• Non ci sono risultati.

3 Practical Setup Steven Dymarkowski

N/A
N/A
Protected

Academic year: 2022

Condividi "3 Practical Setup Steven Dymarkowski"

Copied!
10
0
0

Testo completo

(1)

3 Practical Setup

Steven Dymarkowski and Pascal Hamaekers

S. Dymarkowski, MD, PhD

Department of Radiology, Gasthuisberg University Hospital, Herestraat 49, 3000 Leuven, Belgium

P. Hamaekers, RN

Radiological Technician, Department of Radiology, Gast- huisberg University Hospital, Herestraat 49, 3000 Leuven, Belgium

3.1

Introduction

There are many practical aspects that need to be taken into consideration when performing a car- diovascular MRI examination. Although seemingly trivial, successful image acquisition and interpreta- tion requires a working knowledge of adequate pa- tient preparation, ECG lead placement, breath-hold instructions, and contrast injection. This expertise should be accompanied by knowledge of safety is- sues and precautions in the MRI environment, re- lating to the use of strong magnetic fields, radio frequency (RF) energy, magnetic field gradients, and cryogenic liquids.

The purpose of this chapter is to provide MRI us- ers, in particular those who aim to perform cardio- vascular MRI studies in daily clinical practice, with a practical introduction to setting up and perform- ing an MR scan.

3.2

The Physical MRI Environment and Safety Issues

As with other imaging techniques, MRI has its risks, and precautions need to be taken to ensure that potentially hazardous situations do not occur.

Attraction of ferromagnetic objects due to the static magnetic field is the most important consideration, but the potential unwanted effects of RF-induced heating and peripheral nerve stimulation in the MR environment should also be considered, in particu- lar in patients with permanent pacemakers, auto- matic implantable cardiac defibrillators (AICDs), long lengths of temporary cardiac pacing wires, metallic stents, and prosthetic valves.

Monitoring physiological parameters within the MR scanners can also be problematic. For example, use of MRI-incompatible ECG electrodes may be associated with third-degree burns, and distortion of the ECG by the magnetic field may interfere with monitoring of ischemic changes during MR exami- nations. Specifically designed hardware to reduce this distortion is under development.

Thus, appropriate knowledge of the MR environ- ment is essential to eliminate the risks that can be associated with the specific physical properties of MRI.

3.2.1

Static Magnetic Field

MRI personnel are expected to be familiar with the dangers associated with the presence of ferromagnetic

CONTENTS

3.1 Introduction 51

3.2 The Physical MRI Environment and Safety Issues 51

3.2.1 Static Magnetic Field 51

3.2.2 Alternating Magnetic Field Gradients 52 3.2.3 Bioeffects of Radiofrequency Energy 52 3.2.4 Interaction with Medical Devices 54 3.2.4.1 Cardiac Pacemakers and

Automatic Implantable Cardiac Defibrillators 54 3.2.4.2 Temporary External Pacemaker Wires 55 3.2.4.3 Prosthetic Valves, Surgical Clips, and Osteosynthesis Materials 55 3.3 General Patient Preparation 56

3.3.1 Patient Positioning and Coil Placement 56 3.3.2 ECG Monitoring 56

3.3.3 Breath-Holding and Free-Breathing Scans 57 3.4 Specific Patient Preparation 57

3.4.1 Stress Testing 57

3.4.2 Pediatric and Neonatal Cardiac MRI 58 3.5 MRI and Pregnancy 59

3.6 Contraindications to MRI 60 References 60

(2)

inside the magnet or can seriously damage the mag- net itself. The force exerted on a large metal object of sufficient size induces enough kinetic energy to destroy and injure anything in its path.

Despite numerous safety warnings issued by the manufacturers and the medical profession, sad but true stories of objects being pulled into MR scanners circulate, with, in extreme cases, fatal outcomes.

Special consideration should be given to objects that are associated with patient management, for exam- ple oxygen cylinders, monitors, injection systems, and drip stands (Fig. 3.1).

3.2.2

Alternating Magnetic Field Gradients

Part of the MR-imaging process involves alterna- tions of the magnetic field state of the gradient magnets necessary for spatial encoding. The rate with which the change of magnetic field (B) occurs is dependent on the gradient power of the systems and is expressed as dB/dt. Gradient requirements are usually higher in high-application systems (e.g., for functional brain MRI and cardiovascular MRI), to enable sequence repetition time (TR) to be as short as possible. The United States Food and Drug Administration (FDA) has issued recommendations for dB/dt for each system to be at a level less than that required to produce peripheral nerve stimula- tion.

