• Non ci sono risultati.

Two Ceramic Bearing Surfaces with a Self Adjusting cup:

N/A
N/A
Protected

Academic year: 2022

Condividi "Two Ceramic Bearing Surfaces with a Self Adjusting cup:"

Copied!
6
0
0

Testo completo

(1)

Two Ceramic Bearing Surfaces with a Self Adjusting cup: A New Application of Delta Ceramics to reduce the Risk of Dislocation and Subluxation

J.-Y. Lazennec, L. M. Jennings, J. Fisher and B. Masson

introduction

Dislocation remains one of the most common complications after total hip arthroplasty (THA). Its prevalence ranges from 0.6 to 27% in different series.

Dislocation is reported as a major cause of failure of ceramic-on-ceramic THA prostheses compared to a classical Metal PE bearing couple (0.51% versus 0.14%), Subluxation also appears as an important factor for hard on hard joint surface lesions.

Precise cup position appears to be a main factor as significant variations occur for frontal and sagittal acetabular tilt and anteversion according to sitting or standing positions. Double-mobility polyethylene hip prosthesis have been extensively used in France. Clinical results confirm the efficiency of such systems to prevent dislocation. Hovs/ever concerns remain with regards to polyethylene wear and osteolysis.

An innovative tripolar ceramic system has been investigated to solve these problems.

It has been suggested as a method to reduce the occurrence of recurrent hip dislocation and wear. The performance of delta ceramics from CeramTec has enabled the 3DA tripolar joint to be manufactured (Fig. 1). Using two bearing ceramic surfaces, the intermediate component acts as a "self adjusting cup", dealing with the variations of pelvic orientation and acetabulum anteversion.

Biomechanico! studies

The use of the 3DA tripolar joint seems an interesting alternative to face difficult or unexpected situations for cup adjustment and cases with hip instability.

(2)

The position of the center of rotation influences joint stability

Some publications explain that a tew millinneters inset of the rotation center increases the peak resisting nnonnent against dislocation. This benefit in terms of stability has a significant disadvantage due to the decrease of range of motion (ROM) with classical boll-insert systems. The 3DA tripolor joint allows the movement of the center of rotation much deeper inside the insert without a negative impact on the ROM.

Two biomechonical studies hove confirmed the improved resistance to dislocation with the 3 D A system. Experimental investigations hove been performed using in vivo data of Bergmann et o l at definite implant positions replicating close-to-reality conditions for T.H.A. orientation and loading conditions. A further measuring parameter for the dislocation stability was the torque during subluxation (resisting moment) against levering the head out of the cup.

Relevant combined movements related to posterior and anterior dislocation hove been tested (i.e. internal and external rotation of the leg with adjusted 90°

of flexion and 0° of abduction /adduction as well as with 10° extension and 15°

adduction).

The 3 D A tripolor joint revealed higher torques against subluxation in comparison to the classical AI-AI systems, even with 36mm head diameters, or 41 mm Metal on Metal bearing. More stable situations con be obtained even in poor implant positions, while the classical systems dislocated earlier and spontaneously without previous impingement. This was clearly demonstrated in cose of steep cup position or insufficient onteversion.

The "Self adaptation" of the intermediate cup has been demonstrated with computational models and experimental studies

• The additional outer-bearing surface motion creates a second "adjustable acetabulum" due to the eccentrotion between the rotation center of the ball head and the rotation center of the bipolar head.

• This offset creates a resultant force Fr that rotates the bipolar component. This phenomenon has been evaluated and validated on computational models.

Figure 2:

Illustration of the tripolor prosthesis at a non-equilibrium and equilibriunn position

(3)

Measuring the motion of the intermediate component is important for understanding the mechanism of the tripolar prosthesis and its efficiency against dislocation and microseparation The system was evaluated using a series of video-based motion analysis tests in two types of loading conditions, shear-out and lever-out situations. Shear-out was defined as the situation that leads to dislocation without impingement. Lever-out was defined as the situation that leads to a dislocation, accompanied with impingement. The study provides evidence that the relative motion of the intermediate component is closely related to the eccentricity between the intermediate component and the femoral head.

Mechanical performances

The mechanical characteristics of Biolox® Delta enable the manufacturing of this special device and especially of the intermediate cup with excellent strength properties. In collaboration with CeramTec AG a qualification program has been established to evaluate the mechanical reliability of this device. Standard qualification programs hove been performed on the 22,2 mm Ball Head and the standard XLW fix insert 32/41 mm.

