AGATA

Demonstrator at LNL

Dino Bazzacco

INFN Padova

Symposium Italy- Japan 2012 on Nuclear Physics

Milano 20-23 November 2012

(2)

European g -ray detection systems

TESSA ESS30

EUROGAM

GASP EUROBALL III

EUROBALL IV

1980 1986 1992 1996

(3)

180 hexagonal crystals 3 shapes 60 triple-clusters all equal Inner radius (Ge) 23.5 cm Amount of germanium 362 kg Solid angle coverage 82 %

36-fold segmentation 6480 segments Singles rate ~50 kHz

Efficiency: 43% (M

g

=1) 28% (M

g

=30) Peak/Total: 58% (M

g

=1) 49% (M

g

=30)

AGATA

6660 high-resolution digital electronics channels Detectors operated in position-sensitive mode  Digital EDAQ + Pulse Shape Analysis + g-ray Tracking Coupling to ancillary detectors for added selectivity

Similar project is GRETA in the US

(4)

AGATA

Conventional array

Segmented detectors

g-ray tracking

Energy (keV)

Using the detectors in position-sensitive mode will provide very high efficiency and excellent energy resolution, making AGATA ideal for

spectroscopic studies of weak channels Effective energy resolution maintained at “extreme” v/c

v/c = 50%

(5)

Pulse Shape Analysis to decompose recorded waves Highly segmented

HPGe detectors

· ·

Identified interaction points

(x,y,z,E)

i Reconstruction of scattering sequence from

Compton vertices

Digital electronics to record and process segment

signals

1

2

3

4

Reconstructed gamma-rays

· · · ·

g

Gamma-ray tracking principle

(6)

Pulse Shape Analysis concept

B4 B5

B3

C4 C5

C3

CORE

A4 A5

A3

791 keV deposited in segment B4 measured

(7)

Pulse Shape Analysis concept

B4 B5

B3

C4 C5

C3

CORE

A4 A5

A3

C4

D4

E4 F4

A4 B4

x y

z = 46 mm

(10,10,46)

measured calculated

791 keV deposited in segment B4

(8)

Pulse Shape Analysis concept

B4 B5

B3

C4 C5

C3

CORE

A4 A5

A3

C4

D4

E4 F4

A4 B4

x y

z = 46 mm

(10,15,46)

measured calculated

791 keV deposited in segment B4

(9)

Pulse Shape Analysis concept

B4 B5

B3

C4 C5

C3

CORE

A4 A5

A3

C4

D4

E4 F4

A4 B4

x y

z = 46 mm

(10,20,46)

measured calculated

791 keV deposited in segment B4

(10)

Pulse Shape Analysis concept

B4 B5

B3

C4 C5

C3

CORE

A4 A5

A3

C4

D4

E4 F4

A4 B4

x y

z = 46 mm

(10,25,46)

measured calculated

791 keV deposited in segment B4

(11)

Pulse Shape Analysis concept

B4 B5

B3

C4 C5

C3

CORE

A4 A5

A3

C4

D4

E4 F4

A4 B4

x y

z = 46 mm

(10,30,46)

measured calculated

791 keV deposited in segment B4

(12)

Result of Grid Search Algorithm

Pulse Shape Analysis concept

B4 B5

B3

C4 C5

C3

CORE

A4 A5

A3

C4

D4

E4 F4

A4 B4

x y

z = 46 mm

(10,25,46)

measured calculated

791 keV deposited in segment B4

(13)

Examples of signal decomposition

A B C D E F CC

E g = 1172 keV net-charge in A1

E g = 1332 keV net-charge in C3, E1, E3

x10

x10

(14)

M

g

= 30

MC simulation of a high-multiplicity event Efficiency depends on position resolution

Reconstruction of 0.1÷10 MeV gammas

Position resolution (FWHM, mm)

(15)

Status after ~10 years of R&D

 Germanium detectors

 Pulse Shape Analysis

 Gamma-ray Tracking

 AGATA Demonstrator

 Electronics and DAQ

 Problems encountered

 Performance

 Experiments

 Evolution

(16)

AGATA Detectors

Volume ~370 cc Weight ~2 kg (shapes are volume-equalized to 1%)

AGATA Asymmetric Crystals Manufactured by Canberra France

AGATA Asymmetric Triple Cryostat Manufactured by CTT

80 mm

90 mm

6x6 segmented cathode

Energy resolution Core: 2.35 keV Segments: 2.10 keV (FWHM @ 1332 keV)

A. Wiens et al. NIM A 618 (2010) 223 D. Lersch et al. NIM A 640(2011) 133

Cold FET for all signals

(17)

