15 The Technologist’s Perspective
Bernadette F Cronin
Introduction
From the technologist’s point of view, positron emis- sion tomography (PET) combines the interest derived from the 3D imaging of X-ray CT and MRI with the functional and physiological information of Nuclear Medicine. Until recently PET imaging was performed on “PET only” scanners and any direct comparison between the PET image and an anatomical one was either done by eye or required sophisticated hardware and software as well as extra human power to create a registered image. Now all the key manufacturers have developed and are marketing combined PET/CT scan- ners that allow patients to have both a PET scan and a CT scan without getting off the scanning couch. This creates an exciting imaging modality that offers the technologist the chance to develop a new range of skills.
Although clinical PET has been developing over the last 12 to 15 years, it is still relatively new and the high cost of introducing such a service has limited access for staff wishing to enter the field. As a result it is still the case that few staff entering the field will have previous experience. However, technologists with experience in other modalities will bring with them useful knowl- edge and, although PET in most cases is developing as part of nuclear medicine, recruitment does not need to be confined to nuclear medicine technologists alone. In the UK, training for PET technologists is performed on site with staff learning and gaining experience while working full time. Most recognized undergraduate and post-graduate courses cover PET in a limited capacity, providing theoretical information on a subject that is
intensely practical. As with all new techniques there will, for a while, be a discrepancy between the number of centres opening around the country and the avail- ability of trained staff. It is important for all concerned, and for the viability of PET itself, that training is for- malized to create high national standards which can be easily monitored and maintained.
A common view is to assume that a PET scanner can be sited in an existing nuclear medicine department and that, once commissioned, will function with little extra input. Although there are obvious similarities between PET and nuclear medicine, there are also quite subtle differences and it is not unusual for experienced nuclear medicine staff to find PET quite bewildering.
The advent of PET/CT adds an extra dimension as many Nuclear Medicine Technologists will not have any CT training or background. This too must be ad- dressed in a structured way to ensure that the tech- nique develops optimally. Currently, it is probably the least predictable imaging modality and staff working in a PET facility need to possess certain characteristics if the unit is to be successful. PET is multi-disciplinary, requiring input from various professional groups, and it is important that all these groups work together towards the same end. Teamwork is an essential com- ponent of a successful PET unit and without this, de- partments can easily flounder. More people are gaining the required expertise to work in these units; however, the numbers of people with the necessary expertise are still small and may not always be available to staff newly emerging departments. In the event that depart- ments are unable to recruit people with existing PET experience, it is advisable to send staff to comparable PET centers to obtain initial training.
295
Setting up a PET Service
When setting up a PET service, several factors have to be considered. First, and most important, what is the unit to be used for? If the only requirement is for clini- cal oncology work, then a stand-alone scanner with the tracer being supplied by a remote site may well be sufficient. For the technologist this option is in some ways the least attractive as it will always leave the unit vulnerable to problems over which it has no control, and will provide the least diverse workload. At the moment, most of the work will be done using fluorine- 18-labeled tracers (mainly 2-[
18F]-fluoro-2-deoxy-D- glucose ([
18F]-FDG)) and although the half-life is reasonably long (109.7 minutes) it does not provide much margin for error and therefore scheduling the patients can present added difficulties. If you are fortu- nate enough to be involved with a unit which has both scanner(s) and a cyclotron with chemistry facilities, the opportunities expand enormously. Scheduling, although not necessarily easier, can be more flexible, enabling better use of the equipment at your disposal.
In addition, the variety of scans that can be performed will increase allowing you to provide a full clinical on- cology, neurology, and cardiology service as well as the flexibility required for research work. Another factor to be considered is what service you will offer if you are purchasing a PET/CT scanner. All commercially available PET/CT scanners combine state-of-the-art PET scanners with diagnostic CT scanners. The specification of the CT component varies between manufacturers, but all are capable of high quality diag- nostic work. Opinions will vary as to whether the CT scanner should be used only for attenuation correction and image fusion or whether the patient should have both their PET scan and their diagnostic CT scan per- formed on the one machine during a single hospital/
clinic visit. The concept of “one-stop shops” has also been considered, where patients will have their PET, CT and radiotherapy planning all performed in a single visit.
