• Non ci sono risultati.

4 Fast Imaging with an Introduction to k-Space

N/A
N/A
Protected

Academic year: 2021

Condividi "4 Fast Imaging with an Introduction to k-Space"

Copied!
15
0
0

Testo completo

(1)

4 Fast Imaging with an Introduction to k-Space

Joseph C. McGowan

J. C. McGowan, PhD

Associate Professor of Electrical Engineering, Dept. of Electrical Engineering, United States Naval Academy, Annapolis, MD 21405-5025, USA

CONTENTS

4.1 Introduction 41 4.2 k-Space 42

4.3 Spin-Echo Acquisition with Multiple Echoes 48 4.4 Fast Spin-Echo 48

4.5 Image Quality and Artifacts 50 4.6 Echo-Planar Imaging 52 4.7 Hybrid Techniques: GRASE 53 4.8 Burst Imaging 53

4.9 Parallel Imaging 54 4.10 Conclusion 55 References 55

4.1

Introduction

The quest for speed in diagnostic imaging is un- derstandable from all perspectives. The physician desires a “snapshot” of the anatomy or process in question and is concerned about the effect of physi- ologic motion on the examination. Repetition of the examination is often indicated when the process to be investigated is dynamic. The patient may be limited in ability to remain still, or may be uncomfortable during the imaging procedure. All participants desire a quick answer to the questions being addressed by the study. Finally, economic imperatives often favor a rapid exam.

It has always been possible to trade quality for speed, or vice versa, in magnetic resonance imaging.

For example, repeating an exam and averaging the results is a classic method of improving the quality of some images. This method may fail, however, in the presence of some types of physiologic motion.

Fortunately, much attention has been focused on methods to improve imaging speed while maintain- ing quality. Some improvements have been realized

via advances in technology. Specifi cally, improve- ments in signal preservation, noise rejection and gradient performance have allowed optimization of existing techniques with maximal yield of usable signal. The most impressive improvements, however, have been realized by fi nding means of circumvent- ing the physical limitations imposed by relaxation, some of which will be explained below.

Gradient-echo imaging may be thought of as the “original” fast scan technique. It was known that the inherent spin dephasing via T2 decay was a relatively longer process compared to the signal decay due to the imperfections of the magnet. When magnet technology improved, it was found that suf- fi cient signal to reconstruct an image could be ob- tained without refocusing the spins. Gradient-echo imaging could not, however, replace all conventional spin-echo imaging as it did not produce “pure” T2- weighting. Gradient and spin-echo techniques can be thought of as bases for all fast imaging tech- niques. In fact, it will be shown that all current fast imaging techniques are in some way hybrid forms of these two. To begin, we will trace the development of fast spin-echo imaging.

The success of magnetic resonance imaging (MRI) as a diagnostic modality can be attributed to a num- ber of factors, but prominent among them is that in many disease states the observed transverse relax- ation time (T2) is altered. Thus, the most clinically useful results in MRI are obtained with so-called T2 weighted contrast, where the intensity of an indi- vidual picture element (pixel) in the image refl ects the observed T2 of the tissue. For many years these studies were performed with the spin-warp pulse se- quence, discussed in an earlier chapter. Briefl y, in that sequence one chooses principally the repetition rate (TR) and the echo time (TE). Using relatively long TR and TE values, the signal intensity in regions with long T2 is higher, causing those regions to appear hyperintense on the resulting images. The TR must be long enough so that progressive saturation, which would contribute T1 contrast, is minimized to the de- gree necessary. The compromise, of course, is that the

(2)

exam time increases linearly with TR. It is possible to perform more rapid imaging using short TR con- ventional spin-echo or gradient-echo sequences, but pure T2 contrast cannot be obtained. It should also be noted that the conventional spin-echo sequence is relatively ineffi cient in terms of data acquisition, because the majority of the time spent imaging is actually spent waiting for the decay of longitudinal magnetization.

The resolution to the quandary posed above was found in a pulse sequence that preserved the long time between excitations (TR) but made use of the “dead time” to acquire more data with which to construct an image. These data, taken together, constitute the

“k-space” representation of the image. A basic under- standing of this concept is easily gained and clearly illustrates the evolution from conventional to fast spin-echo imaging. Images obtained with fast spin- echo (FSE, also called RARE and turbo spin-echo) techniques are now the most clinically essential. The ability to interpret and manipulate the contrast ob- tained with FSE is requisite to effective exploitation of the diagnostic capability.