Regarding the bipolar (intermediate piece) component, a new program has been set up, based on a ball head qualification program. Specifications of the bipolar component (diameter, roundness, clearance, etc.) are strictly the same as a 32mm ceramic ball head

The bipolar part shows a particularly high resistance to fracture.

Static Test Post Fatigue test

Load (kN) Average value

129 91

Required Value FDA

>46kN

>46kN

Load (kN) Minimum value 58

82

Min. Required Value FDA

>25kN

>20kN

All 3 components successfully passed the qualification.

A finite element analysis (FEA) has been performed for the complete device in worst case scenario and specially for the bipolar component. This study shows that the stress distribution is optimised by the two bearing surfaces and they have a positive impact on the ceramic strength.

Figure 3:

Finite Element Analysis (FEA) of the Bipolar component

(4)

Regarding the PE ring, dislocation tests have been performed to evaluate its resistance to secure the ball head inside of the intermediate component. Results are comparable to similar PE rings that have been used for more than 18 years for classical double-mobility hip joint. The same tests have been performed using the PE ring after 5 millions cycles with micro separation in hip simulator. Results demonstrate that the locking mechanism is still efficient and intact after 5 millions cycles vs/ith micro separation, even if this test is very challenging for the components.

Dislocation Test of the PE ring Nev^ PE ring before Hip Simulator Tests

PE ring after Hip Simulator Tests with Micro separation

Average Maximum Load (N) 151

175

Tribological tests

The aim was to assess the wear characteristics under standard test conditions and tests incorporating swing phase micro-separation between 200 and 500jjm.

Micro-separation is more appropriate for evaluation of ceramic bearings, as clinical wear rates, wear mechanism and wear debris are reproduced. The simulator was run for a total of 5 million cycles and the lubricant changed every 330,000 cycles. Wear of the ceramic components could not be detected gravimetrically. There was no visual macroscopic evidence of wear.

In a previous study, wear of conventional Biolox Delta components under microseparation conditions in the some simulator was measurable with reported wear rates of 0.32mmVmillion cycles during bedding-in (0-1 million cycles), reducing to a steady state wear rate of 0.12 mmVmillion cycles (1-5 million cycles). Furthermore, a stripe of wear was formed on the standard Biolox Delta heads, which increased the surface roughness Ra from <0.005 |jnn to between 0.02 |jm and 0.13 |jm. However, no stripe wear was observed in the testing of the 3DA tripolar joint.

The wear of the 3DA tripolar all ceramic hip was less than 0.01 mmV million cycles, the detection limit for wear measurement.There was no change in the surface roughness of the inserts. The 3DA tripolar joint showed reduced frictionol torque due to articulation at the smaller diameter 22mm inner femoral head. The wear volume of the PE rings could not be accurately quantified as it was within the systematic error of the soak control ring.

The design of the 3DA tripolar joint with the mobile ceramic head prevented edge loading of the head on the edge of the cup, so significantly reducing wear under these severe, but clinically relevant microseperotion conditions.

Conclusion

Dislocation and microseparation are major causes of failure for ceramic- ceramic hip prosthesis. The use of the 3DA tripolar joint seems an interesting alternative to optimize T.H.A function, as, in some cases, no ideal solution can be found for acetabular implantation. The "self adaptation" of the intermediate cup can be demonstrated: the additional outer-bearing surface motion creates a second "adjustable acetabulum".

(5)

The efficiency againsf dislocation and microseparation can be explained geometrically and experinnentally.

The design and testing carried out on the tripolar bearing with the mobile ceramic head show very high resistance to wear and stripe wear. Reducing the risk of dislocation and reducing wear drastically are two advantages that can place the 3DA tripolar joint as the best choice in primary Total Hip Arthroplasty.

Obviously this choice applies to recurrent dislocation also.

The use of the 3DA device could have a positive impact in terms of cost by significantly reducing the number of revisions.

References

1. An K. N. "Dislocation Stability of CeramConcept Tripolar Total Hip Arthroplasty"

Prosthesis "Division of Orthopaedic Research Mayo Clinic Rochester (USA)

2. Ali Khan M. A., Brakenbury P. H., Reynolds I. S. Dislocation following total hip replacement.

Journal of Bone and Joint Surgery-British Volume, 63-B, 1981: 214-218.

