Journal
of AutomaticChemistry,Vol.10,No. (January-March 1988), pp.31-36Evaluation of the Coulter Counter S-Plus VI
Giuseppe Banff,
MarinaPontillo,
Paola Notti andPierangelo
BoniniIstitutoScientificoS. Raffaele,Laboratorio Analisi, ViaOlgettina 60,Milano, Italy
This article reports an evaluation
of
the CoulterCounter
model S-PlusVI
automaticanalyser for
haematology, and data arepresented
onlinearity, carry-over,precision,accuracyandstabilityof
the instrument, whencompared
withamodel S-PlusIV/D.The
three-part differential
countprovidedby
Coulter S-PlusVI
has beencompared
with manualeyecounting. The results showagood.agreement
withonly
2.5%of
discrepanciesin2271 routinesamples.
Advantages of
the new instrument include: reductionof
runningcosts,
largely
due tomanpower
saving; simpleand easy use, and improved operatorsafety,
there beingno needfor
human contactwithblood.
Introduction
Performance of automatic instruments for
haematology
has beenconsiderablyimproved
inrecentyears,resulting in increased instrumentthroughput (number
ofsamples processed),
areduction inthe volume ofsample required,
andasimplerandmorestandardized interface withEDP.
There are two crucial aspects of automation in haema-
tology:
sampling and differential counting. The advan- tages to automating sampling include a reduction of manual work at atediousphase
intheprocedure
where risks ofsample
mismatching is high; standardization insample
mixing;and,
mostimportantly,areduction in the risksof operator infection. This isespecially
soifaclosed system with automatically pierceable caps is used.Unfortunately,
the need to maintainsample
homo- geneity, whileavoiding haemolysisin thepre-aspiration
andaspiration phases,
creates so many technicalprob-
lems that in most commercial instruments,sampling
is manual.Differential counting has been automated using either
computer-controlled
pattern recognition orby
means of rathercomplicatedandexpensive cytochemicalreactions.Two
newautomatichaematology
instruments havebeendeveloped recently by
Coulter(Models
S-Plus IV/DandS-Plus
VI).
Theseare abletoperform,
in additiontothe traditionalhaemocytometricmeasurements, asimplified
differential count with only threeparameters instead of the classicalfive, usingasample
sizeaslowas1001.
TheCoulter S-Plus
VI
is alsoprovided
with an automaticsampler (CASH),
containing up to32 tubes and with athroughput
of 90samples
perhour.In
the Coulter S-Plus IV/Dsampling
is manual.An
evaluation of these*Correspondence toDrBonini.
instruments is
presented
in thispaper; particular
atten- tionhas been given to the innovative aspects connected with differentialcountingandsampling.
Materialsand methods
The Coulter S-Plus IV/D is an automated multi- parameter
hematology analyser,
based on the Coulterprinciple
ofimpedence
counting. Like most of the previous S-Plus models it has two countingchambers,
operating simultaneously, one forplatelets
and eryth- rocytes and the other forleukocytes.
Each isprovided
withthree apertures, fortriplicate analyses.
Differentialcountingisdone
by
meansofaspecial
lysing reagent, which, in addition tocompletely
haemolysingerythrocytes
in the appropriate countingchamber, pro-
vokes a shrinking of the leukocytecytoplasm;
conse-quently, lymphocytes
are reduced to one half their original volume whilegranulocytes,
because of theircytoplasmic granules,
shrink toonly
two-thirds of their originalsize[2
and3]. A
thirdgroupof cells(monocytes
and other mononuclear
cells)
have volumes in between those oflymphocytes
andgranulocytes.
These three classes ofartificiallymodifiedleukocytes
arethen counted inthe white cell chamber anddifferentiatedonthe basis of their’apparent’
volume.Three
histograms
show. the size distribution for leuko- cytes,erythrocytes
andplatelets respectively.
The leuko- cytes arecategorized
into threesubpopulations" (1) lymphocytes
withvolumes between35-90 fLand claimed to include maturelymphocytes, atypical lymphocytes
andlymphoma cells; (2)
mononuclearcells with volumes between 90 and 160fL,
claimed to include monocytes,blasts, promyelocytes
andmyelocytes; (3) granulocytes
with volumes between 160 and 450 fL and includingpolymorphonuclears, bands, metamyelocytes
and bas- ophils[2
and3].
