• Non ci sono risultati.

HeartMate XVE LVAD

N/A
N/A
Protected

Academic year: 2021

Condividi "HeartMate XVE LVAD"

Copied!
47
0
0

Testo completo

(1)

HeartMate XVE LVAD

VAD Resource Team

University of Washington Medical Center

(2)

VAD Resource Team

pinoy 598-6500

5SE Teddy Villamarin

esuver 598-4500

5NE Eileen Suver

janie 598-7369

5SE Janie Shively

kiwi 598-6500

5SE Nicola Kaye

kegreen 598-4500

5NE Kristina Green

sagreco 994-4535

5NE/5SE Sheryl Greco

sandrus 994-6986 Cardiac Surgery Shauna Andrus Email Pager/Phone

(3)

Heart Failure

Heart failure is defined as the inability of the heart to

effectively pump oxygenated blood to meet the metabolic demands of the body.

This devastating condition affects 4.7 million people a year in the United States alone.

When conventional medical therapies are unsuccessful, cardiac transplantation is an option for treatment and to prolong life.

Unfortunately, only 2300 patients each year receive heart transplants, because the number of patients awaiting

(4)

HeartMate Candidates

Transplant candidate:

dependent on vasopressor support

OR

incomplete response to continued vasopressor support

Approved for ‘Destination Therapy’

(5)
(6)
(7)
(8)
(9)

Pump

Percutaneous Tube

Vent Adapter

Weight: 3.74 lbs Pump volume: 83 ml Rate: Up to 120 Flow: Up to 10 L/min Titatanium

Motor with 2 bearings Vent port

Two tissue valves: Inflow & Outflow

(10)
(11)
(12)

Percutaneous Tube

&

Vent Adapter

AVOID:

(13)

System Controller

(14)
(15)

System Controller

‘Brains’ of system (computer) Alarms: Audible & Visual

Two cables: connect to power source ‘Quick connect’ to percutaneous lead

(16)

System Controller

Change Mode Alarm Reset

(17)

Power Base Unit (PBU)

& System Monitor

(18)

Power Base Unit

PBU Internal battery Battery charger •AC Fail •Low Batt •Alarm reset

(19)

Screen of System Monitor

Pump Rate: Rate PUMP is beating Flow: Cardiac output Stroke volume: ml/beat FUNCTION BUTTONS

(20)

FIXED vs. AUTO MODE

FIXED MODE is a set rate the pump will beat at regardless of other conditions. This rate can be adjusted by using the system monitor, and should

always be set at a rate close to the patient’s baseline.

AUTO MODE is responsive to the filling and emptying of the LVAD pump. The pump fills to a fixed volume at a variable rate depending on the patient’s activity and volume status.

(21)

Display Module

DISPLAY

•Provides information ONLY •No button interface

(22)

Components:Tethered

1. Quick Connect: percutaneous tube to system controller 2. Power Base Unit

Cable: connecting cable from PBU to system controller 3. Twist together

connection: system controller to power base unit cable

4. Twist together

connection: system controller to power base unit cable

(23)

Components: Battery Powered

1. Quick Connect

2. Twist together connect: system controller to battery 3. Twist together connect: system controller to battery

BATTERY •Requires 2 batteries •4-6 hours of power/set •Worn in holster

(24)

Hand Pump: Goes with patient EVERYWHERE

(25)
(26)
(27)

• Steady tone alarm

• Preceded and accompanied by a

Yellow Wrench alarm

• May be reset for 2 minutes

RED HEART: HAZARD ALARMS

• May be silenced for 24 hours

• 1 beep per second

(28)

YELLOW WRENCH

1. Power Cable Disconnect

MOST COMMON

Routine alarm that sounds when power source (battery or outlet) is changed or only one power source connected

(29)

YELLOW WRENCH

2. Controller Malfunction

2nd MOST COMMON

Caused by a malfunction in the system controller

*Inoperative Controller = ½ wrench*

(30)

YELLOW WRENCH

3. Power Limit Advisory

RARE

Occurs when pump has higher than normal pressure to pump against

Kinked or occluded vent line Kinked or occluded vent line ¼

¼ replace the extensionreplace the extension

High blood pressure to pump against High blood pressure to pump against ¼

(31)

YELLOW WRENCH

4. Rate Control Fault

RARE

Occurs when pump is changed to Fixed Mode and the preset rate is:

Less than 50 or

Greater than 120

Pump defaults to 72 beats per minutePump defaults to 72 beats per minute ¼

¼ correct by attaching the system correct by attaching the system

monitor and changing the backup rate

(32)

Steady tone alarm

Preceded and accompanied by a yellow wrench alarm

May be reset for 2 minutes

RED HEART: HAZARD ALARMS

May be silenced for 24 hours

1 beep per second

(33)

