Eur. J. Nucl. Med. 1, 117-122 (1976) Journal European N u c l e a r of

Medicine

© by Springer-Verlag 1976

Radioisotope X-Ray Fluorescence Technique in the Dynamic Study of Thyroid Behaviour under Iodine Load * **

Pietro Pavoni

Istituto di Patologia Medica II °, Universitfi di Roma, Via Bessarione 30, 1-00165 Roma (Italy)

Franco Vittorio Frazzoli and Alberto Magrini

Istituto di Fisica, Facolt~_ di Ingegneria, Universitfi di Roma, 1-00165 Roma (Italy)

Abstract. A presentation is made of equipment based on the principle of excitation of X fluorescence by means of radioisotope sources. The technique has been used to study the evolution of stable iodine con- centration (127I) in the human thyroid in vivo with a minimum variation in weight of ~50 parts per million.

The equipment is characterized by the use of 2 collimated sources of 241Am, each of 45 mCi, and by a proportional gas counter. The measuring times required are of approximately 100 sec each. The dose absorbed for each measurement is less than 30mrem at the level of the neck surface and less than 1 mrem at the thyroid.

Preliminary tests have been carried out over one month in an apparently normal subject given increas- ing doses (6-30 mg) of stable iodine.

Curves of the spatial distribution and of the tem- poral evolution of the counts due to the radiation of X k fluorescence emitted from the stable iodine atoms present in the thyroid were obtained in a pre- established constant area of the thyroid.

Since, under constant geometrical conditions, these counts are proportional to the concentration of iodine the pattern of the curve enables an assess- ment to be made of the increase of iodine in the thyroid; this increase dropped in correlation with the increase in the amount of iodine given, within the range of dose employed.

Further research will be carried out on more patients in order to establish the possibility offered by this technique in the field of thyroid pathophysiol-

o g y .

* Report presented at XIII International Annual Meeting of SNM, Copenhagen, Sef~tember 10-13, 1975

** The full paper has been submitted for publication in "Medical Physics" (USA)

Offprints requests should be sent to Prof. P. Pavoni

Introduction

The radioisotope X-ray fluorescence method [3], as is well known, is based on absorption through the photoelectric effect of X and gamma rays emitted by selected radioisotope sources. Following the deep ionisation thus produced, the atoms of the elements present in the sample exposed emit series of X-ray lines (K, L series etc.) of energies which are character- istic for each element, the intensities being correlated to the concentration of each element in the sample.

Using X-ray spectrometers, qualitative and quantita- tive analyses can thus be performed by measuring the energy and intensity of the characteristic lines detected.

Material and Methods

The experimental equipment consists of a measuring head con- nected to an electronic chain. The measuring head is mounted on a device which enables vertical regulation in order to centre the thyroid of the subject under examination.

The X-ray fluorescence of the stable iodine present in the thyroid is produced following exposure to two collimated sources of 241Am mounted symmetrically so that each one illuminates one lobe of the thyroid. (Fig. i)

The fluorescence KX-rays of the iodine for which the half thickness in soft tissue is about 2.0 cm, are detected by a propor- tional gas counter (gas filling Kripton), which is connected to an electronic chain.

The X-ray energy spectrum relative to a water solution of iodine (7 x 103 micrograms per gram) is obtained by means of a multichannel analyzer. The area of the "escape" peak, isolated and superimposed on a relatively low background, may be corre- lated to the concentration of iodine in water solution. It is thus possible to reach a minimum detectable concentration of 40 micro- grams per gram in 102 s of measurement.

The sensitivity of this technique thus enables the natural iodine content of the human thyroid to be detected. It is worthwhile pointing out that the fluorescence radiation energy of a few tens of KeV is strongly attenuated by the interposed superficial tissues;

(2)

[GAS FILLING'. K~..~.. 10% CH 4)

Am 24

STABLE IODINE - / l

THYRDID

I ///'Z

AI SHI ~ /

I Fig. 1. Schematic drawing o f the measuring

mm head

[detector ~ l

I I I channel II

analyzer I I AE 2 &E 1 E r

• scaler 1

l

timer

AEI: iodine KX escape peak and background A E 2 : background Fig. 2. Block diagram of the electronic chain

this does not occur with gamma rays emitted by radioactive tracers.

On the other hand the radiation emitted from the source gives rise to diffused radiation from the same tissue.

In the quantitative measurement subsequently performed the output signal from the amplifier is sent simultaneously to two single channel analyzers each connected to a scaler (Fig. 2). One o f these analyzers has the " w i n d o w " centered on the iodine escape peak, whilst the second has the " w i n d o w " centered on the back- ground. The difference between the counts accumulated during the preset measuring time, respectively in the iodine channel and the background channel, is proportional to the concentration o f iodine in the thyroid.