The switching of these magnetic gradients also produces high acoustic noise levels, which has been limited by the FDA to 140 dB. It is also recommended that all patients should be provided with hearing protection in the form of earplugs or headphones.

3.2.3

Bioeffects of Radiofrequency Energy

The RF energy from an imaging sequence is nonion- izing electromagnetic radiation in a high-frequency range. Research has shown that the majority of the RF energy transmitted for MR imaging is dissipated as heat in the patient’s tissues due to resistive losses.

Such temperature changes are not felt by the patient.

Nevertheless, MR imaging can potentially cause sig- nificant heating of body tissues, and the FDA recom- mends that the exposure to RF energy be limited.

The limiting measure is the specific absorption rate (SAR) and is expressed as joules of RF energy depos- ited per second per kilogram of body weight (i.e., watts per kilogram). The recommended SAR limit

Fig. 3.1. Typical example of MRI-compatible power injec- tion with dual injection system (contrast agent and saline), equipped with a nonferromagnetic (aluminum) infusion stand

BOGA_03-Dymarkowski,H.indd 52

BOGA_03-Dymarkowski,H.indd 52 18.10.2004 05:24:2918.10.2004 05:24:29

(3)

Fig. 3.3. Floor markings inside a 3-T suite mark the invisible boundary of the 10-gauss line (arrows). No additional monitor- ing unit or power injection, even MRI-compatible ones may be positioned within this line for safety purposes

Fig. 3.2. a Siting diagram for a 3-T unit. The 5-gauss line is marked in red. Note that the public restrooms in the waiting hall (ar- rows) are separated from the magnetic field by a thick reinforced wall. b Similar, but less optimal siting diagram of a 1.5-T unit. The 5-gauss line extends beyond the Faraday cage into the restrooms (arrow).

People with pacemakers are potentially in danger when using this facility

a

b

(4)

accumulation within the human body. Normally, when heat accumulates in the body, it is dissipated by conduction, convection, evaporation, and radiation.

Disease states (hypertension, diabetes) or medication (sedatives, vasodilators) can impair the normal ther- moregulatory responses to a heat challenge.

The heat interaction of pure biological phenom- ena is negligible. However, this is not true in the case of heating in association with devices inside the MR system. Thermal injuries, up to third-degree burns, have been reported with the use of ECG leads, pulse oximeters, and monitoring devices with wires made of conductive materials. Therefore, manufactur- ers’ guidelines should be strictly enforced, and if any doubt exists about safety of a certain device it should not be used.

If leads are used they should not be looped and should not be placed immediately on the individu- al’s bare skin (Fig. 3.4). If available, leads with fiber- optic signal transmission and special outer shield- ing should be used.

performed in patients with implanted cardiac pace- makers. It is generally assumed that presence of a pacemaker is a strict contraindication for MRI.

The effects of the MR environment on pacemak- ers are diverse and dependent on many factors, such as the strength of the external magnetic field, the kind of sequences used, and the body area that is to be imaged.

Practical experience has taught us that though many patients with pacemakers have been inadver- tently placed into MR systems in the past, hazard- ous or fatal outcomes have only occurred in a small number of patients. The following facts, however, summarize the many dangers of performing MRI with pacemaker patients and provide evidence that such practices should be strictly discouraged.

The metal casing of a pacemaker tends to align to the longitudinal axis of the external MR magnetic field. This gives rise to magnet-related translational attraction or torque effects. In different studies de- flection angles were measured from 9° to 90° for

Fig. 3.4a,b. Example of conductive loops inside an MRI unit. a An ECG cable with a loop forms a potential hazard, since this may induce currents and cause burns when the patient’s skin is in contact with this wire. b A patient must never cross arms or legs, since this also causes an inductive loop

a b

BOGA_03-Dymarkowski,H.indd 54

BOGA_03-Dymarkowski,H.indd 54 18.10.2004 05:24:2918.10.2004 05:24:29

(5)

long-bore and from 11° to 90° for short-bore MR systems. Such movement of the pacemaker box may be problematic for patients undergoing MRI, with possible discomfort and the potential for proximal lead fracture.

