• Non ci sono risultati.

8Failures with Bearings 8 8

N/A
N/A
Protected

Academic year: 2022

Condividi "8Failures with Bearings 8 8"

Copied!
6
0
0

Testo completo

(1)

Chapter 8 · Failures with Bearings – K.J. Bozic

8 Failures with Bearings

K. J. Bozic

Summary

Despite the long-term clinical and radiographic success that has been reported with total knee arthroplasty (TKA), failures related to the articular bearing surface continue to be one of the most significant factors that limit survivorship of TKA implants. Many variables in- fluence wear and bearing surface failure,including factors related to the patient,the surgeon,and the implant.Future efforts should be directed at changes in the materials, implant design, sterilization methods, and surgical tech- nique that could potentially lead to improvements in wear properties and an overall reduction in the incidence of bearing surface failures in TKA.

Introduction

Total knee arthroplasty (TKA) has emerged as one of the most successful and cost-effective interventions in or- thopedic surgery. Clinical and radiographic success rates of greater than 90% at 10- to 15-year follow-up have been reported [7, 11]. Despite the high rates of success, howev- er, problems related to the articular bearing surface have been one of the most common causes of failure. Fatigue and delamination are the most frequently reported mech- anisms of bearing surface damage in TKA [19,26].Factors that influence wear rates and bearing surface damage in- clude factors under the control of the surgeon, including

patient selection, implant selection, and surgical tech- nique, and factors under the control of the manufacturer, including material composition, manufacturing tech- nique, sterilization technique, shelf life, thickness, con- formity, and issues related to modularity. The purpose of this chapter is to review the literature regarding the most common causes of bearing surface failure associated with TKA.

Mechanisms of Bearing Surface Failure

Bearing surface failures have been associated with artic- ular wear, osteolysis, and problems related to modularity, including locking mechanism failure and backside wear.

Mechanisms of Wear Debris Generation

McKellop and colleagues have defined four modes of wear debris generation [18] (Table 8-1).Mode 1 describes wear that occurs between the two primary bearing sur- faces, as intended by the designers of the implant. This type of wear occurs when the femoral implant articulates with the polyethylene articular insert.Mode 2 refers to the condition of a primary bearing surface rubbing against a secondary surface in a manner not intended by the de- signer, such as when the femoral component wears through the articular insert and wears against the tibial

Table 8-1.Modes of wear in prosthetic joints (from [18])

Wear mode Description Example

1 Wear that occurs between the two primary bearing surfaces, Femoral component articulating with the tibial insert as intended by the designers of the implant

2 Primary bearing surface rubbing against a secondary surface Femoral component wears through articular insert in a manner not intended by the designer and articulates against the metal base plate

3 Third body wear that occurs between the primary bearing PMMA, metallic debris, or bone chips in joint space between surfaces due to particulate debris femoral component and tibial insert

4 Wear that occurs between two secondary (nonbearing) Backsided wear that occurs between the UHMWPE articular

surfaces insert and the tibial base plate

(2)

8

debris such as polymethylmethacrylate (PMMA) frag- ments, bone chips, or metallic debris from porous coat- ings. Mode 4 refers to wear that occurs between two sec- ondary (nonbearing) surfaces. This would include ‘back- sided’ wear that occurs between the UHMWPE articular insert and the tibial base plate [17, 20, 32].

Wear mechanisms associated with polyethylene ar- ticular inserts have been studied extensively in hip and knee replacement [1, 2, 4, 26, 31]. Wear mechanisms that have been associated with hip and knee arthroplasty in- clude abrasive wear, adhesive wear, third-body wear, de- lamination, and fatigue. Abrasive wear results if particles are generated when imperfections in a harder surface (e.g., metal femoral component) plow grooves into a soft- er material (e.g.,UHMWPE articular insert) [15].This can occur either at the primary articulation or at other sec- ondary surfaces. Adhesive wear occurs when small sub- micron particles adhere to the metallic counterface and are pulled off by the passing of the adjacent articulation surface [15]. Third-body wear occurs when third-body particles, such as cement, hydroxyapetite (HA), metal, and/or polyethylene, become embedded in the articulat- ing bearing [26]. Adhesive and abrasive wear are most commonly associated with total hip arthroplasty, leading to biologically active sub-micron particulate wear debris that can result in osteolysis and peri-prosthetic bone loss [18, 28].

