Investigation of the properties of large volume LaBr

3

:Ce scintillators with high energy gamma rays

N. Blasi2, F.Camera1, S. Brambilla2, C.Boiano2, A. Idini1, B.Million2, R. Nicolini1, L. Pellegri1, O. Wieland2, F. C. L. Crespi1, D. Santonocito3, C.Maiolino3, G.Bellia3,5, A.Rosa3, A.Owens4 and F.Quarati4

1Università di Milano and INFN sez. di Milano, Via Celoria 16, 20133 Milano, Italy 2 INFN sez. di Milano, Via Celoria 16, 20133 Milano, Italy

3 INFN Laboratori Nazionali del SUD, Catania, Italy

4 Advanced Studies and Technology Preparation Division (SCI-PA) ESA/ESTEC Keplerlaan 1, 2201AZ Noordwijk, the Netherlands 5 Dipartimento di Fisica e Astronomia Università di Catania, via Santa Sofia 64, 95123 Catania, Italy

The properties of cylindrical 3’’ x 3’’ and 3.5” x 8” LaBr3:Ce scintillators (see picture on the left) coupled to different Photonis and Hamamatsu Photomultiplier Tubes (PMTs) been investigated with γ-rays with energy up to 15 MeV. The PMTs used are R5300, XP5301, R6233, R6233-100, XP5700, XP3540 and R10233-100.

The tests have been done using the voltage dividers provided with the phototubes and, for the 3.5” x 8”

LaBr3:Ce detector, a specifically designed voltage divider coupled to a R10233-100 PMT. The new Voltage Divider provides both a pre-amplified signal for the energy measurement (from either anode or dynode) and direct anode signal for time measurement.

All the detector/PMT configurations have been tested in the energy range 100 keV - 15 MeV in terms of linearity, energy and time resolution for different values of the high voltage.

LaBr3:Ce 3” x 3” - 350 cm3 LaBr3:Ce 3.5” x 8” ≈ 1300 cm3

PMT Model # Dynodes Diameter Cath. Lum.

Sens. Cath. Blue

Sens. Gain - Typical HV - Typical

Photonis - XP5300 8 3 “ 14.4 µa/lmF 2.4 105 1000

Photonis – XP5301 - Clarity 8 3” 17.3 µa/lmF 1.8 105 1000

Photonis – XP5700 8 3” 13 µa/lmF 2.4 105 1000

Photonis – XP3540 10 5” 11.9 µa/lmF 6.5 105 1200

Hamamatsu – R6233 8 3” 138 µa/lm 2.7 105 1000

Hamamatsu – R6233-100S 8 3” 167 µa/lm 2.4 105 1000

Hamamatsu – R10233-100 8 3.5 “ 137 µa/lm 2.7 105 1000

Table-1

The table shows the properties of the tested PMTs as listed in the general reference sheets and in the final test sheets which is provided with every delivered tube.

As Photonis gives the Cathode Blue Sensitivity while Hamamatsu the Cathode Luminous Sensitivity, the two figures are listed in different columns.

The plots in the top panel show the energy spectra as it was measured using a large volume 3.5” x 8” LaBr3:Ce scintillator detector coupled with a R10233-100 Hamamtsu phototube. The first four panels show the energy spectra measured with sources: 133Ba, 137Cs, 60C and Am-Be-Ni.

The spectrum in the rightmost panel was measured in the reaction12C(p,p’)12C which produces a 15.1 MeV γ-ray from the direct decay of a 1- state in 12C*to the ground state. The measurement was done at the INFN SUD Laboratory (Catania). The calibration of all the spectra has been done using only the 137Cs and 60Co lines. The spectra with sources has been acquired using a specially developed Voltage Divider powered at 800 V. The 15.1 MeV γ-rays spectrum was measured using the standard E1198-27 VD and a voltage of 500 V.

The plot on the leftmost panel displays the amplitude (in millivolt) of the 661.6 MeV line measured at the anode of R10233-100 PMT for different value of the applied voltage. The two lines refer to a new developed voltage divider (black line) and to the Hamamatsu E1198-27 (red line). The plot in the central panel shows the trend of the energy resolution (FWHM) measured between 300 keV and 15.1 MeV using R10233-100 PMT with E1198-27 VD (blue points) and the new developed VD (red points).

The FWHM of the 15.1 MeV line is a little larger than expected but the data are still preliminary and the correction due to detector counting rate and PMT temperature have to be inserted. The plot on the right displays the measured linearity from 661.7 keV up to 9.89 MeV measured for the R10233 + E1198-27 Voltage Divider for different values of the applied voltages. As expected PMT linearity degrades as the voltage increases, a detailed study of linearity with the new developed VD is in progress.

