• Non ci sono risultati.

Immediately after excitation, the magnetization rotates in the xy-plane and is now called transverse magnetization or M

N/A
N/A
Protected

Academic year: 2022

Condividi "Immediately after excitation, the magnetization rotates in the xy-plane and is now called transverse magnetization or M"

Copied!
4
0
0

Testo completo

(1)

2 Relaxation

2 Relaxation

What happens to the spins after they have been excited as just described?

Immediately after excitation, the magnetization rotates in the xy-plane and is now called transverse magnetization or M

xy

. It is the rotating transverse magnetization that gives rise to the MR signal in the receiver coil. However, the MR signal rapidly fades due to two independent processes that reduce transverse magnetization and thus cause a return to the stable state present before excitation: spin-lattice interaction and spin-spin interaction. These two processes cause T1 relaxation and T2 relaxation, respectively.

2.1 T1: Longitudinal Relaxation

Transverse magnetization decays and the magnetic moments gradually re- align with the z-axis of the main magnetic field B0, as discussed previously.

The transverse magnetization remaining within the xy-plane – strictly speak- ing the projection of the magnetization vector onto the xy-plane (▶ Fig. 6) – decreases slowly and the MR signal fades in proportion. As transverse magnetization decays, the longitudinal magnetization, M

z

– the projection of the magnetization vector onto the z-axis – is slowly restored. This process is known as longitudinal relaxation or T1 recovery.

The nuclei can return to the ground state only by dissipating their excess

energy to their surroundings (the “lattice”, which is why this kind of relax-

ation is also called spin-lattice relaxation). The time constant for this recov-

ery is T1 and is dependent on the strength of the external magnetic field,

B0, and the internal motion of the molecules (Brownian motion). Biological

tissues have T1 values of half a second to several seconds at 1.5 T.

(2)

Fig. 6. T1 relaxation. Decay of transverse magnetization and regrowth of magnetization along the z-axis require an exchange of energy

2.2 T2/T2*: Transverse Relaxation

To understand transverse relaxation, it is first necessary to know what is meant by “phase”. As used here, phase refers to the position of a magnetic moment on its circular precessional path and is expressed as an angle. Con- sider two spins, A and B, precessing at the same speed in the xy-plane. If B is ahead of A in its angular motion by 10°, then we can say that B has a phase of +10 relative to A. Conversely, a spin C that is behind A by 30° has a phase of -30° (▶ Fig. 7).

Immediately after excitation, part of the spins precess synchronously.

These spins have a phase of 0° and are said to be in phase. This state is called phase coherence.

Fig. 7. Phase. Vector B has a phase of +10° relative to A while C has a phase of –30°.

Note that all vectors rotate about the z-axis while their phases differ by the respective angles

(3)

2 Relaxation

Fig. 8. T2 and T2* relaxation. Spins get out of phase (lose phase coherence), resulting in the loss of transverse magnetization without energy dissipation

For reasons that we will go into soon, phase coherence is gradually lost as some spins advance while others fall behind on their precessional paths.

The individual magnetization vectors begin to cancel each other out instead of adding together. The resulting vector sum, the transverse magnetization, becomes smaller and smaller and finally disappears, and with it the MR sig- nal (▶ Fig. 8).

In other words, transverse relaxation is the decay of transverse magneti- zation because spins lose coherence (dephasing). Transverse relaxation dif- fers from longitudinal relaxation in that the spins do not dissipate energy to their surroundings but instead exchange energy with each other. Coherence is lost in two ways:

– Energy transfer between spins as a result of local changes in the magnetic field. Such fluctuations are due to the fact that the spins are associated with small magnet fields that randomly interact with each other. Spins precess faster or slower according to the magnetic field variations they experience. The overall result is a cumulative loss of phase. It is a process due to pure spin-spin interaction and as such is unaffected by application of a 180° refocusing pulse (▶ Chapter 7). Dephasing occurs with the time constant T2 and is more or less independent of the strength of the external magnetic field, B0.

– Time-independent inhomogeneities of the external magnetic field B0.

These are intrinsic inhomogeneities that are caused by the magnetic

field generator itself and by the very person being imaged. They contrib-

ute to dephasing, resulting in an overall signal decay that is even faster

than described by T2. This second type of decay occurs with the time

(4)



constant T2*, which is typically shorter than T2. Most of the inhomo- geneities that produce the T2* effect occur at tissue borders, particularly at air/tissue interfaces, or are induced by local magnetic fields (e.g. iron particles). The loss of the MR signal due to T2* effects is called free induction decay (FID). T2* effects can be avoided by using spin echo sequences.

T2 denotes the process of energy transfer between spins, while T2* refers to the effects of additional field inhomogeneities contributing to dephas- ing.

T1 and T2 relaxation are completely independent of each other but oc- cur more or less simultaneously! The decrease in the MR signal due to T2 relaxation occurs within the first 100–300 msec, which is long before there has been complete recovery of longitudinal magnetization M

z

due to T1 re- laxation (0.5–5 sec).

References

Gore JC, Kennan RP (1999) Physical principles and physiological basis of magnetic relaxation. In: Stark DD, Bradley WG Jr (eds). Magnetic resonance imaging, 3rd ed.

Mosby-Year Book no 33, Mosby, St. Louis

Elster AD, Burdette JH (2001) Questions and answers in magnetic resonance imag- ing, 2nd ed. Mosby, St. Louis

Elster AD (1986) Magnetic resonance imaging, a reference guide and atlas.

Lippincott, Philadelphia 1.

2.

3.

Riferimenti

Documenti correlati

Upon setting the measuring mode of approach-retraction curves, calculation of oscillations amplitude, resonance frequency (when maximum amplitude is reached) and phase as a

The former consisted in the absolute phase measurement of a voltage, with respect to the cosine waveform with positive peak aligned with the PPS signal (see Fig. The

of the cases were diagnosed as indolent lymphoma, while the remaining were referable to nodular hyperplasia, either purely lymphoid or complex type. Notably, none of the dogs

La persona non si riconosce più in un’immagine ma si rappresen- ta a raverso un’infi nita serie di immagini, pro- prio come, allo stesso modo, lo spazio e le cose si

For particles with initial positions uniformly distributed in S 2π , while initial veloc- ities are distributed in equally spaced (by ∆v 0 ) beams containing one particle each, i.e.,

In addition, since bal- ance laws and constitutive prescriptions are kept distinct, it becomes easier to generalize the standard Gilbert equation in a manner consistent with the

8 Furthermore, some authors have compared the Perceval S with a new-generation sutureless rapid deployment valve (INTUITY; Edwards Lifesciences LLC, Irvine, Calif) and again

In this paper we are interested in the classification of Killing vector field on the tangent bundle T M endowed with an arbitrary g− natural metric G.. The same subject had been