AGATA Demonstrator

Objective of the R&D phase 2003-2009 Used for Physics at LNL in 2010-2011

5 asymmetric triple-clusters

15 36-fold segmented crystals 540 segments

555 high-resolution digital-channels Eff. 3 – 8 % @ Mg = 1

Eff. 2 – 4 % @ Mg = 30

Real time operation

Pulse Shape Analysis g-ray Tracking

Hosting sites

LNL  2009-2011 (done) GSI  2012-2013 (midway)

GANIL  2014-2015 (preparation)

(18)

AGATA Demonstrator + PRISMA

Infrastructure completed mid 2009, all clusters in place May 2011

Beam line

AGATA PRISMA

Reaction

chamber

(19)

AGATA: Structure of Electronics and DAQ

TRACKING Control,

Storage…

EVENT BUILDER

GL Trigger Clock 100 MHz T-Stamp

Other detectors Fast 1st Level Trigger

Detector Level

Other Detectors

GTS DIGITIZER

PREAMPL.

Global Level

DAQ-NARVAL

RUN- & SLOW-Control

High throughput pre-processing For one crystal:

2 ATCA carriers

1 GTS + 7 processing mezzanines

Other detectors:

interface to GTS, merge time-stamped data into event builder, prompt local trigger from digitisers Digital preamplifier concept

100MB/s/

detector

200MB/s/seg.

GTS

Processing Processing Processing

FARM PSA Core +

36 segments

LAN

(20)

6 crystals in one ATCA crate

(228 high resolution channels)

(21)

GTS : the system coordinator

All detectors operated on the same 100 MHz clock

Downwards 100 MHz clock + 48 bit Timestamp (updated every 16 clock cycles) Upwards trigger requests, consisting of address (8 bit) and timestamp (16 bit)

max request rate 10 MHz total, 1 MHz/detector

Downwards validations/rejections, consisting or request + event number (24 bit)

(22)

Some important issues

• Cross talk among the channels of one crystal

• Need to run at very high counting rates

• Neutron damage and energy resolution degradation of the segments

– PSA can be used to restore resolution, so that

– Annealing of crystals can be done less frequently

(23)

Sum of segment-energy vs number of hit segments

• Crosstalk is present in any segmented detector

• Creates strong energy shifts proportional to fold

• Tracking needs segment energies !

Crosstalk

Segment sum energies projected on fold

2folds : Core and Segment sum centroids vs hit pattern

…All possible 2-fold combinations

Energy [keV]

(24)

Cross talk correction: Results

(25)

High Count Rate Performance

• The detection efficiency of AGATA is, so far, provided by a small number of crystals.

• Experiments want to collect big statistics  need to run at high singles rates (> 50 kHz).

• Digital Signal Processing allows to work at rates “impossible” with analogue electronics.

• Under these “extreme” conditions, the performance of the detectors is still acceptable.

• A limit exists, due to pileup of the signals which exhausts the dynamical range of the FADC.

Can counteract this by reducing the gain of the preamplifiers, but then energy resolution is a bit worse also at low counting rate.

• However, running at high singles rates has consequences …

60Co - fixed

137Cs – 6 positions

Two Sources Method

0 1 2 3 4 5 6 7 8

0 50 100 150 200

Singles Rate (kHz)

FWHM @ 1.3 MeV (keV)

Shaping (trapezoid risetime /2.5) (us) Base Line Restorer (width/4) (us)

(26)

Neutron damage: shape of the 1332 keV line

Blue: April 2010  FWHM(core) ~2.3 keV FWHM(segments) ~2.0 keV Red: July 2010  FWHM(core) ~2.4 keV FWHM(segments) ~3 keV

Damage after 3 high counting-rate experiments (3 weeks of beam at 30-80 kHz singles)

(and a little help by EXOTIC... )

Worsening seen in most of the detectors; more severe on the forward crystals;

segments are the most affected, cores almost unchanged (as expected for n-type HPGe)

(27)

Crystal C002

The charge loss due to neutron damage is proportional to the path length to the electrodes.

The position is provided by the PSA (which is barely affected by the amplitude loss).

 corrected

Knowing the collection path, the charge trapping can be modeled and corrected away (Bart Bruyneel, IKP Köln)

CC r=15mm

SG r=15mm

April 2010

CC r=15mm

SG r=15mm

July 2010

The 1332 keV peak as a function of crystall depth (z) for interactions at r = 15 mm

(28)

Energy resolution at the end of the LNL campaign Before and After charge-trapping correction

Core

Sum of 15 spectra

Segments

Sum of 15*36=540 spectra FWHM 5.5 keV  2.9 keV FWTM 15.5 keV  6.8 keV FWHM 3.2 keV  2.8 keV FWTM 7.2 keV  5.6 keV

Undamaged detectors: FWHM ~2.5 keV, FWTM ~4.8 keV

E

g

(keV)

(29)

Performance of the Demonstrator

Just a few examples

(30)

