• Non ci sono risultati.

Piezoelectric Oxide Semiconductor Field Effect Transistor Touch Sensing Devices

N/A
N/A
Protected

Academic year: 2021

Condividi "Piezoelectric Oxide Semiconductor Field Effect Transistor Touch Sensing Devices"

Copied!
3
0
0

Testo completo

(1)

Piezoelectric oxide semiconductor field effect transistor touch

sensing devices

Ravinder S. Dahiya,1,a兲Giorgio Metta,1,b兲 Maurizio Valle,2Andrea Adami,3and Leandro Lorenzelli3

1

RBCS, Italian Institute of Technology, Genoa 16163, Italy 2

DIBE, University of Genoa, 16145, Italy 3

Bio-MEMS Group, Fondazione Bruno Kesseler, Povo, Trento 38050, Italy

共Received 26 March 2009; accepted 26 June 2009; published online 21 July 2009兲

This work presents piezoelectric oxide semiconductor field effect transistor共POSFET兲 based touch sensing devices. These devices are fabricated by spin coating thin共⬃2.5 ␮m兲 piezoelectric polymer film directly on to the gate area of metal oxide semiconductor共MOS兲 transistor. The polymer film is processed in situ and challenging issues such as in situ poling of piezoelectric polymer film, without damaging or altering the characteristics of underlying MOS devices, are successfully dealt with. The POSFET device represents an integral “sensotronic” unit comprising of transducer and the transistor—thereby sensing as well as conditioning 共and processing兲 the touch signal at “same site.” © 2009 American Institute of Physics.关DOI:10.1063/1.3184579兴

Touch sensing plays an important role in application do-mains such as robotics, electrotextiles, and medical prosthe-sis as it facilitates the detection and measurement of interac-tion parameters of real world objects. Over the years, touch sensing technology has improved and many new touch sen-sors, using different materials and transduction methods, namely, resistive/piezoresistive, capacitive, optical, mag-netic, piezoelectric, etc., have been developed.1 Most of

these sensors are big in size and hence unsuitable for appli-cation areas such as robotic fingertips, where large number of sensors are needed in a limited space. Miniaturized touch sensors using microelectromechanical system approach have also been developed.2 However, inherent fragile nature of such sensors limits their usage to applications involving small contact forces only. Mechanically flexible touch sen-sors using organic field effect transistor共FET兲 have also been

a兲Electronic mail: ravinder.dahiya@iit.it.

b兲Also at: DIST, University of Genoa, 16145, Italy.

FORCE PVDF-TrFE Metal Gate Insulator Source Drain Lower metal over Gate 1 mm (a) (b) Source Drain Top Metal over polymer Lower metal

(i) n-MOS Device

(ii) Polymer film Deposition

(iii) POSFET Device PVDF-TrFE G n+ n+ Si -Sub Oxide Layer Top Metallization PVDF-TrFE G n+ n+ Si -Sub Oxide Layer G n+ n+ Si -Sub Oxide Layer Gate Metallization Top Metal (Au) Lower Metal/Gate (Al) PVDF-TrFE Film (c) (d) (e)

FIG. 1. 共Color online兲 POSFET touch sensing device implementation. 共a兲 Structure and working of POSFET touch sensing device. 共b兲 SEM image of POSFET cross section.共c兲 POSFET before polymer deposition. 共d兲 SEM image of POSFET after polymer deposition. 共e兲 Fabrication steps of POSFET touch sensing device.

APPLIED PHYSICS LETTERS 95, 034105共2009兲

0003-6951/2009/95共3兲/034105/3/$25.00 95, 034105-1 © 2009 American Institute of Physics

(2)

developed for large area skin type of applications.3However, best organics have a mobility of about 1 cm2/V s versus 85 cm2/V s for silicon based MOS devices,4 due to which their usage is limited to detecting and measuring slow vary-ing contact forces only.

This work presents POSFET touch sensing devices, pri-marily developed for the fingertips of robotic hand. These devices are fabricated by spin coating thin共⬃2.5 ␮m兲 poly-vinylidene fluoride trifluoroethylene关P共VDF-TrFE兲兴 film di-rectly on the gate area of MOS transistor. The structure of POSFET device, shown in Figs. 1共a兲–1共d兲, is similar to the metal-ferroelectric-metal-insulator-semiconductor type of ferroelectric random access memory 共FeRAM兲.5 However, the thickness of ferroelectric film in FeRAM is usually much lower 共100–200 nm兲 than that used in POSFET device. The fixed charges ⫾Q, shown in Fig. 1共a兲, appear due to the remanent polarization 共Pr兲 of the piezoelectric polymer film and the charge neutrality condition. The charge carriers are thus accumulated at the surface of the semiconductor accord-ing to the polarization direction. For piezoelectric polymers in thickness mode, as in this work, the mechanical stress 共T3兲, electric field 共E3兲, and electric displacement 共D3兲 are related as6

D3= d33T3+⑀33E3, 共1兲

where d33and⑀33 are, respectively, the piezoelectric and di-electric constants of piezodi-electric polymer. Equation 共1兲 shows that the electric displacement/charge and hence the polarization can be controlled by the electric field 共E3兲 and the applied force共F兲 or stress 共T3兲. While former is used in FeRAM to switch the polarization state, the later is used in POSFET devices to modulate the charge in induced channel. Thus, the force variation is directly reflected into channel current of POSFET devices, which can be further processed by an electronic circuitry that may be present on the same chip. In this sense, POSFET device can also be compared with the mechanoreceptors in human skin—that not only sense the contact parameters, but also partially process the tactile data.7