The most common mechanisms of wear and poly- ethylene damage in total knee arthroplasty are delami- nation and fatigue, whereby the formation of subsurface cracks leads to the generation of particles that are shed from the bearing surface [15]. Delamination can result in the loss of conformity of the articular bearing, leading to an altered pattern of load distribution and ultimately to failure of the articular bearing surface [19] (Fig. 8-1).

Ultra-high molecular weight polyethylene (UHMWPE) has been the articular bearing surface of choice for total knee replacement for the past 30 years [5]. Polyethylene wear can occur at both the articular surface and the undersurface of the articular insert at the interface of the tibial base plate (so-called backside wear). Many factors affect bearing surface wear, including factors associated with manufacturing, sterilization and shelf life of the im- plant, implant design factors, patient factors, surgical technique, and duration of time in vivo.

Manufacturing Technique

Several investigators have examined the effect of specific resins on wear rates associated with articular bearing surfaces in TKA. In a retrieval study,Won et al. examined the effect of resin type and manufacturing method on wear rates of Miller-Galante (M-G) I and II tibial inserts that were gamma irradiated in air [36]. Their analysis revealed that M-G I retrieved inserts,which were made by direct compression molding of Hi-fax 1900 resin,had sig- nificantly more wear damage in the form of scratching and embedded metallic debris,whereas M-G II retrievals, which were manufactured by machining from a ram- extruded rod of GUR 415 resin, had significantly more wear damage in the form of delamination. Light-micro- scopic examination of thin sections of the retrieved im- plants revealed that delamination of M-G II components occurred through a subsurface region of severely oxida- tively degraded UHMWPE, while no such subsurface- degraded region was observed in M-G I retrievals. Based on their findings, the authors concluded that wear of UHMWPE tibial inserts is influenced by both resin type and manufacturing technique.

Sterilization Technique

Numerous in vitro and in vivo studies have been per- formed to evaluate the impact of sterilization techniques, such as gamma irradiation in air and ethylene oxide (EtO) gas,on wear rates in total knee arthroplasty [5,8,16,22,33, 35]. White et al. used qualitative and quantitative tech- niques to evaluate the effects of sterilization methods on the wear and physical and mechanical properties of 29 re- trieved UHMWPE tibial inserts [33]. They reported high- er rates of delamination, lower toughness, and lower per- cent elongation in inserts that were sterilized with gam- ma radiation in air compared with tibial inserts that were sterilized with ethylene oxide. Based on these findings, the authors concluded that EtO sterilization caused less

Fig. 8-1.Explanted tibial insert demonstrating extensive delamination of the articular surface

(3)

Chapter 8 · Failures with Bearings – K.J. Bozic

microstructural damage to polyethylene and resulted in significantly less wear than sterilization with gamma radiation in air.Reeves and colleagues used Fourier trans- form infrared analysis and finite element modeling to compare wear rates between gamma-irradiated in air and gas-plasma sterilized UHMWPE tibial inserts [22]. They found that under high cyclic stresses, delamination oc- curred in the majority of the inserts gamma-irradiated in air but in none of the gas-plasma-sterilized inserts.