Whetheror not toincludeeosinophils
in the class ofmononuclear cells and/orgranulocytes
is a much debatedpoint.
This classification
ofleukocytes
isbasedontheanalysis of some 20000 cells. When the valleys between the threesubpopulations
are notclearly
defined in the histogram, the instrumentproduces
aflag. An
R1flag
indicates the presence ofparticles
smaller than 35fL,
which can be debris ofcells, platelet clumps, erythroblasts,
or non-lysed erythrocytes,
allproducing
a’hightake-off ofWBC
histogram.A
’backlighting’ of absolute values ofWBC
and ofLy, Mo, Gr
alsoappears.An R2 flag
indicates abnormal and/or immaturelymphocytes,
monocytosis, reactive lymphocytosis, eosi-nophilia
and/or blasts.An
R3flag
indicates high neutrophilia and/or abnormalgranulocytes. A
combina- tionof the above anomaliesproduces
aRM flag.
G. Banfietal. Evaluation of the CoulterCounter S-Plus VI
The Coulter S-Plus
VI
differs from the Coulter S-Plus IV/D in beingprovided
with an automaticsampler (CASH)
a mechanical device which simultaneously operates two rotating trays.Up
to 32specimens,
in additiontothecontrol4CPlus,
maybe inserted in each tray.A
CoulterS-PlusVI
hasbeen installed inourlaboratory
for the present evaluation, andoperated
for five weeks according to the manufacturer’s instructions, alongsidean already operating Coulter S-Plus IV/D. Theevalua- tionhas been
performed
accordingtotheICSH protocol [7].
An
internalquality
controlprogramme
operates daily usingthe4C Plusbloodcontrolandcheckingthemoving averages for various parameters.An
externalquality
control programme(Ortho
DiagnosticSystems,
Milan,Italy)
operates at twolevels: normal and abnormal.Samples
were collected in vacuum tubes(Vacutainer, Becton
Dickinson, Milan,Italy)
of 3 ml capacity con-taining
K3EDTA asanticoagulant, and,
except for thestudy
ofprecision
andstability,processed
between and5 hfrom drawing.
For
the study of linearity,plasma
and blood cellsseparated
aftercentrifugationfrom five routinespecimens
were mixed in different
proportions
and the resulting mixtures analysed intheCoulter S-PlusVI.
Within-runprecision
was evaluated by carryingout 10 determina- tionsonfive routinespecimens.Between-run
and overall precisionwascalculatedfrom10 determinations doneon one normal specimen every 60 min, up to 8 h after venipuncture,andfinallyafter24h. Thebloodwasstoredatroomtemperature. The same data have been used to evaluatethestabilityofa
sample
afterblooddrawing.To
evaluate the stability of the instrument, measurements obtained during 19working days on the samelot of4C Plus control materialwererecorded andcompared
with the assigned values.Carry-over
was evaluated on a routinesample
tested induplicate,
beforeand afterthree aspirations ofdiluent(Isoton III);
the sameexperiment
was
repeated
on the samesample,
after centrifugation andpartial
removal ofplasma.
For
checkingtheaccuracy ofdifferentialcounting,results from the Coulter S-PlusVI
werecompared
with those from(1)
areference method consistingofamicroscopic examinationbyawell-trained operator,ofsmearsstained withMay
Grunwald-Giemsa, and differential counting on500cells: 50samples
wereexamined in thisway;(2)
aroutine
method, consisting
ofamicroscopic
examination of smears stained withMay
Grunwald-Giemsa, and differential counting on 100 cells. These measurements wereperformed
on all the laboratory routinesamples (2271
intotal)
for five weeks.In
neither case did the operators know the results obtained with theCoulter S-PlusVI.
To compare
the overallperformance
ofthe CoulterS-PlusVI
and Coulter S-PlusIV/D,
150randomsamples
from inpatientsandout-patients weretestedwithboth instru- 32ments running under routine conditions, with normal quality control and manufacturer’s calibration.
Results Linearity
Plasma and blood cells from five routine
samples
have been mixed in differentproportion
and the followingextremevalueswereobtained: 0"1-15"4
x
109/1forWBC,
0"1-10"6
x
1012/1 forRBC,
10-473x
109/1 forPLT,
0-28"7g/dl
forHb,
0-87"3% forHct.