It is also known that static magnetic fields close the reed switches – tiny electrical relays – of many pacemakers. Reed switch closure has been studied for different orientations and positions relative to the main magnetic field and at different magnetic fields. If reed switches are oriented parallel to the magnetic field, they generally close at 1.0±0.2 mT and open at 0.7±0.2 mT. Different reed switch be- havior has been observed at different magnetic field strengths. In low magnetic fields (<50 mT), the reed switches were found to close; while, in higher mag- netic fields (>200 mT), the reed switches opened in 50% of all tested orientations. Thus, reed switch closure is not predictable with certainty in clinical situations. Despite this unpredictability of the reed switch in these experimental circumstances, tests on isolated pacemakers and in patients with AICDs using continuous monitoring of pacemaker output and temperature at the lead tip at low fields (0.5 T), have shown no pacemaker or AICD dysfunction during MR imaging, and no changes in the pro- grammed parameters in any device tested in vivo or in vitro. Furthermore, the only consequence of reed switches closing is that the pacemaker is switched into asynchronous mode, during which a predeter- mined fixed pacing rate takes over. This effect can be compared with the short period of pacemaker interrogation that takes place in routine pacemaker checkups.

However, it is not the above effects alone that are responsible for the adverse outcome in patients with pacemakers during MR examinations. More con- cerning is induction of currents in the pacemaker lead which could cause action potentials in and sub- sequent contraction of the ventricle. There have in- deed been reports in the literature of cardiac pacing occurring at the TR of the sequence.

Furthermore, temperature increase in the pace- maker lead is a severe and realistic problem. Studies have shown that pacemaker electrodes exposed to MR imaging can have temperature increases of up to 63°C after 90 s of scanning. The potential adverse effect of a thermal injury to the myocardium should thus be considered as the main caveat in conduction MRI examinations in these patients, especially if RF power is to be transmitted in the vicinity of the pace- maker and its leads (i.e., MR imaging of the neck, chest, and abdomen).

Despite reports in the literature that MR exami- nations can be performed in patients who are not pacemaker dependent and have the device switched off, it remains inadvisable to perform such exami- nations considering the potential hazards as stated above. If a patient is somehow accidentally exposed to the MR field, good clinical practice demands that the pacemaker function should be checked by a trained cardiologist. Relatives or other individuals with pacemakers accompanying patients should be asked to leave the MR unit.

3.2.4.2

Temporary External Pacemaker Wires

Often patients are referred early after cardiac sur- gery with temporary external pacemaker wires still in situ. Usually these are cut off close to the skin before discharge. This condition should not be con- sidered as a contraindication to MRI. In our experi- ence, the external portion of these wires should be contained in plastic, test tube-like receptacles, taped to the patient’s chest. These segments of the wires (the straight wire section) can cause significant sus- ceptibility artifacts on the anterior chest wall. This generally poses no concern for evaluation of left ventricular function for example, though the wires do interfere with image quality in the immediate region of the wire, e.g., if an evaluation of the right coronary artery is to be obtained.

3.2.4.3

Prosthetic Valves, Surgical Clips, and Osteosynthesis Materials

There are standard procedures for testing of pros- thetic valves for MRI safety. The term “MRI-safe”

is defined by the American Society for Testing and Materials (ASTM), Designation F 2052. Standard test

method for measurement of magnetically induced displacement force on passive implants in the mag- netic resonance environment. A device is considered

to be MRI-safe when this device poses no additional risk to the patient or other individuals in a static magnetic field, but may affect the quality of the di- agnostic information.

Prior to the approval of the ASTM, prosthetic valves are tested for translational attraction (aver- age deflection angle measured in magnetic fields up to 3 T), tissue heating in the immediate vicinity of the device, and artifacts.

Most currently used prosthetic valves have been

extensively tested in the past and it was found that

(6)

Surgical clips and osteosynthesis material are not considered to be contraindications for MRI exami- nations, with the sole exception of cerebral aneu- rysm clips.

However, in the light of the constant evolution in development of medical devices, the purpose of this chapter is not to provide an exhaustive overview of compatibility of every possible bioprosthetic one may encounter. Several good reference works and pocket guides have been written on this subject and it serves recommendation to keep a copy of such a book in the MRI suite, to be consulted in case of doubt.

3.3

General Patient Preparation

3.3.1

Patient Positioning and Coil Placement

Prior to any cardiac MRI, the patient should be briefly interviewed with regard to MR contrain- dications. We ascertain whether the patient has a pacemaker or any other implanted device, retained pacemaker wires, or other foreign material inside the body (in particular postsurgical subarachnoid hemorrhage clips). Female patients are best advised to remove any make-up, since small metallic resi- dues in cosmetics may cause artifacts in the images or be potentially hazardous.

In general, it is always recommended to remove jewellery and have the patient disrobe completely and wear a hospital gown (ideally without pockets!).

