Determination of automatic weather station self-heating originating from accompanying electronics.

N/A
N/A
Protected

Academic year: 2021

Condividi "Determination of automatic weather station self-heating originating from accompanying electronics."

Copied!
2
0
0

Testo completo

(1)

22 July 2021

AperTO - Archivio Istituzionale Open Access dell'Università di Torino

Original Citation:

Determination of automatic weather station self-heating originating from accompanying electronics.

Terms of use: Open Access

(Article begins on next page)

Anyone can freely access the full text of works made available as "Open Access". Works made available under a Creative Commons license can be used according to the terms and conditions of said license. Use of all other works requires consent of the right holder (author or publisher) if not exempted from copyright protection by the applicable law. Availability:

This is the author's manuscript

(2)

Topic T9.1 Meteorological traceability and uncertainty

Oral or Poster Presentation Oral Presentation

DETERMINATION OF AUTOMATIC WEATHER STATION SELF-

HEATINGORIGINATING FROM ACCOMPANYING ELECTRONICS

P. Pavlasek 1,2, A. Merlone 3, F. Sanna 3, G. Coppa 3, C. G.Izquierdo4, S. Ďuriš 2, Z. Ďurišová 2, J.

Palenčar 2 1

Slovak Institute of Metrology, Bratislava, Slovakia

2

Slovak University of Technology, Bratislava, Slovakia

3

Istituto Nazionale di RicercaMetrologica, Torino, Italy

4 Centro Español de Metrología, Madrid, Spain

E-mail (corresponding author): Pavlasek@smu.gov.sk

Atmospheric air temperature values are fundamental in meteorology and climate studies. To achieve high accuracy in the measurements, features, characteristics and performances of instruments are of high importance. This paper focuses on the most commonly used temperature sensors within automatic weather stations (AWS), with a specific focus on evaluating the self-heating effect. Self-self-heating in AWS originates not only from the temperature sensor itself, but also from the electrical components housed together within. This effect introduces extra heating in the system, causing biases and errors in temperature records. Conducted measurements show the temperature change in the close vicinity of the thermometers over a time period of more than 66 hours with an electric current and voltage supply values recommended by the respective sensor manufacturers. Furthermore, the temperature changes after increasing the voltage supply levels up to 80% of the maximum voltage recommended by the manufacturer is presented as well. The results of overall self-heating indicated a +0.07 °C increase in temperature for the tested sensors when using the manufacturers recommended electric current supply. However, the usage of elevated voltage levels shows a considerably higher temperature increase in the vicinity of the temperature sensors. In the presented study, the measured difference from the initial measured temperature can be high as +0.32 °C.

Acknowledgements

This work is being developed within the frame of the EMRP (European Metrology Research Programme) joint research project ENV58 “METEOMET”. Furthermore the work was supported by the Slovak R&D Agency, projects APVV-15- 0295, APVV-15-0164, the Scientific Grant Agency of the Ministry of Education of the Slovak Republic projects VEGA 2/0610/17, VEGA 1/0556/18 VEGA - 1/0098/18 and KEGA039STU-4/2017.

Riferimenti

Documenti correlati

Three linear resistive tensometers and three thermoelements were placed in the key areas of radiator (the inlet and outlet of cooling liquid to the heat exchanger and separator of

Here we report on the prevalence and fertility rate of hydatid cysts in wild boars that were shot during two hunting seasons (2016 –2017) in the Campania region of southern Italy..

Length of both surgical and VPAC times affected cell viability: the critical surgical time was set around 1 hour and 30 minutes, while cells preserved a good viability when kept

The biological effects of aqueous and organic solvent PM extracts on human epithelial lung A549 were evaluated, and the effects on cell proliferation and lactate

Direttore - Marco Bini - Sezione Architettura e Città - Gian Carlo Leoncilli Massi, Loris Macci, Piero Paoli, Alberto Baratelli, Giancarlo Bertolozzi, Andrea Del Bono, Paolo

Each measurement series started from the acquisition of the voltages and current with no defects in the reservoir; then, progressively one by one, 18 defects (glass spheres with 1.5

The present paper aims to analyze the issues of information quality and sharing in pediatric care pathways in order to promote shared decision-making and improve the knowledge of

Based on all relevant tables, we notice that when treatment is intensive, the customers in segment 2 “checking account only” show a relative high probability to switch towards