Theobservedvalues ofWBC, RBC, Hct, PLT
and Hb have beenplotted
against theexpected
values(figure 1);
the linearity is excellent for all the parameters within the ranges examined(WBC:y
-0"61+
1.04x,r0"996; Hb:y
0"41+
0"99x, r0"999; Hct:y
0.64+
0"99x, r0"999;
RBC:y
-0"12+
l’02x, r 0"994;PLT:y
26"8+
0"96x, r
0"994). As
figure 2 illustrates, a certain elevation oflymphocytes,
andacorresponding
reduction ofgranulocytes,
occurswhen the ratio betweenbloodcells andplasma
islowered,
whilearatherirregular
patternof monocytes canbe observed. Thesame occurs also when haematocrit iskept
constantwhile diluting leukocytesinplasma.
Thisphenomenon
isprobably
duetoamodifica-tionofsome
physico-chemical properties
ofWBC
due tocentrifugation and/or
plasma
removal as it issuggested by
the lack ofcorrespondence
between thetruevaluesof eachcategoryofleukocytes
andthevalues observedat asample
dilutioncorresponding
to the original(before dilution)
haematocrit (figure2).
Ifthis is the case, this fact is irrelevant intermsof thequality
ofresults,
asthere is no need for suchmanipulations
whenperforming
routinehaematology
tests. Furtherinvestigation
is needed.Precision
An
excellent level ofprecision,
even better than manu- facturer’s claims,isillustratedintables and 2.Theonly2C 16.
12
HOI /
12 16 20 2 28
o’"
Ix1
Expected
valuee/dl
(x:o’,;i)
Figure 1. Correlation between
expected
and observedvalues for
leukocytes, haemoglobin,
haematocrit,erythrocytes
andplatelets
inalinearitytest.
G.Banfietal. Evaluation of the CoulterCounterS-PlusVI
70.
gr’/.
10
mo.
4 6 . o
Pointsof dilution
Figure2. Values
of
the threeparametersdifferential
count inthefive samples
studiedfor
linearity,indicated withthe numbersfrom
1to5,at
different
dilutions. Verticalarrowsindicatethe dilution correspondingtothenativehaematocrit;horizontal
arrowsindicate thenativevalueof
each classof
leukocytesfor
eachsample;
insomecases, when values identicalto nativeoneisobtainedatmorethan onedilution,morethanone
horizontal
arrowisreported(see
Gr%and
Ly% for
patientn.3andMo%for
patientn.4).
Obviously,for
eachsample,thehorizontal
and verticalarrows wereexpectedto indicatethesamepoint, which wasnotthecasein mostsamples.exception
is mononuclear cells where an overall coef- ficientof variation of10% has beenobserved.Stability
The results ofstability test on one
sample presented
agood
stability until 6-7 h after blood drawing for all parameters, exceptMPV,
wheresome6-7% of increase isobservedjustafterh,
andMo%whichpresentsawide rangeofpercentvariation. The valuesobservedonthe4C Plus during 19 workingdays (table 3)
are within theassigned
rangesand arevery stable.Carry-over
Practicallynocarry-overhasbeen
observed,
evenin very extremeconditionsas inaconcentratedsample
withthe followingvalues: 18"3x
109/1forWBC,
8"75x
1012/1forRBC,
372x
109/1forPLT
and28"g/dl
forHb. Afteronemeasurement
cycle
on this concentratedsample,,
an analysisperformed
on the diluentIsoton III
gave these results: 0"x
109/1 forWBC,
0"04x
1012/1 forRBC,
3x
109/1 forPLT,
0"1g/dl
for Hb.Two subsequent
deter- minationsperformed
on the sameIsoton III
did not present abackground
forWBC, RBC,
Hb and a negligibleonein thePLT
count(1 x 109/1).
Comparisons
An
excellent correlation exists between the results obtained with Coulter S-PlusVI
and Coulter S-Plus IV/D(figures 3[a]
and3[b])
with a very limiteddispersion
of data for all the haematologic parameters except for mononuclearcells,
which show some disper- sion around the correlation line. Thefollowing regression equationsand coefficientswereobserved:WBC:y
0"9x,r
0"999; RBC:y
l’01x,r 1;PLT:y
1"43+
l’15x,r
0"998; Hb:y
l’01x, r1; Hct:y
0"02+
0"98x,r
0"999; MCV:y
-0"001+
1"058x,r 0"999;RDW:
Table 1. Within-run precisiononsixroutinesamples.