Staffing Requirements and Training
Staffing levels will reflect the aims of the unit. If the in- tention is to have a stand-alone scanner with no on-site
cyclotron and chemistry facilities, then, assuming a throughput of ~800 patients per annum, the depart- ment will need a minimum of two technologists. Two are needed to ensure that there will always be cover for annual leave and sickness and also to share the radia- tion dose. This number of staff assumes that there will be some additional scientific and clerical support staff.
However, with a scanner plus cyclotron and chemistry facilities, the annual patient throughput is likely to in- crease due to greater availability of radiopharmaceuti- cals for clinical studies and, possibly, research work.
Assuming an annual workload of ~1200 patients, there would need to be at least three technologists with a corresponding increase in support staff. Provision of [
18F]-FDG for 8:00 am injection, allowing scanning to start at 9:00 am, requires cyclotron and chemistry staff to begin work very early. This has obvious revenue consequences as it will necessitate out-of-hours or en- hanced payment and may make recruitment difficult.
For this reason it is possible that in many centers the [
18F]-FDG will not be available for injection much before 9:30–10:00 am, with the first scan not starting until nearly 11:00 am. To maximize the resources avail- able it will probably be necessary to extend the working day into the early evening by working a split shift system. Once the PET technologists have been se- lected, consideration must be given to how they will be trained. There is little point sending a technologist who is going to be working in a clinical PET center to train in a center that undertakes research work only. The two units will operate in very different ways with different priorities, and are likely to be using a different range of tracers. To gain an in-depth working knowledge of clinical PET, technologists should spend at least four and preferably eight weeks at their “training” center.
The amount of training required will depend on the existing knowledge of the technologist. It must be re- membered that in order to function successfully in a PET unit, the technologist should be experienced in patient care, handling of unsealed sources, intravenous cannulation, and administration of radiopharmaceuti- cals as well as the acquisition and subsequent recon- struction of the data.
A good basic understanding of physics and comput-
ing will also help, particularly when it comes to
troubleshooting problems. CT training is also likely to
be an advantage as time goes on and there are several
short courses available that will provide a good
grounding in basic CT. If the work is likely to include
full diagnostic CT, there are staffing implications with
regards to state registration and the operation of CT
scanners which cannot be ignored.
Planning the PET Service
Clinical PET is divided into three main categories com- prising oncology, neurology, and cardiology. Most clin- ical PET centers have found that the workload is split among these three areas in a similar way, with oncol- ogy taking about 75 to 85% and the balance being shared between the other two.
When setting up a clinical PET service it is impor- tant to consider the types of scans that will be offered.
When deciding which protocols are going to be used, the most important thing to consider is what question is being asked of PET. This may not always be clear in the initial referral, and it is important that this is estab- lished prior to the study so that PET is not used in- appropriately. Other things to consider are whether quantification (for example, a semi-quantitative stan- dardized uptake value (SUV) [1, 2]) of the data will be required as this can only be performed on attenuation corrected studies. It is also important to know whether the patient is a new patient or is attending for a PET study as part of their follow-up. If they are for a follow- up scan it is essential to know what treatment they have had and, critically, when this treatment finished or was last given. Following the completion of chemotherapy, approximately 4–6 weeks should elapse before the patient is scanned, and after radiotherapy it is ideal to wait longer, up to 3–6 months, although clin- ically this may not be practical. If patients have their PET scan midway through a chemotherapy regime, the aim should always be to scan the patient as close to the next cycle of chemotherapy as possible, allowing at least 10 days to elapse since the previous cycle. Use should also be made of other diagnostics tests which the patient may have undergone, and so for the PET center it is valuable to have details about previous scans (CT, MRI, etc) and preferably to have the scan reports or images available before the appointment is made.
For PET, patient compliance must also be consid- ered. Patients are accustomed to the idea that diagnos- tic tests on the whole are becoming quicker. However, with PET, the scanning times are significantly longer than other scanning techniques. Scans can take any- thing from 20 minutes up to three hours for some of the more complex studies, and it is important to estab- lish whether the patient is going to be able to tolerate this. The design of PET systems, although not as confined as MRI, can mean that patients who are claus- trophobic may not be able to complete the scan
without help and may need to be given some mild sedation in order to undergo this type of investigation.