In earlier chapters the development of gradient- echo imaging was attributed in part to the advances in magnet and signal-processing technology that made it possible to obtain images without spin-echoes. In a similar manner, an ultra fast gradient-echo tech- nique dubbed echo-planar imaging followed further technological developments. This again constituted a means of using most of the scan time to acquire data rather than wait for relaxation. It was also recognized that the spin- and gradient-echo techniques could be combined in a variety of methods that offer opportu- nities to optimize the trade-offs between speed and quality.

4.2 k-Space

During MRI or any imaging examination, data is col- lected for future display. The display can be visualized as a collection of numbers corresponding to bright- ness (or gray scale) arranged in a (typically) square array or matrix. A possible approach in imaging is to acquire an individual point of data representing some feature of anatomy or function and to use that data value to fi ll in the intensity on the image matrix.

Clearly, conventional X-ray imaging works in this way, even as all of the points are acquired simultaneously.

If, for example, a problem with the X-ray fi lm caused only half of the image locations to be obtained, the part of the image that was obtained would be perfect and indistinguishable from the corresponding part of a control image. Mathematically, it is said that there is a one-to-one correspondence between individual data points and spatial locations. It might be sur- prising to know that in magnetic resonance imaging, although it still true that the number of data points is equal to the number of image locations (picture elements, pixels), there is not a one-to-one corre- spondence. Rather, each data point that is acquired infl uences the entire image in some way. The image cannot be fully reconstructed (formed) without all of the data. Further, a mathematical process called the Fourier transform must be employed to translate the data into an image. Figure 4.1 is a diagram of this process in a “black box” sense.

Consider any MR image. If one wanted to send a perfect description of that image to a colleague, it would be possible to write down each pixel intensity on a gray scale from 0-256, agree in which order the numbers would be sent, and proceed to transfer the

Fig. 4.1. a The Fourier transform is used to process the data acquired in an MRI examination into an image. b Data ac- quired in an MRI imaging experiment is shown with the reconstructed image. The acquired data is complex and the magnitude of that complex data is shown

DATA from MRI Coil

Fourier

Transform

Image

a

b

(3)

numbers one at a time. The colleague could receive the numbers, reassemble the image into a matrix form (Fig. 4.2), and then display them as the gray scale equivalent. Clearly, an image is nothing more or less than a matrix of gray scale values (Fig. 4.2).

In “image space”, one can select a particular loca- tion and note the intensity. The coordinates in image space are simply the standard spatial coordinates in the image plane.

Similarly, the data collected in the MRI process is simply a list of numbers acquired over a length of time. During the examination, the voltages present in the RF coil are sampled and recorded by a computer.

A voltage waveform of a particular frequency can be characterized by magnitude and phase and, typi- cally, both parameters of the voltages are obtained using two radiofrequency channels. This is known as quadrature detection and results in data in so- called “complex” form. These data are assembled in what could be called the data matrix (Fig. 4.3). Each spin-echo can be recorded in this way as a collec- tion of numbers including the approach to the echo as well as its decay. Successive spin-echoes with dif- ferent strengths of phase encoding are recorded on their own line (Fig. 4.3c). Arranged in the data ma- trix (Fig. 4.3), the “locations” of individual data cor- respond to time, moving to the right with the prog- ress of the spin-echo, and outward from the center with the integrated magnitude of the phase encod- ing gradient. Since the recording of the spin-echo is associated with the application of the “read” or fre- quency-encoding gradient, it is also correct to refer to the location with coordinates of integrated gradi- ent magnitude. The center of the data matrix corre- sponds to zero gradient in both phase and frequency direction. It is the center, or peak, of the spin-echo acquired with zero phase encoding. Since stronger gradients contribute dephasing that in general low-

ers signal intensity, it is apparent that the center of the data matrix will have the largest intensity.

It may be seen at this point that the data matrix under discussion is no other than “k-space”, and that k-space is in some way an inverted form of image space. The coordinates of k-space can be thought of as times. The coordinates of image space are spatial in nature, but the gradients serve to establish a cor- respondence between frequency (or phase) and po- sition. Thus, the coordinates of image space can be thought of as having units of frequency. It is useful to note that the k-space matrix is theoretically symmet- ric. Clearly, the ideal spin-echo is identical in build- up and decay, yielding, in conventional coordinates, the left-right symmetry of the matrix. Additionally, the magnitude of dephasing due to a phase encode gradient does not depend on the sign (direction) of the gradient. Thus, applying a “positive one unit” of phase encoding is theoretically identical to applying a “negative one unit”, and thus the matrix is symmet- ric top to bottom. This two-dimensional symmetry is exploited in some techniques that shorten acquisi- tion time by acquiring only a part of the actual ma- trix and fi lling in missing data.