Within-runprecision
Samples SD CV% R SD CV% 2 SD CV% SD CV% SD CV% i SD CV%
Wbc x 109/1) Rbc(x1012/1) Ub(g/dl) Hct (%) Mcv (fl) Mch(pg) Mchc(g/dl)
Rdw(%)
Pit(x 109/1) Pct (%) Mpv(fl)
Pdw(%)
Ly(%) Mo(%) Gr(%) Ly( 109/1) Mo (x 109/1) Gr 10U1)
9"18 0"06 0’69 4"48 0"06 1.40 7"62 4"63 0"01 0"25 3"41 0"02 0"58 4.44 14"27 0"06 0"47 8"45 0"05 0"63 13"42 41"96 0"24 0"57 27"44 0"22 0"80 41"54 90"57 0"50 0"55 80"36 0"53 0"66 93"57 30"81 0.16 0"51 24"75 0"26 1"03 30"22 34"02 0"26 0"76 30"76 0.31 1"02 32"30 12"77 0"24 1"89 20"33 0"35 1"72 12’02 339"10 8.14 2"40 246"40 4"14 1"67 298’00
0"203 0"002 0"13 0"245 0"007 2"85
0"09 1"18 5"91 0’07 1"23 0"02 0"40 3"56 0’02 0"48 0"08 0"58 11.06 0"05 0"47 0"24 0"58 33"42 0"19 0"57 0"35 0"37 93"95 0"26 0"28 0"22 0"73 31"09 0"25 0"80 0"29 0"90 33"09 0"25 0"75 0"20 1"66 14"38 0"44 3"06 4"84 1"62 271"50
0’265 0"005 1"88 7"23 0"09 1"31 8"25 0"12 1.43 8"88
16"68 0"25 1.49 16"99 0"28 1"63 16"15 40"37 0.74 1.84 26"74 0"80 2"99 30"07 5"57 0.49 6"79 9"07 0"57 6"28 7"79 54.06 1"06 1"96 64"19 1"12 1.74 62"14 3"71 0"05 1.34 1"21 0"03 2"48 2"29
0"52 0.04 7"69 0"40 0 0 0’59
4"97 0"08 1"60 2’88 0"09 3"19 4"72
0.10 1.13 0.16 0.98 0.42 1.43 0.67 7.70 0.84 1.35 0.06 2.49 0.06 9.66 0.10 2.11
3.57 1.31 158.00
0.254 0.008 3.15
9.30 0.20 2.15
16.57 0.26 1.57
35.60 0.60 1.64
8.27 1.13 13.66
55.23 1.48 2.68
2.16 0.05 2.36
0.49 0.09 1.77
3.27 0.08 2.50
6.52 0.08 1.21 6.54 0-05 0.79
4.87 0.04 0.82 4.79 0.02 0.54
15.20 0.13 0.85 14.37 0.08 0.57
44.69 0.29 0.65 43.34 0.23 0.53
91.70 0.42 0.46 90.52 0.26 0.29
31.17 0.17 0.54 30.00 0.12 0.40
33.99 0.20 0.59 33.15 0.10 0.30
12.83 0.12 0.90 12.82 0.17 1.32
3.71 2.30 217.80 4.39 2.01
0.123 0.003 2.40 0.197 0.005 2.54
7.82 0.10 1.28 9.08 0.12 1.32
16.35 0.20 1.22 16.81 0.19 1.13
37.96 1.22 3.20 31.57 0.45 1.42
7.35 0.57 7.75 8.06 0.66 8.18
54.31 ,1.36 2.50 60.37 0.71 1.17
2.51 0-09 3.46 2.06 0.05 2.42
0.48 0.04 8.70 0.51 0.05 9.80
3.55 0.08 2.40 3.95 0.08 2.02
G.Banffetal. Evaluationof theCoulterCounterS-PlusVI Table2. Overall precisionon one routine
sample.