200 250 300 350 400 450 500

0 200000 400000 600000 800000 1000000

Counts

γ-ray Energy [keV] 0 600 800 1000 1200 1400 1600

50000 100000 150000 200000

Counts

γ-ray Energy [keV]

3000 3500 4000 4500 5000

0 100 200 300 400 500 600 700 800 900 1000

Counts

γ-ray Energy [keV] 70000 8000 9000

30 60 90 120 150

Counts

γ-ray Energy [keV]

133Ba 137Cs + 60Co 12C

4.44 MeV peak

59Ni 8.98 MeV Peak 16 keV

21.5 keV

86 keV

130000 13500 14000 14500 15000 15500 500

1000 1500 2000 2500

Counts

γ-rays Energy [keV]

150 keV 12C(p,p’)12C 15.1 MeV Peak

400 600 800 1000 1200 1400 1600 0

5000 10000 15000 20000 25000

Counts

γ-rays Energy[keV]

3000 3500 4000 4500 5000 5500

0 10000 20000 30000 40000 50000 60000 70000

Counts

γ-rays Energy[keV]

50000 5500 6000 6500 7000

50 100 150 200

Counts

γ-rays Energy[keV] 12500 13000 13500 14000 14500 15000 155000 200

400 600 800 1000

Counts

γ-rays Energy[keV]

7% PMT non linearity

In the top plots the energy and time resolution measured using a 3”x3” LaBr3:Ce detector. The resolution values have been measured for different phototubes and for different high voltages. It is evident that the SBA photocathode of the R6233-100 gives a much better energy resolution and that the time resolution is affected by the high voltage which affects the risetime of the pulse.

PMT 4%

non linearity

The upper plots show the energy spectra as was measured using a large volume 3” x 3” LaBr3:Ce scintillator detector coupled with a XP5300 Photonis Phototube and a Voltage Divider VD202K/01 powered at 500 V. The first three panels of the plot show the energy spectra measured with sources: 137Cs, 60C and PuC. The spectrum in the right panel was measured in the reaction 12C(p,p’)12C which produces a 15.1 MeV γ-ray from the direct decay of a 1- state to the ground state. The measurement was done at the INFN SUD Laboratory (Catania). The calibration of all the spectra has been done using only the 137Cs and 60Co lines. The non linearity of the PMT is evident for high energy γ-rays.

The three plots on the right show i) the behavior of the energy resolution vs the energy of the incident γ-rays as was measured using a large volume 3” x 3” powered at 500 Volts. The LaBr3:Ce scintillator detector was coupled with a XP5300 Photonis Phototube and a Voltage Divider VD202K/01. As expected in a log-log scale the data show a linear trend, ii) The PMT energy linearity plots. Two different PMTs, Photonis model XP5301 and Hamamatsu model 6233-100 have been used. As expected the linearity degrades increasing the high voltage of the PMT. The Hamamatsu PMT shows a much larger non linearity as compared to the Photonis one (the two plots have the same Y scale).

In the top plots the energy and time resolution measured using a 3.5”x8” LaBr3:Ce detector. The resolution values have been measured for different phototubes and for different voltages. The 5” PMT gives the best energy performances. Time resolution is strongly affected by the applied voltage.

Conclusions:

The plots shown in this poster describe the response in terms of time, energy resolution and PMT linearity of large volume LaBr3:Ce crystals. Detector performances have been tested changing the PMT, Voltage Divider, the applied voltage and with gamma radiation ranging from 300 keV up to 15.1 MeV. The Hamamatsu PMTs with SBA photocathode in general provide better energy resolution but a much worser linearity.

Time resolution, which is critical for gamma spectroscopy applications, depends on the applied voltage. Using standard passive voltage dividers, there is a tradeoff between time resolution and linearity. An especially developed Voltage Divider capable to extract a preamplified signal from the anode/dinode and a time signal from the anode is under tests. The results of the tests shows that these detectros can be successfully used in arrays for gamma spectroscopy.

Preliminar

Preliminar

Counts Counts Counts Counts

Counts

Riferimenti

Documenti correlati

Except where otherwise noted content on this site is licensed under a Creative Commons 2.5 Italy License E - 184 with women’s franchise would only increase their

Aggiunge Angeloni (2013): «Simili iniziative andrebbero incoraggiate e replicate in tutto il territorio, perché queste aziende, con la loro storia, sono parte

Therefore, for the each day of the year one can establish precise hours when some volume of the energy should be delivered from the grid/energy system to cover the energy needs of

La cultura della difesa bastionata nel frattempo si è evoluta e la realizzazione delle nuove fortificazioni sarà affidata all’ingegnere militare tedesco don

Further the study categorizes consumers into three segments, namely, high, moderate, and low Spenders based on the preferred wine price for daily drinking, explores

The Monte Carlo simulations have shown that the energy deposit of a g-ray in large volume LaBr 3 :Ce crystal maintains spatial information independent of the energy of the

imaging approaches of colorectal liver metastases colorectal liver metastases (CRLMs) reporting the technical features, the clinical indications, the advantages and the

The distribution of the daily effective temperature over the period considered in this work is shown in Figure 2, the average being T 0.. eff =