• Non ci sono risultati.

P. Lupetti Dept. Life Sciences Siena University

N/A
N/A
Protected

Academic year: 2021

Condividi "P. Lupetti Dept. Life Sciences Siena University"

Copied!
19
0
0

Testo completo

(1)

P. Lupetti

Dept. Life Sciences

Siena University

(2)
(3)
(4)
(5)
(6)
(7)
(8)

The intensity of a small object (diameter much smaller than λ) as viewed through an optical

microscope, also called the point spread function (PSF). (a) The intensity of a small object as

observed in the plane (solid line). To distinguish two small objects from one another, the distance

between them should be approximately as plotted (see the distance between the solid and dashed blue lines). The total intensity is the sum (red line) that has been artificially raised a little. It is common to use the Rayleigh criteria for the resolution. With this criterion, the sum intensity of two close objects (red line) should have an intensity minimum that is 20–27% lower than the peak intensity. (b) The

same graphs shown along the optical axis z. Note that the resolution along the optical (z) axis is

worse than the resolution in the lateral (x,y) plane.

Two axes are shown for each graph. The upper one shows the distance for a numerical aperture (NA) of 1.0 and a wavelength of 500 nm. The lower axis is drawn for a general case and the resolution can be calculated with it for any given value of NA and λ.

(9)

The PSF as observed along the y axis for (a) a conventional

microscope and (b) a confocal microscope. The intensity is

shown along the z axis and along any axis in the xy plane (it has a circular symmetry in the plane).

This function is calculated for NA = 1.0 and λ = 500 nm. Note that in both methods the main spot is much larger along the z axis (optical axis) than in the lateral plane.

(10)
(11)
(12)
(13)

Marvin Minsky

(14)
(15)
(16)

Two-photon microscopy

ž 

Two-photon microscopy is a technique whereby two beams of lower intensity are directed to intersect at the focal point.

Two photons can excite a fluorophore if

they hit it at the same time, but alone they

do not have enough energy to excite any

molecules. The probability of two photons

hitting a fluorophore at nearly the exact

same time (less than 10

-16

) is very low,

but more likely at the focal point. This

creates a bright point of light in the

sample without the usual cone of light

above and below the focal plane, since

there are almost no excitations away from

the focal point.

(17)

Strengths

ž 

Less haze, better contrast than ordinary microscopes.

ž 

3-D capability.

ž 

Illuminates a small volume.

ž 

Excludes most of the light from the sample not in the focal plane.

ž 

Depth of field may be adjusted with pinhole size.

ž 

Has both reflected light and fluorescence modes.

ž 

Can image living cells and tissues.

ž 

Fluorescence microscopy can identify several different structures simultaneously.

ž 

Accommodates samples with thickness up to 100 µm.

ž 

Can use with two-photon microscopy.

ž 

Allows for optical sectioning (no artifacts from physical

sectioning) 0.5 - 1.5 µm.

(18)

Weakness

ž 

Images are scanned slowly (one complete image every 0.1-1 second).

ž 

Must raster scan sample, no complete image exists at any given time.

ž 

There is an inherent resolution limit because of

diffraction (based on numerical aperture, ~200 nm).

ž 

Sample should be relatively transparent for good signal.

ž 

High fluorescence concentrations can quench the fluorescent signal.

ž 

Fluorophores irreversibly photobleach.

ž 

Lasers are expensive.

ž 

Angle of incident light changes slightly, introducing

slight distortion.

(19)

Bibliography

ž 

P. Davidovits and M. D. Egger, Nature, 1973, 244, 366.

ž 

A. Hibbs, Confocal Microscopy for Biologists, Twayne Publishers, Boston (2004).

ž 

M. Minsky, Scanning, 1988, 10, 128.

ž 

J. Pawley, Handbook of Biological Confocal

Microscopy, Twayne Publishers, Boston (2006).

ž 

V. Prasad, D. Semwogerere, and E.R. Weeks, J.

Phys.: Condens. Matter, 2007, 19, 113102.

ž 

D. Semwogerere and E. R. Weeks, Encyclopedia of Biomaterials and Biomedical Engineering Confocal Microscopy, Taylor Francis (2005).

ž 

C. Sheppard, Confocal Laser Scanning Microscopy,

Twayne Publishers, Boston (1997).

Riferimenti

Documenti correlati

[r]

ROC curves showing accuracy in the prediction of a poor prognosis in comatose patients after CA, excluding patients who died from non-neurological causes, thus considering CPC 4 and

Historical situation: The restarting of Higher Education in China leads to a number of government sponsored programmes to go to study in France such as by the Ministry of Trade

Since the Voivod is obligated to make an appropriate referral to the Police, Internal Security Agency or other relevant authorities for information concerning the problem if

“Possiamo vedere nelle fake news un sintomo della crisi dell’universo cognitivo, delle credenze e dei valori condivisi di gruppi sociali, un mezzo volto a contrastare ciò che

De son côté, Susanne Altenberg, chef de l’unité du support au multilinguisme au Parlement européen, explique les spécificités du travail des interprètes au Parlement