Overallprecision
CV% CV% CV% Total
X SD withinrun betweenrun overall CV%
Wbc(x
109/1)
8"90 0"220 1"57 2"47 2"92 2"50Rbc(x1012/1) 4"57 0"099 0"65 2"17 2"28 2"17
Plt(x109/1) 322"58 11"500 2"30 3"49 4"18 3"56
Hb(g/dl) 14" 10 0"195 2"86 1"00 3"05 1"38
Hct(%) 41"50 0"936 0"70 2"24 2"36 2"25
Mcv (fl) 91"40 0"370 0"53 0"36 0"63 0"41
Rdw(%) 12"78 0"335 1"37 2"58 2"89 2"62
Mpv
(fl) 7"90 0"490 1" 17 6"20 6"32 6"20Gr(%) 52"61 1"270 1"82 2’34 2"96 2"41
Mo
(%)
6"67 0"450 7’79 6’29 10"00 6"86Ly (%) 40"76 1"240 2"00 2"96 3"60 3"05
Table 3.
Mean,
rangeandstandard deviation valuesfor WBC, RBC,
Hb,Hct, MCV, PLT
in onesinglelotof
4CPlusduring19workingdays.
Stabilityof
4CPlus-Lot 9635-from
3February
1986to24February
1986.Observedvalues
Assignedvalue range Mean
Range
SDWbc 109/1) 7"9
Rbc (x1012/1) 4"24
Hb (g/dl) 13.1
Hct (%) 37.1
Mcv (fl) 87.5
Plt (x109/1) 211
_+4 7.79 7.4-8.2 0.20
+0.1 4.18 4.13-4.26 0.04
_+0.3 13.2 13-13.5 0.17
+1.5 36.84 36.4-37.5 0.36
_+2 87.96 86.9-89.1 0.53
+25 212 197-232 10.60
vvbc :oL rbc :orL pit m/L
L hct % rncv fL
/
,"
Coulter S-Plus IV/D
Figure
3(a).
Correlation between values obtained with Coulter S-PlusVI
and Coulter S-PlusIV/D on 150routinesamples for
WBC, RBC, PLT, Hb, Hct
andMCV.
CoulterS-PlusIV/D
Figure
3(b).
Correlation between values obtained with Coulter S-PlusVI
and Coulter S-Plus IV/Don 150routinesamples for
RD W, MPV, Gr, Mo
andLy.
y 0.002
+
0.99x,r0"999; MPV:y
-0.001+
1.058x,r
0"999; Gr:y
0"70+
l’02x, r0.999; Mo:y
0"025+
0"96x,r0"990; Ly:y
-0"01+
0"96x,r 0"999. The correlation between themicroscopic
observationby
a well-trainedpersonnel
on 500 cells on one side, and results from CoulterS-PlusVI
andCoulterS-PlusIV/D from the otheroneis ratherpoorfor mononuclearcells;
it isverygood, however,
forgranulocytes
andlymphocytes (figure 4).
The following regressionequations
and coefficientswereobservedfor CoulterS-Plus IV/Dversusmicroscopic
observation:Ly%:
y 2"95+
0"97x, r0"954;
Mo%:y
2"92+
0"32x,r0"259;
Gr% :y 4"38+
0"93x,
r 0"952 and for Coulter S-PlusVI
versusmicroscopic
observation"Ly%’y
3+
x, r0"955;
Mo% :y 2"52
+
0"39x,r0"292;
Gr% :y 2"47+
0"94x, r 0"944.Comparison
between differentialcountingsfromCoulter S-PlusVI
and routinemicrosccopic
observation over a five-week period(table 4)
showsthat, despite
theG.Banffetal. Evaluation of the CoulterCounter S-PlusVI
Coulter S-Plus
VI
Figure4. Correlation between values obtainedwith, respectively, CoulterS-PlusIV/D
(upper part)
andCoulter S-PlusVI (lower part)
andmicroscopicobservation.Table4. Correlation between
differential
counting obtained with the Coulter S-PlusVI
androutine microscopicexamination.Totalnumberofconsidered differentialcounts 2271 Discrepancies betweendifferential counts
from CoulterVIand fromeye-count (100
cells) 363 (16%)
ofwhich 2. Discrepancies dueto eosinophilia 3. Significant discrepanciesbetween
differential countsfrom CoulterVIand fromeye-count(100 cells),accordingto the Rumke’stable
4. Caseswithflagsorbacklighting 5. Caseswithdots ordashes (incomplete
computation)
6. Casescorrelationbetween thoseindicated in3 (above)and4
7. Falsepositives 8. False negatives
295 (13%)
++
(2.37%) 64(2.8%)
16 (0.7%) 39 (72.22%
ofcases in3) 43 (1.9%)
16 (0.7%)
apparently large
number ofdiscrepancies
in results(16%),
mainly due to eosinophilias(13%), only
in 54cases
(2"4%),
modifications of automatic differential countingafteramicroscopic
observation of100 cellsare really significant, according to the Rumke’s table[10]
and most of these cases
(72%)
wereflagged by
the automatic instrument. Sensitivity(false negatives)
andspecificity (false positives)
are0"7 and 1"9%respectively.