Recently developed PET/CT dual-modality systems will exacerbate this even more due to their increased axial length. Some patients, particularly younger subjects, will not be able to keep still for the length of time re- quired for the scan and may well need heavier sedation or perhaps a general anaesthetic.
The other key area that must be considered when setting up a PET service is which tracers are available.
In centers with a cyclotron on site, a full range of clini- cal radiopharmaceuticals should be available. However, when operating a PET scanner at a site remote from a cyclotron, there will be much greater limitation as to what tracers a re available (essentially,
18F-labeled tracers or those labeled with longer-lived radio- isotopes) and that in turn will limit the types of studies possible. The half-lives of these tracers can be very short, which, again, will influence the types of scan possible as well as the way the work is scheduled.
Production times for the various tracers range from just a few minutes up to two or three hours and so scans must be booked accordingly. As with all imaging units, it is important to ensure that the best and most efficient use of the equipment and personnel available is made. PET scans are costly both financially and in terms of the time required, and a single scanner may only be able to accommodate 7–10 scans per day. The way the schedule is arranged should ensure that the scanner is in continuous use throughout the day to maximize throughput. Currently in the UK there are few centers producing [
18F]-FDG for distribution, and those centers which do cannot produce it in vast quan- tities. It is essential that tracer is used in a way that ensures that little, if any, is wasted, and this requires a lot of thought on the part of the people booking the scans.
Information that should be given to the patient ob- viously includes the time, date, and location of where the examination will take place, along with a clear ex- planation of what the scan involves and any special dietary requirements including special instructions for diabetics. An information booklet may be of help but it may need to be tailored to specific scan types and the working practice of individual departments.
It is a good idea to ask the patient to provide
confirmation of their intention to attend for their
appointment. A patient not attending at the pre-
arranged time can cause a huge disruption and waste
of both tracer and scanner time, which is difficult to
fill at short notice.
Patient Preparation and Scanning Protocols
Different centers will have different methods for preparing and performing the tests, and these varia- tions will be determined by several factors, including the type of PET system available to them. However, re- gardless of the type of PET camera being used, there will be some similarities in technique with the greatest overlap in the area of patient preparation and manage- ment of the patient once they are in the department in order to obtain the best diagnostic data.
PET Scanning in Oncology
Today, [
18F]-FDG PET scanning is regarded as useful in many oncological conditions. The most frequent appli- cations are in staging of disease, the assessment and monitoring of treatment response, and in the evalua- tion of tumor recurrence, particularly when morpho- logical imaging techniques are equivocal or difficult to assess for technical reasons.
Patient preparation for [
18F]-FDG PET scans in on- cology is fairly simple. Patient referrals are divided into two categories, those who are insulin-dependant dia- betics (IDD) and all others. Insulin-dependant diabet- ics are asked to drink plenty of water in the six hours leading up to their scan appointment, but they are not
asked to fast because it is both inappropriate and un- necessary to disrupt their blood glucose levels. All other patients are asked to fast for at least six hours prior to their scan but are again encouraged to drink plenty of water. On arrival in the department, the patient’s personal details should be checked and a clear explanation of what the scan involves should be given.
It is important to give the patient the opportunity to ask questions at this stage so as to ensure that they are completely relaxed. For this reason, the appointment time should include a pre-injection period of about 15 minutes. For [
18F]-FDG PET scanning it is advanta- geous to have the patient lying down for both the injec- tion and the uptake period as this will hopefully aid relaxation and reduce unwanted muscle and brown fat uptake. In order to get the best out of the [
18F]-FDG PET scan it is important that certain key areas are given due consideration. The site of the body chosen for injection of the [
18F]-FDG is important. Any ex- travasation at the injection site is unsatisfactory. A small amount of tissued radiotracer at the injection site will result in a local radiation dose caused primar- ily by the positrons themselves due to their short path length and high linear energy transfer. In addition to this, the extravasated tracer will be cleared from the in- jection site via the lymphatic system and may result in uptake in more proximal lymph nodes that could be confused with uptake due to lymph node disease. Even small amounts of extravasated tracer can cause devas- tating artefacts on the 3D reconstruction of the data and, in extremis, can render the scan non-diagnostic if it is adjacent to an area of interest (see example in Fig. 15.1). Iterative reconstruction methods will help by
Figure 115.1. Reconstruction artifacts
caused by extravasated tracer at the
injection site can leave the final image
difficult or even impossible to interpret.
reducing these artefacts but may not remove them altogether.
