• Non ci sono risultati.

Letter by foin et al regarding article, "edge effect from drug-eluting stents as assessed with serial intravascular ultrasound: A systematic

N/A
N/A
Protected

Academic year: 2021

Condividi "Letter by foin et al regarding article, "edge effect from drug-eluting stents as assessed with serial intravascular ultrasound: A systematic"

Copied!
2
0
0

Testo completo

(1)

e70

Article Type

Correspondence

To the Editor:

We read with great interest the article by Wakabayashi et al1 recently

published in Circulation: Cardiovascular Interventions. Wakabayashi et al reviewed the mechanisms of edge restenosis, bringing insights into the possible underlying causes of this phenomenon analyzing re-sults from several drug-eluting stent (DES) trials.

They suggest that the principal mechanism of edge restenosis in bare metal stent as in DES is through plaque shift and intimal hyper-plasia proliferation beyond the stent edges, within the first millimeter from the stent.

Interestingly, the authors noted from several trials, including pacli-taxel-, sirolimus-, zotarolimus-, and everolimus-eluting stents that de-spite significant reduction in restenosis with DES compared with bare metal stent control, stent edge restenosis was nearly equally frequent in both DES and bare metal stent and could not be resolved at the proxi-mal edge of the stents by drug elution from the stent.

The authors suggest that drug distribution downstream of the stent may produce a beneficial stent effect on the lumen distal from DES compared with bare metal stent. They propose this drug diffusion dis-tal to the edge of the stent as a possible explanation of the difference between proximal and distal restenosis, which also might explain the relatively high restenosis rate at the proximal edge of DES observed in trials.

Based on our experience with optical coherence tomography,2,3 we

strongly believe that another important phenomenon may explain the difference: incomplete stent apposition is a particularly frequent prob-lem, and incomplete stent apposition assessed by intravascular ultra-sound has been reported in up to one third of treated segments after DES implantation,4 affecting more frequently the proximal edge of the

stent compared with the distal edge of the stent.2,4,5 Stents deployed

across the origin of a side branch or covering long segments of tapering vessels (left anterior descending) are at particular risk, and postdilation with larger balloon is only effective if the balloon matches the proxi-mal vessel diameter and is also covering the proxiproxi-mal edge. If the stent is incompletely expanded with its struts not in contact with the vessel wall, drug will diffuse from the coating within the bloodstream, in-stead of diffusing to the intima and, therefore, prevent smooth muscle proliferation. Postprocedure malapposition may resolve by the rapid proliferation of neointimal tissue filling the gap between the stent strut and the vessel wall,6 which may thereby limit the risk of stent

throm-bosis but at the same time facilitate edge restenosis. If on the opposite, the stent struts are well apposed and embedded in the tissue (such as almost systematically observed at the distal edge of the stent), the drug can diffuse from the strut to the vessel and thereby prevent intimal hyperplasia.

The incidence of proximal edge incomplete stent apposition and its direct impact of edge restenosis should be assessed using high-resolution intravascular optical coherence tomography. Com-parison of stent segment postimplantation and at follow-up may identify the respective propensity of partially deployed stent, its role on restenosis, and whether complete stent apposition at the proximal edge can actually limit such edge effect.

Disclosures

None.

References

1. Wakabayashi K, Waksman R, Weissman NJ. Edge effect from drug-elut-ing stents as assessed with serial intravascular ultrasound: a systematic review. Circ Cardiovasc Interv. 2012;5:305–311.

2. Tyczynski P, Ferrante G, Moreno-Ambroj C, Kukreja N, Barlis P, Pieri E, De Silva R, Beatt K, Di Mario C. Simple versus complex approaches to treating coronary bifurcation lesions: direct assessment of stent strut apposition by optical coherence tomography. Rev Esp Cardiol. 2010;63: 904–914.

3. Viceconte N, Tyczynski P, Ferrante G, Foin N, Chan P, Alegrıa-Barrero E, Di Mario C. Immediate results of bifurcational stenting assessed with optical coherence tomography. Cathet Cardiovasc Interv. 2012. In press.

4. van der Hoeven BL, Liem SS, Dijkstra J, Bergheanu SC, Putter H, Antoni ML, Atsma DE, Bootsma M, Zeppenfeld K, Jukema JW, Schalij MJ. Stent malapposition after sirolimus-eluting and bare-metal stent im-plantation in patients with ST-segment elevation myocardial infarction: acute and 9-month intravascular ultrasound results of the MISSION! in-tervention study. JACC Cardiovasc Interv. 2008;1:192–201.