The Fourier transform, introduced above as a

“black box” that changes the data matrix to the image, can be described in a number of ways. Fourier theory holds that any function in time (the signal, the spin- echo) can be exactly described as the sum of sinusoi- dal functions. The individual sinusoids (of particular frequency) constitute the frequency components of the signal (Fig. 4.4) and the Fourier transform is sim- ply a mathematical tool used to convert one to the other. Thus, in terms of signal processing, the Fourier transform serves to relate the time domain represen- tation of a signal to its frequency domain equivalent.

The theory can be extended to more than one dimen- sion. In two-dimensional (2D) MRI, the 2D Fourier

Fig. 4.2. An image is a ma- trix of gray scale values, or intensities

(4)

transform serves to translate k-space to image space.

The mechanics of doing so are handled quite effi - ciently and rapidly by modern computers.

A pulse sequence for MRI can be thought of as a scheme for “covering” k-space, that is, for acquiring data such that all point of k-space are known. The spin-warp sequence is perhaps easiest to understand as each horizontal line of k-space is completely fi lled by the acquisition of one spin-echo. Successive spin- echoes with different phase encode values constitute the remaining lines and fi ll the matrix. When k-space is fi lled, the Fourier transform (albeit in two dimen- sions) of that matrix will directly yield the image (Fig. 4.5).

The k-coordinate, establishing the location in the k-space matrix, can be defi ned for constant gradients as:

k =γG(t)

2π (1)

where G is gradient strength as a function of time, γ is the gyromagnetic ratio, and t is the time of gradient application (Fig. 4.6). The read gradient controls kx and is turned on in conjunction with the collection of data. With the passage of time, Eq. (1) can be continu- ously evaluated as the k coordinate increases linearly.

Similar “motion” in k space is associated with all gra- dient manipulations in the pulse sequence.

The k-space matrix can be fi lled line-by-line, as has been described above, or via less conventional means, including as a spiral outward from the cen- ter, simply by varying the application of gradients.

As noted, application of a constant read gradient in the absence of a phase encode gradient is equivalent to moving to the right with time in the k-space ma- trix. Application of a constant phase encode gradient along with the read gradient would provide a diago- nal motion. To move back to the left one would apply a negative read gradient. Curved paths would be real-

b a

c

Fig. 4.3. a Intensity values represent signal strength at a point in time. A hori- zontal line in the data matrix represents a single spin-echo, with the center corresponding to the peak of the echo. b The spin-echo corresponding to the indicated line in the data matrix. c Different phase encode values correspond to different horizontal lines of the data matrix. The center in the vertical direction corresponds to the line obtained with zero phase encoding

(5)

3 2

1

0

1

1500

1000

500

0

0 2000 4000 6000 8000 10000 49 50 51 52

a b

Fig. 4.4a,b. Time-domain and frequency domain plots are related by the Fourier transform. Fourier theory holds that any func- tion may be represented by adding up a series of sin and cos functions. In MRI, the function is the echo, a complicated sum of components at different frequencies corresponding to the different spatial positions from which the signal is obtained. The echo is obtained in the time domain and depicted in (a), the time-domain plot. Recall that frequency encoding gradients establish a correspondence between frequency and spatial position. Extracting the signal that corresponds to a particular frequency is equivalent to isolating the signal from a particular position. In (b), the frequency-domain plot, the frequency components can be identifi ed by frequency and magnitude. It is diffi cult to discern these individual behaviors from the time-domain depiction

Fig. 4.5. When all echoes associated with all values of the phase encode gradient are acquired, k-space is covered and an image can be reconstructed. K-space can be covered by a variety of trajectories. A single echo in a spin-warp imaging sequence can be visualized as starting on the left edge of k-space and moving to the right edge

G

t

Fig. 4.6. The k-coordinate is proportional to the area under the gradient- time curve

(6)

ized by more complex time-varying gradients. Some of these are desirable from the standpoint of accu- rate and physically realizable fast gradient switching.

That is, one may choose a k-space trajectory based upon the physical limitations of the gradient ampli- fi ers. The original echo-planar implementation (dis- cussed below) is based upon applying to the gradi- ent amplifi ers the most basic waveform possible, a sinusoid. Curved trajectories may be less intuitive than the visualization of individual spin-echoes, but from the standpoint of the cumulative gradient effect they are relatively straightforward to interpret. The complicated k-space trajectories that are employed by modern pulse sequences, and indeed any k-space trajectories, can be created by controlling the gradi- ents in this manner.