In a retrieval study of 1635 UHMWPE tibial inserts, Williams et al. found that those gamma irradiated in air had a high incidence of delamination and cracking, lead- ing at times to complete wear through of the bearing, while inserts that were sterilized with EtO showed no evidence of fatigue damage, even after in vivo clinical use longer than 15 years [35]. Collier and colleagues used mechanical testing and Fourier transform infrared spec- troscopy to study how gamma sterilization leads to degra- dation of the physical and mechanical properties of UHMWPE tibial inserts [8]. Their results indicated that gamma sterilization in air resulted in elevated oxidation of polyethylene, which reduced static strength and elon- gation properties and significantly decreased fatigue resistance of polyethylene bearings.

The overwhelming conclusion from most of the in vitro and in vivo studies regarding sterilization technique in total knee arthroplasty has been that gamma irradia- tion in air has an adverse effect on the physical and me- chanical properties, thus leading to higher wear rates of UHMWPE articular bearing surfaces. In fact, in a review article on the subject that was published in 2002, Blunn and colleagues indicated that “delamination of loaded UHMWPE…will not occur in normal use over ten years if the UHMWPE has been well compacted and has not been sterilized by gamma irradiation in air” [5].

Shelf Life

Shelf life is defined as the period of time between when the implant is sterilized and packaged and when it is in- serted into the patient [6]. Kurtz et al. examined the ef- fects of shelf aging following gamma irradiation in air on the mechanical properties of UHMWPE tibial and ac- etabular inserts [16]. Using finite element modeling tech- niques, they reported that post-irradiation aging during shelf storage of UHMWPE inserts is likely to worsen long- term wear. Similarly, Bohl et al. investigated the effects of shelf aging on the in vivo performance of 188 gamma- sterilized UHMWPE Synatomic tibial inserts [6]. They found clinical failure (defined as component retrieval be- cause of polyethylene degradation) rates ranging from 20.8% for prostheses that had shelf lives before implanta- tion of 8–11 years to 0% for prostheses that had shelf lives of less than 4 years. The authors concluded that longer

shelf life had an adverse effect on the mechanical proper- ties of UHMWPE tibial inserts. Based on these findings, they recommended component expiration dates should be placed on implant packages and that implants that are shelf aged beyond this date should be discarded.

Other Design Factors (Implant Thickness, Conformity, and Type of Polyethylene)

Bartel et al.investigated the effect of conformity and plas- tic thickness on contact stresses in metal-backed hip and knee implants [4]. Using both analytical and finite element methods, the investigators found that for metal- backed components, minimum thickness of less than 6 mm for non-conforming surfaces and less than 4 mm for nearly conforming surfaces may result in excessively large contact stresses. Furthermore, their results demon- strated that bonding of the plastic to the metal backing decreased the tensile stresses at the edge of the contact zone and the maximum shear stress immediately under the load, suggesting a potential benefit of non-modular metal-backed tibial components.

In a follow-up study from the same research lab, pub- lished in 1986, Bartel and colleagues used elasticity and finite element solutions to estimate the effect of confor- mity, thickness, and material on contact stresses and sur- face damage in UHMWPE hip and knee implants [3].

They found that stresses associated with surface damage in the tibial component of the total knee replacement were much greater than those in the acetabular compo- nent of the total hip replacement. Also, their analysis of contact stresses of the polyethylene insert for tibial com- ponents suggested that a thickness of more than 8–10 mm should be maintained whenever possible. Other interest- ing findings were that contact stresses in the tibial com- ponent were reduced most when the articular surfaces conformed in the medial-lateral direction, and were much less sensitive to changes in geometry in the anteri- or-posterior direction. Finally, they found that the use of carbon-fiber reinforced PE resulted in stresses that were greater by as much as 40%.

In a separate study, Wright and Bartel analyzed sur- face damage in retrieved TKA tibial inserts and compared these findings with analytical studies to assess the influ- ence of thickness, articular conformity, and type of poly- ethylene on wear patterns [38]. With respect to these de- sign parameters, they found greater surface damage in thin components (less than 4–6 mm), components with relatively flat tibial articulating surfaces, and carbon-re- inforced polyethylene components.