This avoids the accidental retention of materials in trouser or shirt pockets, as patients are not always completely aware of exactly which objects pose a po- tential risk.

For cardiac MRI, most systems have specific car- diac or torso coils. These are often multi-element phased-array coils, required for parallel imaging.

The number of elements in such dedicated coils

can give an indication as to whether coil placement was performed correctly (Fig. 3.6).

3.3.2

ECG Monitoring

As already stated in Chap. 1, accurate peak detection in the ECG is critically important for good quality MR scans. Most MR systems currently use advanced triggering modules based on vectorcardiography (VCG) to improve R-wave detection in the MR en- vironment. They use a three-dimensional (3D) and orthogonal lead system. The ECG electrodes should be attached in a cross-shaped or triangular pattern, depending on the number of electrodes of the sys- tem (Fig. 3.7). Prior to application of the electrodes, the skin should be cleaned with alcoholic pads or an abrasive gel to improve surface contact. Excess gel should be carefully removed, and if necessary the chest may need to be shaved. Nonmetallic, pregelled electrodes should be used for optimal result. Note that when the electrodes are removed from their

Fig. 3.5. Example of a 6-element phased-array cardiac sur- face coil. The markings on the coil surface facilitate adequate positioning of the coil elements with regards to the body

BOGA_03-Dymarkowski,H.indd 56

BOGA_03-Dymarkowski,H.indd 56 18.10.2004 05:24:3018.10.2004 05:24:30

(7)

sealed container for a long period, the gel pad tends to dry out and lose good signal transduction.

During positioning, the quality of the ECG trace should be checked, and a constant, high-amplitude R-wave and a low T-wave identified. If the tracing is unsatisfactory, restart and reposition the elec- trodes.

3.3.3

Breath-Holding and Free-Breathing Scans

Prior to the placement of the cardiac or torso coil and introduction into the magnet bore, the patient should be made properly aware of what is going to happen during the MR examination.

It is essential to explain the breath-hold com- mands to the patient and, if necessary, to briefly rehearse these instructions. Breath-holding can be performed during inspiration or expiration. If mea- surements are performed during expiration, repro- ducibility among the different views is higher and the chance of partial volume effects due to slice mis- registration is decreased; though patients generally get the impression they have less “air” and are likely to give up a breath-hold prematurely. Therefore some investigators prefer to do breath-hold studies in inspiration.

If any sequences using navigator gating are to be used, ask the patient not to move and if possible to breathe as regularly as possible, to avoid deep in- spirations, and to try to avoid falling asleep. Lastly provide the patient with headphones so breath-hold commands and other communications can be effi- ciently conveyed. The patient is now ready for image acquisition.

3.4

Specific Patient Preparation

3.4.1

Stress Testing

As demonstrated in the following chapters, cardiac MRI offers several clinically usable approaches for the assessment of ischemic heart disease. Using echocardiography, the analysis of wall-motion ab- normalities under pharmacological stress with in- travenous administration of dobutamine has been established as a sensitive method. It has been shown

Fig. 3.7. Typical setup of vectorcardiogram (VCG) system with four skin electrodes. (Image courtesy of Philips Medical Systems, Marketing Division, Best, The Netherlands)

Fig. 3.6. a Coronal scout image in a patient where the coil placement is too low on the body. The erroneous placement can be clearly noted, since the area of the higher signal intensities, i.e., higher coil sensitivity, is centered under the heart. b After reposi- tioning of the patient, the signal distribution in the scout image reveals better coil placement

a b

(8)

patient is more limited than in echocardiography.

Using low-dose protocols, side-effects are minimal, but at higher doses side-effects can occur in 0.25%

of patients, including myocardial infarction (0.07%), ventricular fibrillation (0.07%), and sustained ven- tricular tachycardia (0.1%). The MR unit should be equipped with facilities to allow a rapid extraction of the patient, such as a removable tabletop, to en- able rapid clinical assessment and treatment in case of adverse events. Resuscitation equipment and medication should be in the immediate vicinity to be administered by experienced staff in case of an emergency.

Physicians performing stress studies should be aware of the patient’s medical history and strictly observe the known contraindications. These are: se- vere arterial hypertension, unstable angina pectoris, aortic stenosis, severe cardiac arrhythmias, obstruc- tive cardiomyopathy, endocarditis, and myocardi- tis. Patients are asked not to take their beta-blockers and nitrates 24 h prior to the examination.

Before a stress study is initiated, standard ECG leads are applied, as in any other cardiac MR scans.