The contribution to the fi nal image of specifi c data points may not be intuitively obvious, because

the form of the fi nal image is not discernible from the appearance of the k-space matrix. However, the effect of a single data point on the entire image can be estimated and simulated. The nature of that effect is related to the position of the data point in the k- space matrix. The implications of this observation are profound and essential to the understanding of artifact as well as to the proper prescription of the imaging study. For example, an artifact featuring par- allel stripes or a herringbone pattern may be seen on an image when a “pop” of static electricity is recorded during image data collection, and where only one or two points are abnormally high values. The orienta- tion and number of the stripes are functions of the location of the bad point(s) in k-space (Fig. 4.7). This arises from the fact that each data point can be as- sociated with a sinusoidal function that is part of the Fourier representation of the data, where the fre-

Fig. 4.7a–d. Striping (a,b) and Herringbone (c,d) artifacts result from a few “bad”, that is, abnor- mally high values in the MRI acquisition. The number and orientation of stripes are determined by the position of the bad points (arrows) in k-space

a b

c d

(7)

quency of the sinusoid (striping function) is higher for larger values of the k-coordinate.

An accurate and simple generalization is that the central area of k-space contributes primarily to im- age contrast while the peripheral area of k-space contributes primarily to the defi nition of edges. This can be illustrated by the simulated reconstruction of an image using only a portion of the data, with the remaining data points set to zero. As can be seen in Fig. 4.8, a reconstruction using only 6% of the original data points results in a blurry yet interpre- table image that would in some cases be diagnostic.

Omitting the center data points yields an image that

shows only edges (Fig. 4.8c). These observations give rise to some techniques for rapid image acquisition (for example, so-called keyhole imaging where outer points are “reused” in subsequent images).

To summarize, the signal is acquired over time by sampling the voltage present in the radiofre- quency coil. The frequency content of that signal reflects the encoding of the gradients, which estab- lish a correspondence between frequency and spa- tial position. The Fourier transform is employed to convert the time-domain signal into the frequency domain information that is decoded into spatial position.

a b

c

Baseline Reduced k-space

Fig. 4.8a–c. Reconstructing an image with a reduced set of points. a The control image reconstructed with all of k-space. b Using only the central 6% of k-space the blurry, yet recognizable, image (b) is formed. c Using the outer 94% of points, image (c) results

(8)

4.3

Spin-Echo Acquisition with Multiple Echoes Recall that conventional spin-echo imaging can be used to obtain T2-weighted diagnostic images. A long repetition time (TR) must be used to allow for the recovery of longitudinal magnetization and to avoid T1 weighting. Proton density images are typi- cally obtained in these studies because they come

“free”, that is, without any time penalty. Specifi cally, the time spent collecting the data to acquire proton density images is completely contained within the time spent waiting for T1 decay. A slight extension of this technique gives the general “multiple spin-echo”

acquisition, as discussed in Chap. 3 and diagrammed in Fig. 4.9. In this technique, the spin magnetizations are repeatedly refocused by successive inversion pulses. In the example four echoes are produced but more or fewer are possible. During each TR period, multiple lines of k-space data are acquired, one for each spin-echo that is formed. Each of these echoes is associated with the same phase-encode gradient, but they differ in echo time (TE). Thus, in each of the k-space matrices that correspond to specifi c echo times, the same line is fi lled in during the TR. The k-

space data are used to build up k-space matrices for each of the images as diagrammed in Fig. 4.9b. At the end of the acquisition there is enough information to reconstruct images for each of the TEs for which echoes were collected. A single completion of the ac- quisition sequence yields data for four images, and the time taken is the same as that required to collect data for the single longest image. This technique thus improves the effi ciency of image data collection by yielding multiple images in the same time that would be required to obtain one. However, greater than two multiple spin-echo images, for example, series with TE values of 30, 60, 90, and 120 ms, have not typi- cally been found to be clinically useful. The images that are most valuable are typically the long-TE (T2- weighted) images and perhaps the very short TE (or proton-density weighted) images. Standard imaging protocols for many years included proton-density images whenever T2-weighted images were obtained, even though they might have been considered of lim- ited utility. These images were considered “free” in that no time was added to the study to obtain them.

4.4

Fast Spin-Echo

Fast spin-echo imaging (originally called RARE, rapid acquisition with relaxation enhancement and also “turbo spin-echo”; Hennig and Friedburg 1986) was introduced to exploit the “available” time in the TR period and to apply this time to the acqui- sition of a single image rather than acquisition of many images. In principle, the idea is to take a mul- tiple echo sequence as illustrated above, and instead of writing the four lines of k-space in four separate k-space matrices, to use them all in a single k-space matrix, which thus is fi lled four times faster. In order to accomplish this the phase encoding must vary be- tween each of the echoes – otherwise the four lines of k-space will occupy the same points.