Discussion
It
wasemphasized
in the Introduction that for haema-tology,
automation of thepre-analytical phase (sampling)
and differentialcountingare the crucialproblems.
For sampling,
the excellent correlation between data observedonthe Coulter S-PlusVI
andtheCoulter S-Plus IV/D with the samesamples,
demonstrates that the adoption ofCASH
in the Coulter S-PlusVI
resulted ingood mixing
ofsamples
before theiraspiration
andproduced
manyoperational
benefits,compared
with manualsampling.
These included: reduction of manual work andpossible
errors; reduced risk of infection because of theuseofvacuumtubes and directaspirationby
the needle of theCASH,
and standardization ofsample
mixing. The adoption ofthree-parameter
dif- ferentialcounting, even ifgreatly
modifyingthe classical five-parameter pattern, in our experience isacceptable
both from an analytical and clinicalpoint
of view.Granulocyte
andlymphocyte
counts correlatevery well with manual countingperformed
on 500 cells by a well-trained operator. This agreeswith other datafrom the literature[1,
4 and5].
Thesophisticated
system offlags indicating
clinical abnormalities from thehistogram patterns givesconfidencefor theadoption
of this system- note, in table 4, the very low number of significant modifications of automatic differential counting neces- saryafterchecking by microscopic
examination.A
smaller number of falsepositives
andnegatives
were observed thanreported by
other authors[4
and9];
this wasprobably
due to the limited number ofcases with abnormalhaematology
inourpopulation. In
agreement withother authors[1,
4 and5]
we alsoobserved apoor correlation between automatically determined mono- nuclear cells andmicroscopic
counting of monocytes;even the precision is poorer than those of
granulocytes
andlymphocytes
and a greaterdispersion
of data is observed incomparing
data ofCoulter S-PlusVI
with Coulter S-PlusIV/D, partially
duetothe low number of thiscategoryof cells.As
discussedby
other authors[4
and6],
themononuclear cellsfraction,
ratherthancorrespondingto awell-defined category ofcells,
should be considered as a remainder fraction, including cells other than monocytes, such asblasts, promyelocytes
and myelocytes. These must be identified, when necessary, under themicroscope.Important advantages
of automatic differentialcounting are the greater reliability of results based on the examination of a muchlarger
number ofcells than isusually
counted under themicroscope (20
000for Coulter against 100 formicroscope),
and the elimination ofthe fastidiousandveryrepetitive microscopic
examinationof alarge
number ofsmears, many of which are normal inmostlaboratories.In
ourlaboratory,
after the presentevaluation, we have decided to limit microscopical differential counting tospecimens
where aflag
for differential counting and/or alterations of an other haematologic parameter are present(approximately
30% of our routinework). In
addition, theclinicalpathologist
obtainsmuchclinically useful information from carefulstudy
of thehistograms.This aspect will be
reported
in future papers.Overall,
there is a considerable time saving for clinicalpatholo-
gists. The clinician receives an haematologic report for eachpatient
based on reliable differential counting ofG. Banffetal. Evaluation of theCoulterCounterS-PlusVI
granulocytes
andlymphocytes [8]
andon acriticalstudy
of the Coulter histograms.In
summary, and in agreement with manufacturer’s claims and data inliterature, ourfindings
confirmgood
linearity, stability(up
to8 h afterdrawing)
andprecision, with negligible carry-over. The instrumentthroughput,
thevery lowsample
sizeand reagentconsumption have notbeenreported
in detailbutcorrespond
tothe claimsof manufacturer. The instrumentwas easy tooperate and was wellaccepted by
the laboratorystaff.6.