Doppler correction using center of crystals

FWHM ~20 keV Detector FWHM = 2.2 keV

Doppler correction using center of hit segments FWHM = 7 keV

Doppler Correction Capability, “small” v/c

220 MeV

56

Fe 

197

Au (ATC1 + DANTE, July 2009)

Doppler correction using PSA (AGS) and tracking FWHM = 3.5 keV

(3.2 keV if only single hits)

Eg (keV)

v/c 8%

E(2+) = 846.8 keV

56Fe 2+  0+ 846.8 keV

56Fe 4+  2+ 1238.3 keV 4.8 keV FWHM Au recoils also seen by Dante

Original Corrected

(31)

Doppler Correction Capability, “large” v/c

inelastic scattering

17

O @ 20 MeV/u on

208

Pb

F O

N C

B Be

a Li

No Doppler Corr.

Crystal Centers Segment Centers PSA+Tracking

16 O

TKE (MeV)

dE (MeV)

F. Crespi, Milano

v/c 20%

Recoils detected in TRACE

(32)

gg capabilities

135 MeV

32

S 

110

Pd (6 AGATA crystals)

138Sm

6 gates on:

347keV,545keV,686 keV,775keV,552keV, 357keV

871 keV 22+ - 20+ The performance of AGATA using g-ray tracking is comparable to conventional arrays with a much larger number of crystals

(33)

Imaging of E g =1332 keV gamma rays

AGATA used as a big Compton Camera

F. Recchia, Padova Far Field Backprojection

Near Field Backprojection

All 9 detectors One detector

All 9 detectors One detector

2 0

1 1 1

cos m c

E E  

 

 

g g

Source at 51 cm  Dx ~Dy ~2 mm Dz ~2 cm

(34)

neutrons

protons

Neutron drip-line

Lifetime of the

6.792MeV state in 15O

n-rich nuclei

Lifetimes in n-rich Ni, Cu and Zn isotopes Lifetimes of the

n-rich Cr isotopes Lifetimes near the

island of inversion

The Experimental Campaign at LNL 2010-2011

Pygmy and GQR states

Neutron-rich nuclei populated by fission

N=51 nuclei

Lifetime of 136Te Neutron-rich nuclei in

the vicinity of 208Pb

Proton drip-line Molecular

structure of 21Ne

Coulex of 42Ca

Isospin Mixing in 80Zr

Octupole-deformed Ra and Th nuclei

Shape transition in 196Os

N=84 isotone

140Ba

n-rich Th and U

g.s. rotation in Dy, Er, Yb High-lying states

in 124Sn and 140Ce

20 experiments 148 days

Order-to-chaos transition in 174W

(35)

IKP University Köln TU Darmstadt

PLUNGER

AGATA-PRISMA

Differential RDDS measurements at LNL

72

Zn(-2p-2n)

76

Ge +

238

U 577 MeV

E

g

E’

g

bbef~10%

E’g

baft~8.5%

Eg

RDDS

measurement

G

=55º

Spectra at 50 kHz singles rate

(36)

Lifetime measurement of the 6.79 MeV state in 15 O

15O 6.79 MeV

τ?? 15N 7.3 MeV τ=0.65 fs

15N 8.3 MeV τ=2 fs

source(*) (*)

14N(2H,n)15O and 14N(2H,p)15N reactions @ 32 MeV (XTU LNL Tandem)

Direct lifetime measurement with 4 ATCs at backward angles (close to the beam-line)

C.Michelagnoli, Padova

(37)

15O 6.79 MeV

τ?? 15N 7.3 MeV τ=0.65 fs

15N 8.3 MeV τ=2 fs

source(*) (*)

14N(2H,n)15O and 14N(2H,p)15N reactions @ 32 MeV (XTU LNL Tandem)

Direct lifetime measurement with 4 ATCs at backward angles (close to the beam-line)

Detected gamma- rays can be sorted

in few degrees θ

“slices” with a continuous distribution

Lifetime measurement of the 6.79 MeV state in 15 O

C.Michelagnoli, Padova

(38)

The 6.79 MeV level in 15 O

15 O

Transfer +

CN reaction

Sub-fs lifetime?

Sensitivity to sub-fs lifetimes due to position resolution of the AGATA crystals

1.1 fs

4 fs

0 fs

(39)

Loss of selection rules on K with temperature

K–quantum number

K is robust at T=0 (K-isomers …)

Order-to-chaos in 174 W

High-Spin Fusion Evaporation

50Ti on 128Te @ 217 MeV, I ≥ 60

V. Vandone, S. Leoni, Milano HELENA: 27 detectors  5

clusters of BaF2 (3”×3”, exagonal) Total solid angle: 25% of 4π

Total efficiency: 16% @ 500keV 4 Triple Clusters 2 and 3 folds:

e2g=30%, e3g=10%

(Mg = 30)