Many of the patients attending for these scans will have already undergone chemotherapy regimes that may have made their veins extremely difficult to access.
It is therefore essential that the technologist giving the injections is well trained and well practiced in venepuncture. A good, precise injection technique is essential with the emphasis on accuracy. Unfortunately, there will occasionally be mistakes, but with careful thought it is possible to ensure that these mistakes do not make the scan completely worthless. It is always advisable to keep the injection site as far away as possi- ble from the area of interest. If the abdomen is under investigation it is best to avoid the ante-cubital fossae and likewise the hands or wrists when the pelvic area is significant. For many patients the arms may not be at all appropriate for the injection; for example, patients with a history of carcinoma of the breast, patients with bilateral axillary, neck, or supra-clavicular disease, and patients whose primary lesion is on their arm. In any of these instances it is advisable to inject the patient in their foot so that if there is a problem it will not affect the scan to such a degree that it may cause confusion. A note should always be made of the site of the injection so that the reporting doctor can refer to this if clarification is needed. A note should also be made if an unsuccessful attempt at venepuncture has been made prior to the successful one, as leakage of injected tracer can occasionally be seen at this initial site. If it is suspected that tracer has extravasated it is advisable to perform a quick test acquisition at that site to establish if this is the case, as it is not helpful to remove the in- jection site from the field of view without establishing whether it is necessary. If it is, then remedial action can
be taken; for example, lifting the arm above the patient’s head and out of the field of view. With PET/CT it is likely that patients will be routinely imaged with their arms raised above their head. This is done to prevent beam hardening artefacts in the CT image.
However, an image where the injection site ( the arm for example) is not seen but which does include axil- lary uptake will be difficult to assess unless the report- ing clinician can be sure that the injection was without incident. For this reason a quick acquisition of the injection site will avoid any subsequent confusion.
Normal physiological muscle and brown fat uptake of [
18F]-FDG can also give rise to confusion when re- porting PET scans (Fig. 15.2) [3, 19]. The causes of this increased uptake are not fully understood, however, it is clear that physical exertion and cold will inevitably give rise to increased levels of [
18F]-FDG in these areas.
In addition to this, stress and nervous tension also play their part.
Unfortunately it is not always manifestly clear which patients are going to produce high levels of muscle/brown fat uptake, and the calmest of patients can produce a very “tense” – looking scan. When uptake of this type occurs it is only the distribution that may give a clue as to the origins of the pattern of uptake, and it is almost impossible to distinguish this physiological response from any underlying pathologi- cal cause. In order to reduce this uptake it is suggested that patients be given 5–10 mg of diazepam orally one hour prior to the [
18F]-FDG injection. Clearly this is going to increase the time they spend in the depart- ment, as well as influencing their homeward journey, as they should not drive following this. It is not some- thing that can be done at short notice as patients will have to be brought in for their appointment an hour
a b
Figure 115.2. FDG uptake into tense muscle can mask
or simulate underlying pathology (a). The use of
diazepam prior to FDG injection can significantly
reduce this effect (b).
earlier than usual and an area will have to be provided where they can rest comfortably following the adminis- tration of the diazepam. Although the dosage sug- gested is relatively low, patients’ reactions range from barely negligible to sleeping throughout most of their visit to the department. The unpredictability of this re- action means that it is generally advisable to suggest that the patient bring someone along with them for their scan to ensure that they are accompanied once they leave the department. Too low a temperature may also increase the incidence of this physiological uptake so it is also important that, on arrival, patients are in- jected and allowed to rest in a warm comfortable area.