GP.ISWOLD, D.J.
and CHAiVIPAGNE,g.D.,AmericanJournal of
ClinicalPathology, 84(1985), 297.
7. InternationalCommitteefor Standardization inHaematol- ogy (ICSH), Clinical andLaboratory Haematology, 6 (1984), 69.
8.
KOEPK.,J.
A.,LaboratoryMedicine, 11 (1980), 371.9. NELSON,
L.,
CHARACHE,S., KEYSER, E.andMm’zER, P.,
American
Journal of
ClinicalPathology,83 (1985),547.10.
RUMKE,
C. L., inDifferential
Leukocyte Counting. Edited by Koepke,J.
A. (Aspen/College of American Pathologists, 1977), 39.References
1. CORNBLEET, S. and KESSINGER, S., American
Journal of
ClinicalPathology,84 (1985),620.
2. Coulter Electronics,Hematology Analyzer, 5 (1984), 1.
3. CoulterElectronics,Product
Reference
Manual;S-PlusIVwith3-part
differential
andautotransfer
(Hialeah,Coulter Electron- ics, 1983).4. Cox,
C.J., HABERMANN,
’I’.M., PAYNE,
B.A., KLEE,
G. G.and
PIERRE,
R.V.,
AmericanJournal of
ClinicalPathology,84 (1985),297.5.
GREENDYKE,
R.M., KANTER,
D.R., DEBOOVER,
L., SAVAGE, L.andVANGELDER,
S.,AmericanJournal of
ClinicalPathology,84
(1985),
348.Listof abbreviations
WBC
orwbc whitebloodcells;
RBCorrbc redbloodcells;
HborhborHGB
haemoglobin;Hct
orHCT
or hct haematocrit;PLT
orplt platelets; MCV
or mcvmean
corpuscolar volume;
mch meancorpuscolar haemoglobin;
mchc meancorpuscolar haemoglobin
concentration;RDW
or rdw red cell distribution width;MPV
or mpv meanplatelet volume; pdw platelet
distributionwidth; pct thrombocrit;Ly
orlylymphocytes; Mo
or mo orMono
monocytes;Gr
orgrgranulocytes.
FIRST INTERNATIONAL SYMPOSIUM ON DRUGS IN COMPETITIVE ATHELETICS To
be heldin Yugoslaviafrom
29May
to2June
1988Thesymposium
(which
issponsored
byIUPAC
andorganizedunder theauspicesof theIFCC)
will assemble scientists andphysiciansofvaryingbackgrounds
withacommoninterest intoxicologyandsportsmedicine.Thesymposium
willbeorganizedina manner so as tofacilitate discussion withexchange
of information and ideas.There will be
plenary
sessions tointroduceand reviewatopic,
followed by openforum discussions of various aspects of such topics.Plenarylectures include:
D. H.
Catlin(USA) Drugs
incompetitive athleticsD. A. Cowan (UK)
Specimenacquisition,
quality assuranceandchainofcustodyR. Dugal (CND) Analyticalrequirements
for detection ofdrugs
of abuse and anabolic steroidsH. W.
Durbeck(FR Germany) Gas chromatography-
mass spectrometryprocedures R. Hampl (CS)
Endocrineeffects and immunoassayprocedures
A. Ljungquist (S)
Health risks of steroiduseS. Rendi(:
(YU)
Identification ofdrugmetabolitesby gaschromatography-
mass spectrometryJ. R. Shipe (USA) Mass
spectrometryinstrumentation in the 1990sS. D.
Vesselinovitch(USA)
Syntheticsteroidsaspotential carcinogens.There will also beanopenforumon
drug
testingatmajorathleticevents.Principal
organizersofdrug
testingfor theOlympic Games
held inMontreal(1976);
Sarajevo(1984), Los Angeles (1984)
and Universiade ’87Zagreb
willparticipate aspanel
members in this openforum.Further
information from
eitherProfessor M.
Mikac-Devid, University Hospital’Dr
Mladen Stojanovi{’,Department of
Clinical Chemistry,41000
Zagreb,
Yugoslavia;orDr John Savory, Box
168,Department of Pathology,
Universityof
VirginiaMedical
Center,
Charlottesville, Virginia 22908,USA.
36