Clearly, the diffi culty with this idea is that the four echoes yielding data have different values of TE. In fact, in FSE imaging the k-space matrix is fi lled with lines representing echoes that differ widely in TE. To understand how this is possible one must recall that only a small portion of k-space determines the con- trast in the image, and that the contrast is the param- eter that is expected to change most dramatically be- tween echoes with different TE values. On the other hand, if the subject is stationary, one does not expect the edge information to change. Thus, in FSE imaging TE = 15

frequency – 128

0

phase

TE = 45

TE = 30

TE = 60 frequency

frequency frequency

– 128

– 128 – 128

0

0 0

phase

phase phase

RF

signal PE Gradient

Fig. 4.9a,b. In a multiple spin-echo sequence, one phase-en- code gradient application per TR is used (a). Four images are acquired when the PE gradient is incremented through full range, and the contrast in each image refl ects the TE of the echo associated with that image. K-space maps are shown in (b) for the fi rst TR period

a

b

(9)

there is not defi ned a single TE but rather an effective TE. The effective TE is the TE corresponding to the information obtained with low values of the phase encode (PE) gradient. Equivalently, the effective TE is the TE of the echoes where the dephasing due to the PE gradient is minimal, giving the higher intensity, central part of k-space. The decision regarding which echo to use for the contrast information is made au- tomatically as part of the prescription of the pulse sequence when the effective TE is chosen.

Consider FSE acquisition with four echoes. The rf excitation part of the pulse sequence is essen- tially identical to the four-echo multiple spin-echo sequence discussed above. The essential difference between the sequences is in the phase encoding gra- dient, applied before every echo. In each TR period, four lines of k-space are fi lled in the same image. An example is depicted in Fig. 4.10. Note in the example that the overall signal intensity increases over the TR period, that is, that the strongest echo is the last one.

That may be counter-intuitive given that T2 decay (dephasing) progresses during TR, an effect that con- tributes to loss of signal over time. Pure T2 decay is indeed present and is irreversible but it is relatively slow on the time scale of interest. In contrast a largely reversible dephasing is contributed by fi eld gradients, and it is this type of dephasing, specifi cally that of the phase encode gradient, that dominates the over- all strength of successive spin-echoes in FSE imaging.

Smaller values of PE gradient contribute less dephas- ing and thus result in stronger signals. In the exam- ple of Fig. 4.10 the value of the PE gradient decreases with successive echoes and thus the signal strength

increases. In viewing this example, recall that nega- tive gradient application can be used to decrease the overall integrated magnitude of any gradient. For completeness we note that T2* decay, the observed signal decay during the echo, includes magnet inho- mogeneities that make up most of the observed de- phasing of the individual spin-echo. This dephasing is to a large degree reversible and is in fact reversed when each spin-echo is formed.

Fast spin-echo can be employed with as many echoes as desired, as long as the signal is suffi cient at the end of the sequence to allow collection of data.

The number of echoes collected in each TR period is known as the echo train length (ETL) and represents the factor by which the total image acquisition time is divided. Equivalently, it is the factor by which the exam is speeded up. Repetition time (TR) is defi ned as in any pulse sequence and effective TE is defi ned as noted above. Echo spacing is defi ned as the time between successive echoes and any of these factors are determined by the other three. A useful relation- ship is:

Ta = TR* # PE

ETL (2)

where Ta is the acquisition time, #PE is the number of phase encode steps and ETL is echo train length.

FSE imaging is now routine clinical practice and produces diagnostic quality T2-weighted images in brain and spine (Fig. 4.11). The choice of ETL and other timing parameters is made with consideration of local equipment and constraints. Typically the

a

120 56

–8 –72 0

phase

b

RF PE Gradient

signal

frequency

Fig. 4.10. a In a fast spin-echo image, multiple phase encode gradient applica- tions are used in each TR. One image is acquired after all phase-encode values are applied. b A k-space map for the fi rst TR is shown. Here, the values of the phase-encode gradient are 120, –72, 56, –8. Ordering in this way makes the signal intensity of the echoes increase during TR, and makes the contrast of the resultant image refl ect the later echo

(10)

quality of images obtained at any individual center will dictate the ETL, made, all else equal, as large as possible.