Sathasivam and Walker used computer modeling techniques to simulate static and dynamic loading of tib- ial inserts in total knee arthroplasty [25]. Their results in- dicated that much lower contact stresses, and therefore

(4)

8

Patient Factors, Length of Time In Vivo

Several authors have suggested that wear of tibial articular inserts is affected by patient factors, including weight and duration of time in vivo [37, 38]. In 1986,Wright and Bartel observed patterns of surface damage from retrieved total knee polyethylene components and compared their find- ings with analytical studies of contact stresses in articular inserts [38].They reported that the amount and severity of surface damage to the tibial insert increased with patient weight and with length of time in vivo.

In a separate study, Wright and colleagues used light microscopy to evaluate surface damage to carbon fiber- reinforced and non-reinforced UHMWPE tibial inserts that had been retrieved from failed posterior stabilized total knee prostheses [37].They reported that the amount of surface damage was directly proportional to the dura- tion of time in vivo for both types of implants.

Surgical Technique

Wasielewski et al. examined 55 unconstrained polyethyl- ene tibial inserts that were retrieved at revision TKA for evidence of wear [31]. They found the most severe wear patterns in inserts with third-body wear from metal- backed patellar failures and cement debris. Furthermore,

the varus knee and lateral in the valgus knee).The authors concluded that wear associated with unconstrained TKA is affected by clinical and mechanical factors under the surgeon’s control,including component size and position, knee alignment, and ligament balance (Fig. 8-2).

D’Lima and colleagues used finite element modeling techniques to assess the effect of malalignment on poly- ethylene contact stresses in total knee arthroplasty [10].

They found that increased conformity significantly re- duced contact stresses in neutral alignment, liftoff signif- icantly increased contact stresses under both low- and high-conformity conditions, and malalignment in rota- tion was especially detrimental with the high-conformi- ty insert design.Based on these findings,the authors con- cluded that both implant design and surgical technique can play a critical role in wear rates in TKA, and that mo- bile bearing designs could result in lower contact stress- es by minimizing the effect of rotational malalignment.

Problems Related to Modularity

Another issue that has been debated in the literature is the benefits and drawbacks associated with modular tibial implants [24, 27, 29]. Proponents of modularity cite the ability to alter soft-tissue balancing intra-operatively and the potential for less traumatic revision surgery as bene- fits of modularity [27,29].Opponents of modularity point to potential problems associated with backside wear and incompetent locking mechanism designs as arguments against the use of modular tibial inserts [21, 24].

Backside Wear

Many investigators have demonstrated that in addition to articular surface damage and wear, backside wear be- tween the articular insert and the metal tibial base plate is a major contributor to wear debris in TKA [17, 20, 24, 30] (Fig. 8-3). Li et al. performed an in vitro analysis of 55 retrieved tibial inserts from four different TKA designs [17]. Their results suggested that backside wear of tibial inserts can be a significant contributor to polyethylene wear in TKA, and as a result, surgeons and manufactur- ers should pay close attention to the fixation of tibial in- serts to metal trays.

In a review article on the subject published in 2002, Wasielewski reported that the forces at the undersurface articulation created during physiological loading are in- fluenced by insert type, articular design, and surgical technique [30]. He noted that increasing articular insert constraint, while reducing contact stresses at the point of contact on the articular surface, can actually increase

Fig. 8-2.Anteroposterior radiograph of a TKA demonstrating medial compartment wear secondary to incomplete ligament balancing of a varus knee. Note the extensive osteolysis in the proximal tibia

(5)

Chapter 8 · Failures with Bearings – K.J. Bozic

backside wear,due to transfer of forces to the interface be- tween the articular insert and the tibial base plate. Also, designs with a cam-post mechanism, such as posterior cruciate substituting designs, create a significant shear force at this interface (Fig. 8-4). Furthermore, he re- ported that factors under the control of the surgeon, in- cluding component alignment and position and ligament balance, may also influence backside wear.