Fig. 3 shows the iodine escape peak relative to the thyroid of a normal subject; this spectrum is superimposed on the back- ground spectrum obtained in a p h a n t o m h u m a n thyroid not con- taining iodine. The difference in the two spectra is due to the

iodine content which, in a normal subject, is a few hundred micro- grams per gram [1].

In order to define the characteristics o f the experimental equip- ment a p h a n t o m thyroid was set up; the calibration curve was then obtained by correlating the difference between the counts accumulated in 103 s, respectively in the iodine channel and in the background channel, with the water iodine solution in the p h a n t o m (Fig. 4). The pattern o f the relative resolution power enables an evaluation to be made o f the minimum detectable con- centration o f iodine which results as approximatively 100 micro- grams per gram, in corrispondence to 100% relative resolution power.

F r o m a dosimetric point o f view the radioisotopic source of 241Am is particularly convenient as can be seen in Fig. 5 in which the p h o t o n flux needed to determine a l mrem soft tissue absorp- tion per second exposure is plotted against the energy.

(3)

P. Pavoni et al. : Radioisotope X-Ray Fluorescence Technique 119

J~ E/ u ul g o 7

200,

0 I~ 150.

¢)

¢ 100.

0

C

50.

t "escape" peak

- - - . ~ ~

-.. _ . . . - : ' ~ - . ~ •,',

• 4 . - r . ' • . . - : - • ~ . . - ~ r . ! . 0

p - % . - - . . ~ , ; .

. . - . - . - . ~ . ; - . . . .

..."' . ,.,,..,:-."~-"

, , . . . r . . , / , , " "

y

E (kev)

T H Y R O I D P H A N T O M

[]H20

[]Iodine in H20

I100

80

60

40

20

0",2

.00

¢)

= 0

h

Fig. 3. Iodine escape peak from human thyroid superimposed on background obtained from the

" p h a n t o m " n o t containing iodine

relative resoMng power

~ ' 0

i t I ~ ~ [ I I t , I I I I I I I I I I I I" I I I I I I I I I ~ I I I I h I IV

1000 2000 3000 4000

i o d i n e c o n t e n t i n w a t e r ( p p r n )

Fig. 4. Calibration and resolving power curves obtained from the

" p h a n t o m "

Bearing in mind the absorption of the superficial tissues, the dose in the thyroid is 1 mrem per 100 s o f exposure. This value is so low that several measurements can be carried out in a same subject without approaching the maximum permissible dose.

Results

The information obtainable with the described equip- m e n t is relative to the fraction of thyroid limited

by the penetration of radiations involved and the col- limation o f the sources.

With this information no absolute evaluation can be m a d e at present of the iodine concentration inas- m u c h as this is greatly influenced by the shape of the thyroid and the thickness of the superficial tissues of the neck. Nevertheless the same information allows us to follow the local evolution of the iodine con-

(4)

10 7

10 e

; ' ( 5 9 , 6 k e Y ) f r o m Am-241

/ I

I l

10 5 ' I J

i I I i

0 20 " 40 60 80 100 E (keY)

Fig, 5. P h o t o n flux necessary to determine I m r e m / s dose rate in soft tissue

The optimal position of the measuring head on the neck was previously established by determining the axial distribution of the iodine KX-ray counting rate on the neck of the subject under examination.

The above mentioned distribution in the two sub- jects are shown in Fig. 6. It can be seen that even the concentration of iodine normally present in the thyroid already enables a good localisation of the gland to be made.

Preliminary determination of axial distribution is necessary not only to improve the useful information/

background ratio but, above all, to ensure the best possible reproducibility.

Some preliminary tests were performed under stable iodine load in subject B (Fig. 6) who presented a plateau in the axial distribution of the counts which therefore ensures less critical measuring conditions.

Fig. 7 shows the uptake curve obtained after oral administration of 25 ml Lugol (containing ~ 6 mg of iodine). The points correspond to the mean of series of 3-5 measurements lasting 102 s each. This uptake pattern looks far different f r o m the typical pattern

o

2

50-

4 0 -

30-

20

apex subject A

I I I !