Although the likelihood of physiologically increased muscle/brown fat uptake is difficult to predict, there are certain categories of patients where diazepam should be given prophylactically. Obviously, any patient who has previously demonstrated physiological uptake with [
18F]-FDG should be given diazepam on all subse- quent visits. For patients attending for their first PET scan, those with lymphoma where the axillae, neck, and supraclavicular area is of particular interest may well be given diazepam, and patients with carcinoma of the breast are also likely to be considered. Adolescents also give rise to significantly increased amounts of physio- logical muscle/brown fat uptake and they too are routinely given diazepam, although in this group of pa- tients it is less successful than with adults. In addition to the diazepam, every effort should be made to ensure that the total environment and experience is as relax- ing as possible for the patient and that they feel com- fortable and well informed about the procedure. It is difficult to evaluate the true effect of the oral diazepam because for each patient there is no control study.
However, it does appear to reduce the incidence of this type of normal uptake and is therefore used frequently by many centers.
Uptake of [
18F]-FDG into normal myocardium is both unpredictable and difficult to manage. When investigating the thorax of a patient, particularly when looking for small-volume disease close to the myocar- dial wall, high levels of myocardial uptake can cause such severe artefacts that this can be almost impossible to assess (Fig. 15.3).
The use of iterative reconstruction methods will help decrease the magnitude of the artefact, but ideally it is desirable to minimize the myocardial uptake of the [
18F]-FDG when necessary. The physiology of uptake of [
18F]-FDG into the myocardium is complex and there are, as yet, no straightforward methods of reducing it.
One suggestion is that by increasing the free fatty acid levels in the patient the heart may be persuaded to utilize this for its primary energy source and ignore
the available glucose and [
18F]-FDG. One way of doing this is to ensure that the patients have a high caffeine intake prior to, and immediately after, the [
18F]-FDG injection. This can be achieved simply by encouraging the patient to drink unsweetened black coffee instead of water in the hours leading up to the scan and during the uptake period or, more palatably, a diet cola drink as this contains no sugar but high levels of caffeine.
Unlike glucose, [
18F]-FDG is excreted through the kidneys into the urine, and artefacts can be produced if the radioactivity concentration in the urine is high.
These artefacts can be seen around the renal pelvis and the bladder and can make it very difficult to assess these areas with confidence. In patients where the upper abdomen or pelvis is in question, it is advisable to give 20mg intravenous frusemide at the same time as their [
18F]-FDG injection. The frusemide should be injected slowly and the patient should be encouraged to drink plenty of water to ensure that they do not become dehydrated. Patients should be given an oppor- tunity to void immediately prior to the start of the [
18F]-FDG scan and if the pelvis is to be included in the scan, then the patient should be scanned from the bladder up. Obviously, it is not easy for patients to be given a diuretic and then have limited access to a toilet, but in the main, with a good explanation, patients tol- erate this quite well.
Finally, in patients where the laryngeal and neck area is being investigated, normal physiological muscle uptake into the laryngeal muscles can cause confusion.
Patients who speak before, during and after injection
are likely to demonstrate this type of uptake, whereas
patients who are silent for this period do not [4]. For
this reason it is recommended that when this area is of
Figure 115.3. High myocardial uptake in a scan looking for small-volume
disease in the chest can render the scan uninterpretable.
clinical significance, the patient should be asked to remain silent for about 20 minutes before injection, during injection and for the majority of the uptake period (the first half hour being the most important time after injection). Although this is not as easy as it sounds, most patients, if they understand why it is nec- essary, are sufficiently well motivated to comply.
PET Scanning in Neurology
For clinical PET in neurology, the most frequently used tracer is [
18F]-FDG. It features in the general work-up for patients who have intractable epilepsy for whom the next option would be surgery, and PET is one of the many investigations these patients will undergo to confirm the exact site of the epileptogenic focus. PET is also used in the investigation of primary brain tumors, mainly in patients who have had tumors previously treated by surgery, radiotherapy, or a combination of both. In these patients, standard anatomical techniques – for example, CT and MRI – are often difficult to in- terpret because of the anatomical disruption caused by the treatment. [
18F]-FDG PET scanning may also be helpful when investigating patients with dementia-like symptoms, as it can provide a differential diagnosis for the referring clinician.