5. Cook S, Wenaweser P, Togni M, Billinger M, Morger C, Seiler C, Vogel R, Hess O, Meier B, Windecker S. Incomplete stent apposition and very late stent thrombosis after drug-eluting stent implantation. Circulation. 2007;115:2426–2434.

6. Guagliumi G, Bezerra HG, Sirbu V, Ikejima H, Musumeci G, Biondi-Zoccai G, Lortkipanidze N, Fiocca L, Capodanno D, Wang W, Tahara S, Vassileva A, Matiashvili A, Valsecchi O, Costa MA. Serial assessment of coronary artery response to paclitaxel-eluting stents using optical coher-ence tomography. Circ Cardiovasc Interv. 2012;5:30–38.

(Circ Cardiovasc Interv. 2012;5:e70.)

© 2012 American Heart Association, Inc.

Circ Cardiovasc Interv is available at http://circinterventions.ahajournals.org DOI: 10.1161/CIRCINTERVENTIONS.112.972927

Letter by Foin et al Regarding Article, “Edge

Effect From Drug-Eluting Stents as Assessed With

Serial Intravascular Ultrasound: A Systematic

Review”

Nicolas Foin, MSc, PhD Justin E. Davies, MD, PhD International Centre for Circulatory Health Imperial College London London, UK Carlo Di Mario, MD, PhD, FESC Biomedical Research Unit Royal Brompton & Harefield NHS Trust London, UK

by guest on March 7, 2015 http://circinterventions.ahajournals.org/

(2)

Nicolas Foin, Justin E. Davies and Carlo Di Mario

Assessed With Serial Intravascular Ultrasound: A Systematic Review''

Letter by Foin et al Regarding Article, ''Edge Effect From Drug-Eluting Stents as

Print ISSN: 1941-7640. Online ISSN: 1941-7632

Copyright © 2012 American Heart Association, Inc. All rights reserved.

Avenue, Dallas, TX 75231

is published by the American Heart Association, 7272 Greenville

Circulation: Cardiovascular Interventions

doi: 10.1161/CIRCINTERVENTIONS.112.972927

2012;5:e70

Circ Cardiovasc Interv.

http://circinterventions.ahajournals.org/content/5/5/e70

World Wide Web at:

The online version of this article, along with updated information and services, is located on the

http://circinterventions.ahajournals.org//subscriptions/

is online at:

Circulation: Cardiovascular Interventions

Information about subscribing to

Subscriptions:

http://www.lww.com/reprints

Information about reprints can be found online at:

Reprints:

document.

Answer

Permissions and Rights Question and

under Services. Further information about this process is available in the

permission is being requested is located, click Request Permissions in the middle column of the Web page

Clearance Center, not the Editorial Office. Once the online version of the published article for which

can be obtained via RightsLink, a service of the Copyright

Circulation: Cardiovascular Interventions

in

Requests for permissions to reproduce figures, tables, or portions of articles originally published

Permissions:

by guest on March 7, 2015 http://circinterventions.ahajournals.org/

Riferimenti

Documenti correlati

Key words: lexical relationships, probabilistic annotation, word sense disam- biguation, label normalization, semantic resource framework..

In [1] we proposed an innovative Word Sense In- duction method based on multilingual data; key to our approach was the idea that a multilingual context representation, where the

Cite this article as: Molicotti et al.: Molecular identification of Mycobacterium avium subspecies paratuberculosis in oral biopsies of Crohn’s disease patients. Gut Pathogens

The topic of this paper is the analysis of the flow through porous media contam- inated by some chemical species in presence of a growing biomass feeding on the pollutant.. In turn,

To compare the vitrectomy time, clinical outcomes, and complications between 27-gauge (27-G) and 25-gauge (25-G) vitrectomy in patients with primary rhegmatogenous retinal

We analysed a representative plot irrigated through a shallow well drilled in the sand dam aquifer, in terms of yield, Crop Water Productivity (CWP), Crop Water Productivity based

We further compared the detailed acoustic structure (including relative amplitude and vocal tract resonances) of the two main types of roars identified in our repertoire

[I1] Si, dove possiamo arrivare noi cerchiamo di sanarle, perché possono provare con bollette delle utenze, con eventuali contratti d’affitto, buste paga, dichiarazioni della