Many experienced users, however, prefer to place electrodes in such a way that permits the acquisi- tion of a standard 12-lead ECG before the MR scan;

a policy that can save a considerable amount of time in case of an emergency.

It is important to be aware that ST segment changes indicating ischemia are deemed not to be reliable in the static magnetic field, so the ECG during the scans can only be used for monitoring of the cardiac frequency. Further physiology moni- toring requires the use of MR-compatible devices.

Nonferromagnetic monitoring devices exist, which can be brought up to a distance of 1–2 m from the entrance of the magnet bore opening. Using such devices, blood pressure can be measured every min- ute and pulse oximetry continuously monitored (Fig. 3.8).

Furthermore, audiovisual contact with the pa- tient is maintained throughout the MR examina-

tion, to be aware of any symptoms that may occur during the dobutamine infusion. At the same time, the acquired images are immediately assessed to de- tect wall-motion abnormalities; the most sensitive sign of a positive stress test.

Indications for termination of a stress study are: a systolic blood pressure decrease greater than 20 mmHg below baseline; a systolic blood pres- sure decrease greater than 40 mmHg from a pre- vious level; a blood pressure increase greater than 240/120 mmHg; intractable symptoms and/or new or worsening wall-motion abnormalities in at least two adjacent left ventricular segments.

3.4.2

Pediatric and Neonatal Cardiac MRI

Given the inherent benefits of tissue contrast and resolution in MRI, cardiac studies are frequently used as a complementary technique in children and infants with congenital heart disease to elucidate complex cardiac anatomy. However, the require- ment of complete immobility and the ability to follow breathing instructions is often not obtain- able in children, due to either age or developmental stage. Therefore general anesthesia or prolonged sedation is often necessary to overcome these limi- tations and to complete these procedures success- fully.

In our own centers, general anesthesia is pre- ferred due to previous experiences with failed sedation or perceived medical risk. The greatest concern during prolonged sedation is respiratory

Fig. 3.8. Example of MR-compatible monitoring unit

BOGA_03-Dymarkowski,H.indd 58

BOGA_03-Dymarkowski,H.indd 58 18.10.2004 05:24:3018.10.2004 05:24:30

(9)

depression and hypoxemia. A study by Lamireau et al. has found an 89% incidence of oxygen desatura- tion in children who were sedated for endoscopic procedures. For less invasive procedures, a study performed by Malviya et al., performed in 922 chil- dren undergoing MRI or CT procedures, has found a 2.9% incidence of hypoxemia and failure rate of 7% in children who received sedation. In contrast, the procedure is successful in all of the children re- ceiving general anesthesia, with only one incident of laryngospasm (0.7%). Although general anes- thesia is a more expensive procedure than seda- tion, it is an essential aid in successful completion of MRI studies with a minimum of adverse events (Fig. 3.9).

3.5

MRI and Pregnancy

In 2002, the American College of Radiology Blue Ribbon Panel on MR Safety finalized its review of the MR Safe Practice Guidelines as detailed in the so-called White Paper on MR Safety. With regard to safety of MRI procedures in pregnancy, a distinc- tion was made between health care practitioners and patients.

Pregnant health care practitioners are allowed to work in and around the MR unit throughout the en- tire pregnancy. This includes the positioning of pa- tients, injecting contrast, and entering the MR scan room in response to an emergency. Pregnant MR

Fig. 3.9. a Patient installation for pedi- atric MR examination under general anesthesia. The child is mechanically ventilated by a nonferromagnetic gas- inhalator and covered with blankets to decrease body temperature loss.

b Physiological parameters are con- stantly monitored by use of the MR- compatible unit

a

b

(10)

has therapeutic consequences during the duration of the pregnancy. MR contrast agents should not be administered to pregnant patients, since there is in- sufficient data regarding the pharmacokinetics of MR contrast agents passing through the placenta.

3.6

Contraindications to MRI

Any patient with any of the following devices im- planted should, on a routine basis, not be imaged by means of MRI:

• Cardiac pacemaker

• Automatic implanted cardiac defibrillator

• Aneurysm clips

• Carotid artery vascular clamp

• Neurostimulator

• Insulin or infusion pump

• Implanted drug infusion device

• Bone growth/fusion stimulator

• Cochlear, otologic, or ear implant

Furthermore, any patients presenting with the fol- lowing medical conditions should also preferentially not be brought into the MRI device:

• Patients with severe claustrophobia in which medical sedation is contraindicated or unable to resolve anxiety sufficiently

• Patients with ocular foreign body (e.g., metal shavings)

• Patients with unstable angina or New York Heart Association functional class IV heart failure Finally, the following patient groups form no contra- indication for MRI but should not be administered intravenous contrast agents:

• Pregnant and lactating women

• Patients with hemoglobinopathies

• Patients with severe renal disease (creatinine clearance less than 20 ml/min)

Gimbel JR, Kanal E (2004) Can patients with implantable pacemakers safely undergo magnetic resonance imaging?