Rao et al. analyzed 29 retrieved modular tibial com- ponents from 12 fixed bearing designs with regard to backside wear and relative motion between the polyeth- ylene insert and the metal base plate [20]. Their results suggested that backside wear was correlated with the rel- ative motion between the polyethylene insert and the metal base plate.Based on their findings,the authors con- cluded that new locking mechanism designs directed toward better securing of the polyethylene insert to the tibial tray are needed to minimize the generation of particular wear debris at the modular interface.

Dissociation of Tibial Insert

Although rare,several case reports have been published re- garding patients who suffered dissociation of the modular tibial insert from the metal base plate in fixed bearing knees [9, 23]. In a report of two cases of dissociation of a modu- lar PCA insert,Davis et al.postulated that anterior liftoff of the insert during knee flexion was the cause of the dissoci- ation [9]. Ries recently reported a case of a constrained condylar insert-base plate dissociation that occurred as a result of anterior cam-post impingement,leading to failure of the posterior locking mechanism and posterior liftoff [23] (Fig. 8-5).These authors suggested that use of a mod- ular constrained condylar knee may not be appropriate for patients with a deficient extensor mechanism.

Osteolysis

Osteolysis, although less common in TKA than in total hip arthroplasty (THA), is a major cause of TKA failure leading to the need for revision surgery. The mechanism of periprosthetic bone loss associated with particulate debris has been critically evaluated in both in vitro and in vivo studies [12–14]. Willert and Semlitsch first pro- posed that wear-particle generation and migration into the joint cavity and periprosthetic space may stimulate macrophage recruitment and phagocytosis [34]. Since that time, further research has revealed that sub-micron wear particles are phagocytosed by macrophages, result- ing in the release of various pro-inflammatory factors and

Fig. 8-4. Explanted tibial insert from a posterior stabilized TKA demon- strating fracture of the tibial post. (Courtesy of Steve Haas)

Fig. 8-5.Explanted tibial insert and base plate demonstrating disso- ciation and displacement of the tibial insert secondary to failure of the base-plate locking mechanism. (Courtesy of Michael Ries)

Fig. 8-3. Explanted tibial insert demonstrating backside wear sec- ondary to motion between the tibial insert and the metal base plate

(6)

play a significant role in osteolysis are interleukin-1 (IL-1), interleukin-6 (IL-6), tumor necrosis factor-α (TNF-α) and prostaglandin E2(PGE2) [13]. Proliferation of these cellular mediators then leads to stimulation and differentiation of osteoclasts and inhibition of os- teoblasts. These factors work synergistically, ultimately leading to the dissolution of bone at the prosthetic inter- face, allowing for prosthetic micro motion that leads to further generation of wear debris [14].

In vitro studies have demonstrated that many factors influence the biological response to wear debris, includ- ing the size,volume,surface chemistry,and material com- position of the particles [2]. Many investigators have not- ed that osteolysis is less common in TKA than in THA [12, 15, 26]. In THA, adhesive and abrasive wear mechanisms dominate, resulting in the formation of high volumes of sub-micron particulate debris. Conversely, in TKA, fa- tigue and delamination are the most common mecha- nisms of bearing surface damage. These modes of failure produce wear debris particles that are larger than the wear particles observed around total hip replacements. It has been hypothesized by Ayers and others that the more aggressive biological response seen with THA could be explained by the fact that sub-micron particles provide a greater stimulus to the macrophage to produce inflam- matory mediators that result in osteolysis [2].