When investigating partial epilepsy, [
11C]-flumazenil may also be used. This is a radiopharmaceutical that looks at the distribution of benzodiazepine receptors and it can be useful in more accurately localizing the epileptogenic focus [5] than [
18F]-FDG alone. Likewise, in the investigation of primary brain tumors, the use of [
18F]-FDG alone may not always be sufficient to answer the clinical question. Normal cortical brain has a high uptake of [
18F]-FDG and so makes visualization of tumor tissue over and above this background level quite difficult. The grade of the brain tumor is corre- lated with the degree of uptake into the tumor [6],with low-grade tumors having a lower than normal uptake, intermediate-grade tumors having a similar uptake to normal brain, and high-grade tumors having a higher than normal uptake. For this reason it is common for patients who are being assessed for recurrent tumor to have a [
11C]-L-Methionine scan in addition to their [
18F]-FDG scan. Methionine is an amino acid that demonstrates protein synthesis rates, so, whilst the [
18F]-FDG gives information about the grade of the tumor, the methionine will provide information about the extent of any tumor present. To use any
11C-labeled tracer obviously requires that the scanner be sited
close to the cyclotron and radiochemistry unit where the tracer will be produced, as the 20-minute half-life along with the low quantities currently produced pre- cludes transport to a distant site.
The exact protocol for brain scanning will vary from center to center but again there are certain key points that can be taken from one site to another. For all of these scans, patients may have to keep still for quite a significant period of time. When dealing with a patient suffering from epilepsy, particularly if they are very young, it may be that the risk of them having a seizure during the scan is quite high. To ensure that this causes the minimum amount of disruption to the total study, it is advisable to acquire the data as a series of sequen- tial dynamic time frames rather than as one long, single, static acquisition. This means that if the patient does move during the scan it is possible to exclude that frame from the total scan data. If the patient is clearly marked at various points corresponding to the posi- tioning laser lights, they can be repositioned so that only the affected frame is lost. However, any re- positioning needs to be done with extreme accuracy to ensure co-alignment of the transmission data for atten- uation correction, which is often recorded prior to the start of the emission scan. Patient movement is a par- ticularly important issue when dealing with the shorter-lived
11C tracers. If a patient moves during a single static acquisition, not only is the scan going to be corrupted, but the short half-life of the tracer will make it very difficult to repeat the scan as the resultant scan quality will be very poor. An alternative which is becoming increasingly popular on PET scanners with the ability to operate in 3D mode, is to acquire these as high-count, short-duration 3D scans only. Emission scan times in 3D for a brain may be of the order of only 5–10 minutes.
Preparation of patients for brain scans is again con-
cerned with ensuring that the injected [
18F]-FDG does
not have to compete with high levels of normal glucose
in the patient. So, as for oncology scans, the patients
are fasted and asked to drink water only in the few
hours preceding the study, but because of the high
avidity of uptake in the brain this fasting period is re-
stricted to three hours only. As with all other areas of
clinical PET, insulin-dependent diabetics should be
told to maintain their normal diet and insulin intake
and should not be fasted. Following injection it is im-
portant that the patient remains quiet and undisturbed
for the uptake period. External stimuli should be
avoided, as these may cause “activation” within differ-
ent areas of the brain, which, in extreme cases, might
be evident in the final image produced. Ideally, it is best
to ask the patient to remain in a darkened room with
their eyes closed and no distractions. This applies primarily to [
18F]-FDG scanning but should also be adhered to when using other tracers so that the prepa- ration is standard and reproducible. Many patients being investigated for epilepsy may well be young children, and may also have concurrent behavioural problems. As a result it can be difficult to keep these patients still for the length of time required for the scan, and it is prudent therefore to consider scanning them with some form of sedation. In general, it is preferable to scan under a full general anaesthetic rather than a light sedation. This is because the latter can be unpredictable and difficult to manage. However, a general anaesthetic requires good cooperation with the anaesthetics department and an anesthetist who has a reasonable understanding of PET and the time constraints associated with the technique. One option is to have a regular general anaesthetic time booked, the frequency of which can be governed by the individ- ual demands of the PET center. The anaesthetic will unfortunately result in a globally reduced [
18F]-FDG uptake, which will give a slightly substandard image, and the best results will be achieved if the patient is awake for their [
18F]-FDG uptake period but under anaesthetic for the scan. This is not always possible, particularly when patients are having a dual-study protocol with [
18F]-FDG scans in conjunction with an
11