3. Bader R., Mittelmeier W. Dislocation Tests-Klinik fur orthopadie der Technischen Universitat Munchen - (Germany) "Evaluation of the range of motion and dislocation stability of implant systems for total hip replacement" Report # 030929.010.93

4. Chen Q., Lazennec J. Y., Guyen O., Kinbrum A, Berry D. J., An K. N. Validation of a Motion Analysis System for Measuring the Relative Motion of the Intermediate Component of a Tripolar Total Hip Arthroplasty Prosthesis Clinical Biomechanics 2005

S.Chen Q., Lazennec J. Y., Prabhakar P. P., Berry D. J., An K. N. Relative Motion of the Intermediate Component of an Eccentric Tripolar Total Hip Arthroplasty Prosthesis Division of Orthopaedic Research Mayo Clinic Rochester (USA)

6. Fackler C. D., Poss R. Dislocation in total hip arthroplasties. Clinical Orthopaedics and Related Research, 1980: 169-178.

7. Jennings L. M., Fisher J., Stewart T. D. "Wear Simulator Study of the Tripolar All Ceramic Hip Prosthesis" Mechanical Engineering Institute - University of Leeds (UK ) Report # 03/05-01 8. Jolles B. M., Zangger P., Leyvraz P. F. Factors predisposing to dislocation after primary total

hip arthroplasty: a multivariate analysis. Journal of Arthroplasty, 17, 2002: 282-288.

9.Kavanagh B. F., Fitzgerald R. H. Jr., Multiple revisions for failed total hip arthroplasty not associated with infection. Journal of Bone and Joint Surgery, 69, 1987: 1144-1149.

10. Lazennec J. Y., Chariot N., Gorin M., Roger B., Arafati N., Bissery A., Saillant G. Hip-spine relationship: a radio-anatomical study for optimization in acetabular cup positioning.

Surgical & Radiologic Anatomy. 25(7), 2003

11. Morrey B. F., Instability after total hip arthroplasty. Orthopedic Clinics of North America, 23,1992:237-248.

12.Sanchez-Sotelo J., Berry D. J., Epidemiology of instability after total hip replacement.

Orthopedic Clinics of North America, 32, 2001: 543-552.

13. Scifert C. F., Brown T. D., Pedersen D. R., Callaghan J. J., A finite element analysis of factors influencing total hip dislocation. Clinical Orthopaedics and Related Research, 1998:

152-162.

14. Stewart T. D., Tipper J. L., Insley G., Streicher R. M., Ingham E., Fisher J., "Long-Term Wear of Ceramic Matrix Composite Materials for Hip Prostheses Under Severe Swing Phase Microseparation". JBMR 66B (2003); 562-573.

15. Stewart T. D., Tipper J. L., Streicher R. M., Ingham E., Fisher J., "Long-Term Wear of HIPed Alumina on Alumina Bearings forTHR Under Microseparation Conditions". J MatsSci: Mats in Med 12 (2001); 1053-1056.

(6)

16.Toni A., Sudanese A., Ceramic on cerannic : long term clinical experience Bioceramics in joint arthroplasty Thieme Ed March 2001

17. Von Knoch M., Berry D. J., Harmsen W. S., Morrey B. F., Late dislocation otter total hip arthroplasty. Journal of Bone & Joint Surgery - American Volume. 84-A(l 1):1949-53, 2002 18. Woo R. Y., Morrey B. F., Dislocations after total hip arthroplasty. Journal of Bone and Joint

Surgery, 64, 1982: 1295-1306

Riferimenti

Documenti correlati

Three different synthetic approaches are reported starting either from 4-isobutylbenzoic acid (via benzyl alcohol, or via acyl chloride), or by Suzuki-Miyaura cross-coupling

After block of the creatine transporter with guanidine acetic acid (GPA) 10 mM, creatine, creatine-gluconate and creatine-ascorbate increased the tissue content of creatine,

Tenga presente che i primi tre valori riflettono la sua salute fisica (AF: Attività Fisica, RF:       limitazioni delle attività specifiche del suo Ruolo dovute a problemi Fisici,

At present, transversity has been extracted [ 2 – 6 ] in Semi Inclusive Deep Inelastic Scattering (SIDIS) processes in combination with the Collins fragmentation function (FF) [ 7

Quando una storia inizia i partner sono propensi ad innamorarsi, hanno bisogno di intimità, confidano nel fatto di essere amati e hanno desiderio sessuale, in questa fase,

  Il   libro   non   tradì   le   mie

flective view, we can see how vital force is consumed in the course of an activity, such as when you are busy thinking but, in turn draws nour- ishment from the activity that makes

Anche Sereni 18 , d’altra parte, compone una storia di assetti territoriali, cioè di paesaggi storici in quanto ricostruiti nella loro struttura funzionale; al contempo, però,