References

1. Ayers DC (2001) Maximizing ultra high molecular weight polyethylene performance in total knee replacement. Instr Course Lect 50:421–429 2. Ayers DC (1997) Polyethylene wear and osteolysis following total knee

replacement. Instr Course Lect 46:205–213

3. Bartel DL, Bicknell VL, Wright TM (1986) The effect of conformity, thick- ness, and material on stresses in ultra-high molecular weight compo- nents for total joint replacement. J Bone Joint Surg [Am] 68:1041–1051 4. Bartel DL, Burstein AH, Toda MD, Edwards DL (1985) The effect of confor-

mity and plastic thickness on contact stresses in metal-backed plastic implants. J Biomech Eng 107:193–199

5. Blunn G, Brach del Preva EM, Costa L, Fisher J, Freeman MA (2002) Ultra high molecular-weight polyethylene (UHMWPE) in total knee replace- ment: fabrication, sterilisation and wear. J Bone Joint Surg [Br] 84:946–949 6. Bohl JR, Bohl WR, Postak PD, Greenwald AS (1999) The Coventry Award.

The effects of shelf life on clinical outcome for gamma sterilized poly- ethylene tibial components. Clin Orthop 376:28–38

7. Colizza WA, Insall JN, Scuderi GR (1995) The posterior stabilized total knee prosthesis. Assessment of polyethylene damage and osteolysis after a ten-year-minimum follow-up. J Bone Joint Surg [Am] 77:1713–1720 8. Collier JP, Sperling DK, Currier JH, et al (1996) Impact of gamma steriliza-

tion on clinical performance of polyethylene in the knee. J Arthroplasty 11:377–389

9. Davis P, Bocell J, Tullos H (1999) Dissociation of the tibial component in total knee replacements. Clin Orthop 272:199–204

10. D'Lima DD, Chen PC, Colwell CW Jr (2001) Polyethylene contact stresses, articular congruity, and knee alignment. Clin Orthop 392:232–238 11. Faris PM, Ritter MA, Keating EM, Meding JB, Harty LD (2003) The AGC all-

polyethylene tibial component: a ten-year clinical evaluation. J Bone Joint Surg [Am] 85:489–493

pathogenesis. Instr Course Lect 49:71–82

14. Jacobs J, Rosebuck K, Archibeck M, Hallab M, Glant T (2001) Osteolysis:

basic science. Clin Orthop 393:71–77

15. Jacobs J, Shanbag A, Glant T, Black J, Galante J (1994) Wear debris in total joint replacements. J Am Acad Orthop Surgeons 2: 212–220

16. Kurtz SM, Bartel DL, Rimnac CM (1998) Postirradiation aging affects stress and strain in polyethylene components. Clin Orthop 350:209–220 17. Li S, Scuderi G, Furman BD, et al (2002) Assessment of backside wear from

the analysis of 55 retrieved tibial inserts. Clin Orthop 404:75–82 18. McKellop H, Campbell P, Park S-H, et al (1995) The origin of submicron

polyethylene wear debris in total hip arthroplasty. Clin Orthop 311:3–20 19. Muratoglu OK, Mark A, Vittetoe DA, Harris WH, Rubash HE (2003) Poly- ethylene damage in total knees and use of highly cross-linked poly- ethylene. J Bone Joint Surg [Am] 85 [Suppl 1]:S7–S13

20. Rao A, Engh G, Collier M, Lounici S (2002) Tibial interface wear in retrieved total knee components and correlations with modular insert motion. J Bone Joint Surg [Am] 84:1849–1855

21. Rao KS, Siddalinga Swamy MK (1989) Sensory recovery in the plantar as- pect of the foot after surgical decompression of posterior tibial nerve.

Possible role of steroids along with decompression. Lepr Rev 60: 283–287 22. Reeves EA, Barton DC, FitzPatrick DP, Fisher J (2000) Comparison of gas plasma and gamma irradiation in air sterilization on the delamination wear of the ultra-high molecular weight polyethylene used in knee replacements. Proc Inst Mech Eng [H] 214:249–255

23. Ries MD (2004) Dissociation of the UHMWPE Insert from tibial baseplate in total knee arthroplasty. A case report. J Bone Joint Surg [Am]