The ultimate extension of FSE is single-shot FSE (SSFSE), whereby the entirety of k-space is fi lled in following a single initial excitation (π/2) rf pulse. This pulse sequence actually has no TR value, since there is no repetition, but may require only tens of seconds.

In current implementation the symmetry of k-space is exploited in SSFSE, with slightly more than half of the phase encode values suffi cing to fi ll in the data matrix.

The resulting images are of lower quality than gen- erally expected for diagnostic images, but are highly useful for “go/no go” evaluation of certain lesions and particularly in patients whose ability to remain still is limited. As shown in Fig. 4.12, SSFSE was employed to produce a diagnostic image in an uncooperative pa- tient where conventional SE and even FSE were simply too slow. Single-shot FSE in brain may not typically provide the quality needed by clinicians for routine evaluation, but continued improvements in scanner technology and signal processing should result in standard protocols featuring this technique.

4.5

Image Quality and Artifacts

A reasonable concern is the extent to which FSE im- ages are equivalent to those acquired with conven- tional spin-echo techniques. Clearly, to make this comparison the images must be acquired with the FSE effective TE values equal to the conventional TE values. That being assumed, FSE images are in fact used like the conventional spin-echo images that they were designed to replace. However, some differences do arise from a number of physical factors. First, the signals that fi ll k-space do originate from spin-echoes with different values of TE, and the effects of pure T2 decay make those echoes different. In the case of FSE with the longest echoes assigned to the low phase encode values, the effect is minimized because the strongest echoes (low phase encoding) will be the least affected by T2 decay during the TR period.

Thus, in FSE images with long effective TE, the im- age quality is highest because the high phase encode acquisitions, which are characterized by maximal dephasing and lowest signal strength, occur early in

a

b

Fig. 4.11a,b. Spine FSE imaging with near-isotropic voxels (a) at 1.5 mm resolution and (b) demonstrating excellent resolution of nerve roots

(11)

the T2 process. These acquisitions provide the high spatial frequency edge information. In the alternate scenario, PD-weighted FSE, the higher phase encode acquisitions must be assigned to the later echoes.

If signal strength is insuffi cient at this point in the sequence, the information from those echoes will be contaminated or dominated by noise, disrupting the reconstruction of edge information. The resulting images will be relatively blurred.

Another source of blurring common to all imaging modalities is described by the point-spread function, which relates the theoretical distribution of signal originating within a single pixel to adjacent pixels. In FSE, the point-spread function takes the form:

FW1/2 =∆y * 2 ETL * Es

π T2 (3) with FW representing the full width and half height of the function, this being a standard measure of linewidth such that a wider line corresponds to more

“spread” of the information, hence more blurring.

In this equation, ∆y is the pixel dimension, and the remaining parameters have been previously defi ned.

A similar issue related to the point spread function is ringing artifact, manifest as repeating bands related to the discontinuous T2 weighting function. Again,

this artifact results from the fact that signifi cant T2 decay can occur on the time scale of the phase encod- ing. Methods have been proposed and implemented to reduce these effects. The importance of the point spread function to the physician prescribing the MR scan is that the amount of blurring and ringing due to T2 decay can be controlled through the ETL and the Es.

Another concern in 2D FSE is magnetization transfer (MT). Consider that when multiple slices are prescribed, the data to reconstruct them is acquired in an interleaved fashion. Thus, the repetition time contains all of the echoes being acquired for one slice, and also the echoes for all slices in the study. For a given slice, the rf excitation for its adjacent slices can be considered “off-resonance” in the same sense as for the design of a magnetization transfer sequence. In conventional spin-echo, these effects are present but insignifi cant. In FSE, the rf pulses occur much more frequently. Here, the MT effects are measurable and tend to enhance the contrast achieved by T2 weight- ing. A related concern is the observation that many successive inversion pulses carry a heavy absorbed power load, and may be limited by United States Food and Drug Administration (FDA) or other govern- ment agency constraints. This limitation increases in severity with static magnetic fi eld strength, and is of

Fig. 4.12a–c. FSE (TR 4000 ms ETL 8) imaging was unsuccessful in an uncooperative patient (a).

SSFSE with 19-s acquisition time (6-mm slices, TE 97) was diagnostic (b,c) a

b c

(12)

note as the number of higher fi eld clinical scanners continues to rise. However, it is possible with cur- rent technology to acquire FSE data at high fi eld with careful choice of scan parameters (Fig. 4.13).

Susceptibility artifact has been observed to be mitigated by FSE imaging as opposed to conventional spin-echo. This may be related to the information that is incorporated into the image from the short echo phase encodes, presumably less affected by suscepti- bility-induced dephasing. One application where this observation is important is in post-surgical spine im- aging, where metallic fragments can cause diffi culty

in interpretation of conventional images. An example demonstrating the susceptibility advantage in the knee is provided in Fig. 4.14.