RED HEART

1. Low Beat Rate or No Op Low Beat Rate:

Pump slows to < 30 bpm No Op: may be stopped

THIS IS A CRITICAL ALARM

AND

(34)

Check connections

¼

most likely culprit is the percutaneous tube connection to the system controller

Check your patient

¼ you may need to hand pump

Check the vent port to insure it is not occluded

Try changing the power source

(35)

If pumping does not resume immediately, remove all power sources and start hand pumping

STAT page the perfusionist on call

STAT page Surg C Attending or Fellow

Call for pneumatic console (5NE) with Stroke Volume Limiter (SVL)

Do NOT restart the pump if it has been off > 5 minutes except with order from

(36)

RED HEART

2. Low Stroke Volume: < 25 ml 3. Low Flow: < 1.5 L/min

STAT page perfusion

STAT page Surg C Attending or Fellow

Consider changing to Fixed Mode to ↑ flow

Assess whether this is a patient or equipment problem

(37)

RED HEART CAUSES

Severe RV failure ¼ consider inotropes

Extreme low blood volume ¼ assess for bleeding or fluid loss, replace fluids

VT, VF or Asystole ¼ ACLS protocols

EXCEPTION: NO chest compressions

(may dislodge outflow conduit)

LV inflow cannula obstruction

(38)

XVE System Controller

disconnected from the percutaneous tube

FLASHING YELLOW BATTERY

RED HEART

YELLOW WRENCH

Steady tone

CANNOT BE SILENCED

Check system controller cable connection to percutaneous tube

(39)

BATTERY ALARMS

Flashing Yellow Battery

No tone

Yellow Battery

Once-per-second

Red Battery

(40)

FLASHING YELLOW BATTERY

No tone

Battery Module

The battery that controls the alarm lights and tones on the system controller is low

This battery controls ONLY the alarm tones and lights, NOT the pump

Change the battery, then do a controller self-test to clear the alarm

(41)

Low Voltage Advisory

Less than 15 min of battery power remain

YELLOW BATTERY

Once-per-second BEEP

(42)

RED BATTERY:

Steady tone

Pump defaults to POWER SAVER MODE: 50 bpm

9 Patient may be symptomatic @ this rate if used to faster rate

Immediately switch to an alternate power source or prepare to hand pump

Low Voltage Alarm

(43)

EMERGENCIES

VT/VF

Monitor BP; Flows; Cardiovert/Defibrillate

BLOOD FROM EXIT SITE

STAT page CT surgeon &/or fellow

RED HEART

Trouble shoot; be prepared to hand pump STAT page CT surgeon &/or fellow

(44)

IF LVAD MOTOR STOPS

< 5 minutes

Troubleshoot and get it working

again

Try hand pumping

Replace system controller

Try pneumatic console (with SVL)

> 5 minutes

DO NOT attempt to restart motor

STAT page perfusionist

STAT page Surg C Attending/Fellow

(45)

Pneumatic Console

Can drive XVE HeartMate if pump failure

Must use

STROKE VOLUME LIMITER Pneumatic tubing with flutter diaphragm

FIXED only available mode No electric connections to provide pump information Consider transfer to 5SE

(46)
(47)

University of Washington Medical Center

1997-2006

2 1 7 HM II 66 102* TOTAL 1 4 XVE HeartMate DT/HM II DT 1 2 Thoratec IVAD 5 24 Thoratec PVAD 1 23 26 XVE HeartMate 20 22 VE HeartMate 17 19 IP HeartMate Waiting/ Current TX Total Device

Riferimenti

Documenti correlati

•  When measuring particle spectra binned in energy, the energy dispersion originates the phenomenon of the event migration, or bin-to-bin migration—namely the fact that events

In the recent decade the initial employment rate of Chinese university graduates stays around... put university graduates in the first place in employment

(2002) compared local tumor control and complications of external beam radia- tion therapy with or without interstitial 137 Cs needle brachytherapy boost doses in 201 patients with

Class Ic antiarrhythmic agents (flecainide, encainide, moricizine) should be avoided in patients with coronary heart disease owing to the increased mortality shown in the

numero di transizioni nell’unità di tempo è una sequenza di tutti 1 (come NRZI). • Richiede un miglior rapporto S/N rispetto

© 1999 Pier Luca Montessoro ( si veda la nota a pagina 2) 2 Questo insieme di trasparenze (detto nel seguito slide) è protetto dalle leggi sul copyright e dalle disposizioni

Le stecche dei sedili sono unite a traverse laterali per mezzo di viti inossidabili a legno, le cui teste debbono essere bene affogate nello spessore del

[r]