use anterior segment OCT to detect either the angle opening width, which depends on identifying the scleral spur, or Schwalbe line. Both may be unclear depending on the quality of the image.6 Additionally, ambi- ent lighting, the brightness of the instrument monitor, or a patient’s not having adapted to the dark can cause the angle to appear open, because its width is directly propor- tional to the pupil’s size.

CONCLUSION

SD-OCT has tremendous potential for glaucoma man- agement, but eye care providers must be aware of possible artifacts. Re-acquiring the scans or manually correcting seg- mentation errors, though time consuming, may be needed for better clinical care. Continued improvements in soft- ware and segmentation algorithms may provide increas- ingly reliable quantitative thickness data. n

1. Asrani S, Essaid S, Alder, B, Santiago MC. Artifacts in SD OCT in glaucoma. JAMA Ophthalmol. 2014;132(4):396-402.

2. Ghasia FF, El-Dairi M, Freedman SF, et al. Reproducibility of spectral-domain optical coherence tomography measurements in adult and pediatric glaucoma. J Glaucoma. 2015;24(1):55-63.

3. Asrani S, Moore D, Jaffe G. Paradoxical changes of retinal nerve fiber layer thickness in uveitic glaucoma. JAMA Ophthalmol. 2014;132(7):877-880.

4. Moore, D, Jaffe G, Asrani S. Retinal nerve fiber layer thickness measurements: uveitis; a major confounding factor.

Ophthalmology. 2015;122(3):511-517.

5. Rosdahl J, Asrani S. Glaucoma masqueraders: diagnosis using spectral domain optical coherence tomography. Saudi J Ophthalmol. 2012;26(4):433-440.

6. Wen JC, Freedman SF, El-Dairi MA, Asrani S. Microcystic macular changes in primary open-angle glaucoma.

J Glaucoma. 2016;25(3):258-262.

Sanjay Asrani, MD

n professor of ophthalmology, Duke University, Durham, North Carolina

n head, Glaucoma OCT Reading Center, Duke Eye Center, Durham, North Carolina

n director, Duke Eye Center of Cary, Cary, North Carolina

n (919) 684-8656; sanjay.asrani@duke.edu

n financial disclosure: receives lecture honoraria from Heidelberg Engineering

• Clinicians need to confirm a few rules of accurate opti- cal coherence tomography imaging before accepting the results: signal strength, motion artifact, and proper acquisition of the image.

• When assessing progression, physicians must deter- mine whether or not changes are clinically significant and, if so, whether they are a result of glaucoma or something else.

AT A GLANCE

Riferimenti

Documenti correlati

3(a), in the recorded cross section, two spatial points were randomly selected from the regions of static solid-gel ðz s ; x s Þ and dynamic fluid ðz d ; x d Þ, respectively, and

Invited papers from Massachusetts Institute of Technology, ‘‘The Development, Commercialization, and Impact of Optical Coherence Tomogra- phy,’’ and the Medical University of

de Carlo TE, Chin AT, Joseph T, et al., Distinguishing Diabetic Macular Edema from Capillary Nonperfusion Using Optical Coherence Tomography Angiography, Ophthalmic Surg Lasers

In vivo dynamic human retinal blood flow imaging using ultra-high-speed spectral domain optical coherence tomography.. Phase- resolved optical frequency

Retinal nerve fiber layer imaging with spectral-domain optical coherence tomogra- phy: analysis of the retinal nerve fiber layer map for glaucoma detection.. Medeiros FA, Zangwill

Optical coherence tomography had its origin in femtosecond optics, but used optical communications technologies and required advanced engineering for early OCT prototypes,

New Insights Into Microaneurysms in the Deep Capillary Plexus Detected by Optical Coherence Tomography Angiography in Diabetic Macular Edema.. Moore J, Bagley S, Ireland G,

Additional signal processing steps (such as interpolation) may also be required in FDOCT to prepare the spectral interferogram for the inverse Fourier transform, so that the data