4.6

Echo-Planar Imaging

We have seen that acquiring more than one line of k-space from a single excitation pulse allows spin- echo to become “fast spin-echo”. In a completely

Fig. 4.134,b. Susceptibility artifact is mitigated by FSE (a) compared with conventional spin-echo imaging of the same knee. A surgical pin demonstrates a much more severe artifact in the con- ventional spin-echo image (b)

a b

Fig. 4.13a–c. Imaging at high fi eld in MS offers the potential for higher signal-to-noise ratio, better resolution and perhaps the detection of incipient MS lesions. A diagnostic scan at 1.5 T (a) is compared with a 4-T scan (b), also enlarged for clarity (c)

1.5 T 4 T 4 T

a b c

(13)

porate more shots by acquiring less echoes per repeti- tion period. The same is true in EPI. It is also possible to combine the techniques by, for example, incorpo- rating a refocusing rf pulse into an EPI sequence in order to restore strength of signal. This is the basis of so-called GRASE (gradient-echo and spin-echo) and related techniques (Feinberg and Oshio 1991).

All of the techniques can be viewed in light of their position on a 2D grid, with one axis assigned to the number of spin-echoes per shot (ranging from one to full k-space coverage) and the other axis assigned to the number of gradient-echoes per shot. Single shot techniques of FSE and EPI represent the maximum extent of the two axes, multi shot FSE and EPI oc- cupy intermediate points on those axes, and hybrid techniques including GRASE can reside anywhere within the plane defi ned by the axes. It is reasonable to surmise that an optimal imaging protocol can be found on this grid for all possible constraints.

4.8

Burst Imaging

Burst imaging is another class of ultra-fast imaging methodology, based upon the generation of a series of spin (and stimulated) echoes. A single shot tech- nique; it differs from those discussed above by the employment of a burst pulse originated as DANTE (delays alternating with nutations for tailored exci- tation). In this composite pulse a large number of evenly spaced, low fl ip-angle individual pulses are used. The basic pulse sequence is known as DUFIS (DANTE ultrafast imaging sequence) and employs the burst pulse given in the presence of a constant gradient (Hennig 1998). An inversion (180º) pulse follows to refocus (rephase) the transverse magneti- zation. Clearly, the train of excitation pulses creates a large number of echoes. However, as the spacing

Fig. 4.15. Timing diagram example for echo planar imaging

RF

Gz

Gy

Gx analogous manner, repeated gradient echoes can be refocused and obtained from a single excitation pulse resulting in so-called echo planar methods. As proposed in 1977 by Mansfi eld, the original echo- planar sequence was based upon a single excitation pulse (shot) and a zigzag trajectory through k-space brought about by the oscillation of one gradient (pro- ducing gradient-echoes) in the presence of a constant orthogonal gradient (Fig. 4.15). In current practice all similar techniques are referred to as EPI. A practical challenge with this technique is posed by the non- uniform coverage of k-space, but this is overcome by sophisticated reconstruction techniques. Other methods of covering k-space include the use of very rapid gradient switching (blipped gradients) to allow a k-space trajectory that is equivalent to spin-warp (at the cost of greater demands on system engineer- ing). In yet another variation, spiral trajectories are achieved by oscillations of both x and y gradients at more reasonable speeds. This reduces the need for extremely rapid gradient switching at the cost of additional complexity in image reconstruction and sparser sampling of the higher k-space regions.

Echo-planar imaging (EPI) as originally imple- mented is the fastest of the fast techniques owing to its reliance on a single excitation and having no need for other rf pulses for refocusing. As a variation, the single excitation can be replaced by a spin-echo or by some other brief preparation of the magnetization.

The EPI acquisition follows using all gradient-echoes and k-space is covered. Better quality can be obtained if necessary by performing two or more EPI acquisi- tions, with attendant linear increases in acquisition time. As with SSFSE, in single shot EPI the effective TR is infi nitely long but the entire study is accomplished in just a few seconds. Contrast is characterized by T2* or by whatever magnetization preparation was employed. The most rapid echo-planar imaging is vulnerable to artifacts including signifi cant blurring in the direction orthogonal to readout, as well as mo- tion and susceptibility. These challenges may impose limitations on the brevity of the acquisition time.