Goal: populate 174W at the highest possible spins (≥60ћ), in order to make the statistical fluctuation analysis of the ridge-valley structures in the γ-γ matrices, to

estimate the number of low-K and high-K bands and their correlation

(40)

1879 686

1328

1314 13491000

1000 1100 1200 1300 1400 1500 1600 1700 1800 1900 0

250 500 750 1000 1250 1500

Counts

Energy [keV]

1879 keV 1000 keV

1314keV 1328keV 1349keV 965 keV

K = 8 K = 12

Gate on γ-isomer enhanced delayed bump

Gate on ENERGY : delayed γ

prompt γ

AGATA vs Helena time-spectrum

Gate on DELAYED time : Isomeric gammas

background 174

W selection

Helena: selection of high-K bands

(41)

Quasi-Continuum gg matrices

: Number of bands below 1 MeV

Is the K quantum number

conserverd up to E*< 1 MeV ?

E g 1 [keV]

400 1200

800 1200

400

Eg2 [keV]

Cut 45°

similarity of different cascades, test of the selection rules

Similar number Low-K and High-K Bands

As in the only studied case of 163Er

174

W

Previous studied case

of 163Er

<r>

ORDER

CHAOS

in 174W ?

(42)

LNL: 2010-2011 15 crystals (5TC) Total Eff. ~6%

GSI: 2012-2013

25 crystals (5DC+5TC) Total Eff.~10%

GANIL: 2014-2015 45 crystals (15 TC) Total Eff. ~15%

AGATA+VAMOS

“backward”

From the Demonstrator to AGATA 1 p

Plans for the next few years

Demonstrator + PRISMA

“backward”

AGATA + FRS

“forward”

(43)

Lund-York-Cologne CAlorimeter (LYCCA)

• 12 weeks of beam

• New FRS tracking detectors (>106 s-1 at S2, 105s-1 at S4)

• New LYCCA-0 particle identification and tracking system

• Higher SIS intensities and fast ramping 109(U) to 1010 (Xe, Kr) ions/spill

• IKP-Cologne Plunger (under construction)

AGATA at the GSI-FRS in-flight RIB

AGATA

GSI-FRS

First part of GSI campaign ended this morning (21/11/2012) Four experiments performed , using up to 19 crystals:

- Coulomb Excitation of n-rich Pb, Hg and Pt isotopes - Pygmy resonance excitation in 64Fe,

- Isomer Coulex in 52Fe

- Lifetimes in the heavy Zr-Mo region

+ M1 excitation in 85Br, 131In + studies of HE background HECTOR

AGATA

Double Cluster ADC

(44)

The AGATA Collaboration

Bulgaria: Sofia

Finland: Jyväskylä

France: GANIL, Grenoble, Lyon, Orsay, Saclay, Strasbourg Germany: Darmstadt, GSI, Köln, München

Italy: Firenze, LNL, Milano, Padova Poland: Krakow, Warsaw

Romania: Bucharest

Spain: Valencia, Madrid, Salamanca

Sweden: Göteborg, Lund, Stockholm, Uppsala Turkey: Ankara, Istanbul

UK: Brighton, Daresbury, Edinburgh, Liverpool, Manchester,

Paisley, Surrey, York

(45)
(46)

First implementations of the g-ray tracking array concept

AGATA Demonstrator GRETINA

15 crystals in 5 Triple Clusters

Commissioned in 2009 at LNL (with 3 TC) Experiments at LNL in 2010-2011

Now working at GSI with 20 crystals

28 crystals in 7 Quadruple Clusters

Engineering runs started early 2011 at LBNL Experiments at LBNL in 2011

Now working at at NSCL

Riferimenti

Documenti correlati

In conclusion, the present calculations suggest an overall improvement of the predicted properties of wurtzite III-V nitrides through the GGA compared to the LDA. In particu- lar,

Finally, we must ensure that people with emerging issues or ongoing mental health problems can benefit both from in-work support, and from a cultural shift to supporting

• Under EU legislation, particularly stringent measures — in addition to those required for other dangerous substances — must be taken by employers to prevent harm:

Out of 148 confirmed SNPs, 43.9% perfectly agreed with the RAD-seq data in all the rese- quenced samples, while 51.3% showed from one to three different genotypes ( S4 Table ).

La Lombardia risulta essere la regione dove la durata media delle recessioni è più ridotta (poco oltre 10 mesi, anche se le fasi espansive non sono molto lunghe) seguita da

The argu- ment takes inspiration from the ideas used in the proof of Proposition 3.1 in [15] and, even if it is essentially the proof of [4] in disguise, in the sense that the

The spectrometer is equipped with entrance and focal plane detectors that allow the complete identification and trajectory reconstruction of the reaction products by using the Time

This protocol, designed by Artyll – Società Lombarda di Ingegneria, has been created to reduce application time and, thereafter, costs, in order to make accessible behavioral