10 cm

base apex

subject B

base

I I I I I I h i _

- t 10 cm

Fig. 6. Axial distribution o f counting rate due to iodine in the thyroid o f subjects A a n d B

(5)

P. Pavoni et al. : Radioisotope X-Ray Fluorescence Technique 12

in 0 0

¢

0 0

C 4500-

4000 -

3500-

3000-

expected pattern on the cma[ogy

of radioiodine thymid uptake T

ore[ adrnini~trcttion of 025 m[ Lugo[to subject B /

/

i I

1 2

T

I I I

3 4 5

t i m e ( d a y s )

Fig. 7. Uptake curve in subject B corresponding to administration of 0; 25 ml Lugol (6 mg iodine)

5000-

4 5 0 0 -

4 0 0 0 -

3 5 0 0 -

3 0 0 0 -

/ / / / / 1 I I

/ / " T ~

! / t

I J t I f I

T I

t

6.2

t t f f

12.4 12.4 30.0 60.0

I I I / I I

0 5 10 15 20 25

t i m e ( d a y s )

i o d i n e m a s s a d m i n i s t e r e d ( r a g )

3'0 & P

Fig, 8. Uptake curve in subject B corresponding to periodic administration of increasing quantities of Lugol

(6)

0.1

\

\

I I I I 1 I f

1.0 1.1 1.2 1.3 1.4 1.5 C / C o

Fig. 9. Pattern o f m a x i m u m uptake (per m g o f iodine administered) versus the concentration o f iodine in the thyroid at the time o f administration

Table 1. Dosimetric data characteristic o f the technique p r o p o s e d c o m p a r e d with those f r o m radioactive tracer m e t h o d s

Tracer Radia- Typical Thyroid Whole

tion activity dose b o d y utilized adminis- (rein) dose

tered (mrem)

(gCi)

1127 (stable) Charac- 0 10 - 3 b 0

teristic X-rays a

1 lz5 (radioactive) y-rays 5 c 6 c 8-19 °

I T M (radioactive) y-rays 5 c 6-10 c 2-18 c

Tc 99 (radioactive) y-rays 250 c 2 - 6 . 1 0 - ~ 2 M . 1 0 - 3 c

a Excited by 24~Am (2 x 45 mCi).

b F o r each m e a s u r e m e n t lasting 100 s.

c D a t a obtained f r o m I C R P Report 17 (1971).

obtained with radioactive iodine tracers, because of a transitory phenomenon of accelerated discharge which takes place between the 6th and 24th hour.

thyroid.

Measurements were subsequently carried out in subject B submitted to periodic administration of in- creasing amounts of Lugol; the results (Fig. 8) clearly demonstrate that there is a reduction in the increases of maximum uptake, i.e. the concentration of iodine in the thyroid tends to a saturation level.

A quantitative analysis of the results was carried out bearing in mind the linear correlation between net counting rate and iodine concentration. It was thus possibile to calculate the pattern of the maximum increases of the iodine concentration (per mg iodine administered) in percent of the initial concentration, related to the thyroid uptake level before each new administration (Fig. 9).

Discussion

The results so far obtained are particularly encourag- ing and it seems worthwhile developing the technique described for use in the field of thyroid physiology.

Furthermore it should eventually be possible to draw up parameters useful for the recognition of func- tional disorders of the thyroid and for dynamic stud- ies on single nodules. In fact, it seems that the out- standing feature of this technique, as a diagnostic tool, will be the low dose administered as is clearly visible from Table 1 showing a comparison of data related [2] to radioactive tracers currently used in diagnostic techniques.

It is hoped that tests of thyroid function based on the technique described will be forthcoming in order to overcome the present limitations of "in vivo" studies, i.e. in children and in pregnancy.

References

1. Andreoli M. : La Tiroide Ed. I1 Pensiero Scientifico (1974) 2. " P r o t e c t i o n o f the patient in the radionuclide investigations"

I C R P Report 17 (1971)

3. Woldseth R. : X-ray Energy Spectrometry. K E V E X Corporation (1973)

Received October 7, 1975

Riferimenti

Documenti correlati

Novikov, Nonlinear equations of Korteweg-de Vries type, finite-band linear operators and Abelian varieties, Uspekhi Mat.. Druma, On analytic solution of the two-dimensional

The coherence, clearness and open-minded nature of Jervis’s thought about scientific and cultural innovations, which were main features of his intellectual research,

For positive integer n and i ∈ {0, 1}, let D i (n) be the number of derangements on n elements whose number of cycles has the same parity

We now turn to the general, two-dimensional case, and we ask again: Can two wave fields possessing different co- herence properties give the same optical intensity at

In this sample, we selected pLLSs and LLSs for which we could derive N H I reasonably well (specifically with a 1σ error less than 0.3 dex; see Section 3.2 ) and estimate

In the analysis as a reference point the largest intersection of streets located in the immediate vicinity of the plant was selected, from which virtual circles were drawn every

Così, a distanza di qualche anno, nel 2005, l’allora maggioranza di centrodestra giunse all’approvazione parlamentare di un testo di legge costituzionale recante

But the probabilistic approach to knowledge has also affected other areas of human inquiry and activity, especially the development of technology and consequently the rise of