4.7

Hybrid Techniques: GRASE

We have seen that all fast imaging techniques can be evaluated based upon the number of acquisitions and the number and type of refocusing rf pulses. In all cases it is possible to take more time to acquire the image in order to gain quality. In FSE one can incor-

(14)

of the pulses is constant, many of the echoes are su- perimposed upon one another. Reconstruction algo- rithms account for the phase imposed by the constant gradient and assign the data to the proper location in k-space. Burst imaging is among the most quiet of MR sequences as it does not require rapid gradient switching.

4.9

Parallel Imaging

There has been a great deal of recent development and interest in parallel imaging, achieved by simul- taneously acquiring additional signals from coils arranged in an array around the tissue of interest (Sodickson and Manning 1997; Pruessman et al.

1999). Although parallel imaging is not based upon rapidity of acquisition, it deserves mention in this chapter because any increase in quality may be

“traded off ” in an engineering sense to permit an increase in speed. Parallel imaging may be viewed as an evolution of the employment of phased array coils, such as those in common use to image long struc- tures in the body (e.g. spinal cord). In pulse sequences that use these coils, each in turn acquires data from the area of the body in its vicinity. The reconstruc- tion algorithm combines the data from coils to make a seamless image of the entire structure. Although there is overlap, it is generally true that each receive coil is “responsible” for a different part of the image.

In contrast, parallel imaging uses multiple receive coils to provide simultaneous data on the same part of the anatomy, giving the benefi t of a repeat study for averaging but taking effectively no more time.

The reconstruction of this data into an image is chal- lenging but has been accomplished using techniques known as SMASH and SENSE. An example of parallel imaging is given as Fig. 4.16 and was obtained using 12 signal processing channels and a purpose-built parallel imaging head coil with 12 coils.

Fig. 4.16a,b. Comparison of standard imaging (a) and parallel imaging (b)using a 12 channel head coil (courtesy E. Knopp, New York University)

a b

(15)

4.10 Conclusion

Fast imaging is employed for much of the current clinical imaging in brain, offering the diagnostic power of long TE imaging, rapid acquisition times, and the ability to manipulate contrast as well as to tailor the study to the desired speed/quality compro- mise. An understanding of the essential parameters that defi ne these techniques can enhance the ability of the physician to prescribe and interpret imaging studies. Novel scan sequences based upon FSE, EPI, and other techniques, coupled with continued ad- vances in magnet and system technology, will con- tinue to extend diagnostic capabilities.

Acknowledgments.

The author is grateful for magnetic resonance images provided by his clinical colleagues at the University of Pennsylvania, and in particular to Dr. Robert I.

Grossman, friend and mentor.

References

Feinberg DA, Oshio K (1991) GRASE (gradient- and spin- echo) MR imaging: a new fast clinical imaging technique.

Radiology 181:597–602

Hennig JHM (1998) Burst imaging. MAGMA 1:39-48 Hennig JNA, Friedburg H (1986) RARE imaging: a fast imag-

ing method for clinical MR. Magn Reson Imag 3:823-833 Mansfi eld P (1977) Multi-planar image formation using NMR

spin echoes. J Phys C Solid State Phys 10:L55-L58

Pruessman KP, Weiger M, Sheidegger MB, Boesiger P (1999) SENSE: sensitivity encoding for fast MRI. Magn Reson Med 42:952-962

Sodickson, DK, Manning WI (1997) Simultaneous acquisition of spatial harmonics (SMASH):fast imaging with radiofre- quency coil arrays. Magn Reson Med 38:591-603

Riferimenti

Documenti correlati

The collection of ESEEM time and frequency domain data at different magnetic field values allows a more accurate determination of the relative orientation of the hyperfine

Figure 8.6: Tomographic image reconstruction of the content of the first metallic box, using the NEDA detectors inside the neutron beam cone

 verification approach: tries to assure the correctness of a detailed model M of the system under validation..  debugging approach: tries to find errors in a

For the detection and staging of a soft tissue mass, several pulse sequences are utilized: spin-echo T1-weighted, spin-echo T2-weight- ed, or fast spin-echo T2-weighted with

Gass A, Barker GJ, MacManus D et al (1995) High resolution magnetic resonance imaging of the anterior visual pathway in patients with optic neuropathies using fast spin echo

Spin-echo pulse sequence timing diagram where a pair of 90º and 180º ra- diofrequency (RF) pulses are used to cre- ate spin echo at echo time (TE)... The entire sequence is

b Phase encoding: Instead of acquisition of the complete signal during a single readout, the discrete data points can be acquired separately after applying gradients of

tial relevance and hence admissibility of brain images will vary, according to the specific legal issue at hand within civil and crim- inal contexts.. Put another way, the