86:1522–1524

24. Rodriguez J, Baez N, Rasquinha V, Ranawat C (2001) Metal-backed and all- polyethylene tibial components in total knee replacement. Clin Orthop 392:174–183

25. Sathasivam S, Walker PS (1998) Computer model to predict subsurface damage in tibial inserts of total knees. J Orthop Res 16: 564–571 26. Schmalzried T, Callaghan J (1999) Wear in total hip and knee replace-

ments. J Bone Joint Surg [Am] 81:115–136

27. Shaw J (1992) Angled bearing inserts in total knee arthroplasty. A brief technical note. J Arthroplasty 7:211–216

28. Sinha R, Shanbhag A, Maloney W, Hasselman C, Rubash H (1998) Os- teoylsis: cause and effect. Instr Course Lect 47:307–320

29. Surace M, Berzins A, Urban R, et al (2002) Backsurface wear and def- moration in polyethylene tibial inserts retrieved postmortem. Clin Orthop 40:14–23

30. Wasielewski RC (2002) The causes of insert backside wear in total knee arthroplasty. Clin Orthop 404:232–246

31. Wasielewski RC, Galante JO, Leighty RM, Natarajan RN, Rosenberg AG (1994) Wear patterns on retrieved polyethylene tibial inserts and their relationship to technical considerations during total knee arthroplasty.

Clin Orthop 299:31–43

32. Wasielewski RC, Parks N, Williams I, et al (1997) Tibial insert undersurface as a contributing source of polyethylene wear debris. Clin Orthop 345:53–59

33. White SE, Paxson RD, Tanner MG, Whiteside LA (1996) Effects of steriliza- tion on wear in total knee arthroplasty. Clin Orthop 331:164–171 34. Willert H, Semlitsch M (1977) Reactions of the articular capsule to wear

products of artificial joint prostheses. J Biomed Mater Res 11:157–164 35. Williams IR, Mayor MB, Collier JP (1998) The impact of sterilization method

on wear in knee arthroplasty. Clin Orthop 356:170–180

36. Won CH, Rohatgi S, Kraay MJ, Goldberg VM, Rimnac CM (2000) Effect of resin type and manufacturing method on wear of polyethylene tibial components. Clin Orthop 376:161–171

37. Wright T, Rimnac C, Faris P, Bansal M (1988) Analysis of surface damage in retrieved carbon fiber-reinforced and plain polyethylene tibial compo- nents from posterior stabilized total knee replacements. J Bone Joint Surg [Am] 70:1312–1319

38. Wright TM, Bartel DL (1986) The problem of surface damage in polyeth- ylene total knee components. Clin Orthop 205:67–74

8

Riferimenti

Documenti correlati

To determine the causes of the emulsion mist effectiveness, the SEM analysis of the worn surfaces, which include the main flank (Fig. Studies were realized for the cutting speed

It remains to show that the partition exists when n is multiple of 8 and n ≥ 16.. For any non-negative integer x, we have a partition of {1

For example, if a chemical engineer is interested in investigating the effects of reaction time and reaction temperature on the yield of a process, and if two levels of time 1 and

Gli Autori descrivono un caso di artrite reattiva, caratterizzata da oligoartrite asimmetrica e dattilite, presentatasi in un paziente, affetto da carcinoma della vescica, sottoposto

La storia clinica del paziente risale a circa 20 anni fa quando comparivano i primi segni dell’in- grandimento acrale, iperidrosi, palpitazioni e ma- croglossia; ma è solo nel 1991

Il virus dell’epatite C può essere implicato nell’etiopatogenesi di alcune forme di artriti, della fibrosi polmonare e in rari casi nell’insorgenza o nel peggioramento

Although the new, high-sensitivity troponin assays offer greater diagnostic accuracy than the stan- dard assays (2), false positive results have been reported on various

In this study the evaluation focused on those parameters that (1) were previ- ously identified as intrinsic risk factors of ante- rior knee pain in prospective studies, and (2) can