J Am Coll Cardiol 43:1325–1327

Greatbatch W, Miller V, Shellock FG (2002) Magnetic reso- nance safety testing of a newly-developed fiber-optic car- diac pacing lead. J Magn Reson Imaging 16:97–103 Kanal E (2002) American College of Radiology white paper

on MR safety. Am J Roentgenol 178:1335–1347

Kanal E (2004) Clinical utility of the American College of Radiology MR safe practice guidelines. J Magn Reson Imaging 19:2–5

Lamireau T, Dubreuil M, Daconceicao M (1998) Oxygen saturation during esophagogastroduodenoscopy in chil- dren: general anesthesia versus intravenous sedation. J Pediatr Gastroenterol Nutr 27:172–175

Luechinger R, Duru F, Zeijlemaker VA et al (2002) Pacemaker reed switch behavior in 0.5, 1.5, and 3.0 Tesla magnetic resonance imaging units: are reed switches always closed in strong magnetic fields? Pacing Clin Electrophysiol 25:1419–1423

Malviya S, Voepel-Lewis T, Eldevik OP et al (2000) Seda- tion and general anaesthesia in children undergoing MRI and CT: adverse events and outcomes. Br J Anaesth 84:743–748

Martin ET, Coman JA, Shellock FG et al (2004) Magnetic resonance imaging and cardiac pacemaker safety at 1.5- Tesla. J Am Coll Cardiol 43:1315–1324

Nahrendorf M, Hiller KH, Hu K, Zeijlemaker V et al (2004) Pacing in high field cardiac magnetic resonance imaging.

Pacing Clin Electrophysiol 27:671–674

Partain CL, Price RR (2004) MR safety. J Magn Reson Imag- ing 19:1

Pohost GM. (2001) Is CMR safe? (Editor‘s page) J Cardiovasc Magn Reson 3:9

Shellock FG, Kanal E (1991) Policies, guidelines, and recom- mendations for MR imaging safety and patient manage- ment. SMRI Safety Committee. J Magn Reson Imaging 1:97–101

Shellock FG, Tkach JA, Ruggieri PM et al (2003) Cardiac pacemakers, ICDs, and loop recorder: evaluation of trans- lational attraction using conventional („long-bore“) and

„short-bore“ 1.5- and 3.0-Tesla MR systems. J Cardiovasc Magn Reson 5:387–397

Shellock FG (2001) Pocket guide to MR procedures and metallic objects update 2001. Lippincott Williams and Wilkins, Philadelphia

Sommer T, Vahlhaus C, Lauck G et al (2000) MR imaging and cardiac pacemakers: in-vitro evaluation and in- vivo studies in 51 patients at 0.5 T. Radiology 215:869–

879

BOGA_03-Dymarkowski,H.indd 60

BOGA_03-Dymarkowski,H.indd 60 18.10.2004 05:24:3118.10.2004 05:24:31

Riferimenti

Documenti correlati

Analysis of light emission spectra of bacteria exposed to Bo, VHF and UHF magnetic field irradiation demonstrated more pronounced changes in the bioluminescence pattern as

The best sedimentation was noted in variant 1 of series 1, where magnetizers were used on activated sludge bioreactor and secondary settling tank.. The application of

At the territory of the Far East there are geological and economic potential to create raw material bases for development of important economic complexes such as agrochemical

1 (the real names of objects of the Unified gas supply system in this article are replaced by conditional serial numbers), the average number of pairs for all objects

Se i principi della conservazione fossero applicati a tutto il territorio si potrebbe ricostruire l’originario rapporto città-campagna attribuendo finalmente un

In modern Astrodynamics and Celestial Mechanics, thanks to the rise of interplanetary exploration that has brought a renewed interest in Lambert’s problem, it has application mainly

Figure 5.6: Numerical simulation of HBB-II display: flux density at the middle plane of the fluid when 2 opposite pistons operate.. acting both electrically, varying the value of

The adapter is the device which provide to enlarge the section of the linear support to make it perfectly fit inside the hole of the ceramic tile.. The linear support fit perfectly