In contrast with the conventional and extended gate approaches,8,9 where transducers and conditioning electron-ics are separate entities connected through wires, the POSFET touch sensing devices allow sensing and processing at same site which will improve signal to noise ratio and the overall performance. A conformal electronic surface can be obtained by embedding such devices, as mechanically dis-tinct and stiff subcircuit islands, in a soft and compliant

poly-mer such as polydimethylsiloxane 共PDMS兲 and interconnect the islands with flexible and stretchable metallization.10 Alternately, they can be developed on flexible electronic devices.11

The fabrication steps for implementing POSFET touch sensing device are shown in Fig.1共e兲. In first step, the MOS device is obtained using the n-MOS technological module of a nonstandard complimentary MOS 共CMOS兲 technology, based on 4 ␮m p-well ion sensitive FET/CMOS process.12 A Si3N4/SiO2 double layer is used as a gate dielectric. To match the tactile spatial acuity of human skin, POSFET de-vices are designed to have size of 1⫻1 mm2—fairly big when compared to the MOS devices obtained with present day technology. To obtain large transconductance, the n-MOS devices are designed to have W = 7500m and L = 12 ␮m. A 2.5 ␮m thick PVDF-TrFE piezoelectric poly-mer film is then spin coated, at 3000 rpm for 30 s, from 10% concentrated solution.13 Further steps for in situ processing of polymer film include annealing at 120 ° C for 3 h, fol-lowed by vacuum deposition of 200 nm metal共Au/Cr兲 on top of the polymer film and thereafter dry etching the polymer film. The P共VDF-TrFE兲 film is then be polarized in situ by applying a voltage at the rate of 80– 100 V/␮m共200–250 V, in present case兲. With such a high voltage there is risk of damaging or altering the characteristics of underlying MOS device. To overcome this challenge, all metal layers and the substrate under the polymer film were grounded before

ap-0 0.5 1 1.5 2 2.5 3 0 0.5 1 1.5 2 2.5 3 3.5 Vgs(V) I d (mA ) 0 1 2 3 4 5 0 0.5 1 1.5 2 2.5 3 3.5 4 4.5 Vds(V) I d (mA) 0 0.5 1 1.5 2 2.5 3 0 0.5 1 1.5 2 2.5 3 3.5 Vgs(V) I d (mA ) Vds= 0.5 V Vds= 1.0 V Vds= 1.5 V V gs= 0.5 V V gs= 1.5 V Vgs= 2.5 V Before Poling After Poling X Before Poling After Poling X

FIG. 2. Input characteristics (left) and output characteristics (right) before and after poling.

Charge Amplifier

Data Acquisition/ Oscilloscope Signal from POSFET

Signal from Force Sensor Force Sensor POSFET Vibration Vibrator/Shaker Mechanical Stimulus 100 K -5 V Vout -0.6 -0.4 -0.2 0 0.2 0.4 0.6 -0.04 -0.03 -0.02 -0.01 0.00 0.01 0.02 0.03 0.04 0 0.0005 0.001 0.0015 0.002 A pp lied Force (N ) PO SF ET Output (V) Time (sec) POSFET Output Applied Force

FIG. 3. 共Color online兲 (top-left) POSFET connected in source-follower ar-rangement with floating gate共lower metal兲; (top-right) response of POSFET to a 1N 共peak-peak兲, 925 Hz sinusoidal force; and (bottom) experimental arrangement.

034105-2 Dahiya et al. Appl. Phys. Lett. 95, 034105共2009兲

(3)

plying the voltage across the polymer film. Poling was per-formed at 85 ° C, to reduce the voltage needed for polariza-tion, and 200 V was finally used. Further care was taken by applying the voltage in four cumulative steps of 50 V each.13 The similar characteristics of POSFET touch sensing devices before and after the poling step, shown in Fig.2, confirm that the steps taken to prevent device damage are sufficient.

The response of POSFET touch sensing device, to dy-namic normal forces, is obtained by connecting it in a source-follower floating gate arrangement, as shown in Fig.

3. Source-follower configuration results in less than unity gain and hence POSFET devices can be tested for wide range of forces. Alternately, common source configuration can be used to have better force resolution. The experimental ar-rangement used to apply sinusoidal forces in the normal di-rection is also shown in Fig. 3. The POSFET device is pressed by vibration generator or shaker against the force sensor, which is equipped with a probe of dimensions 1.3 ⫻1.3 mm2. In this way the force applied by shaker is re-corded by the force sensor and the output of POSFET is read on the oscilloscope. The output of a POSFET touch sensing device due to 925 Hz sinusoidal normal force applied on top of the device is shown in Fig.3. When applying forces, the sensor is covered with a thin 300 ␮m protective PDMS rub-ber. The response of three POSFET touch sensing devices, shown in Fig. 4共a兲, to a range of dynamic normal forces 共0.15–5 N, sinusoidal, 20 Hz兲 is linear with a slope of 50 mV/N. Considering various polymer and device parameters the expected response of POSFET device is approximately 78 mV/N—a value little higher than the observed response. This range of forces very well covers the range of forces 共0.15–0.9 N兲 involved in normal manipulative tasks in hu-mans. It should be noted that the drain current共Id兲 of MOS devices also changes with stress mainly as a result of mobil-ity variations.14 In fact this piezoresistive property has been exploited to obtain MOS based pressure sensors. To under-stand the extent to which such effects contribute to the POSFET response, the P共VDF-TrFE兲 film was removed and forces up to 25 N were applied with and without a thin PDMS film on the gate. For comparison the device was con-nected as in Fig.3 and the gate voltage varied to obtain the response in linear and saturation regions. The response due to piezoresistive effect varies with biasing condition,15 as evident from plots in Fig.4共b兲. With POSFET connected as

in Fig. 3共a兲 ⌬Id/Id of 0.01 is needed to obtain 50 mV/N output. The ⌬Id/Id due to piezoresistive effect is about 0.0001 for 1 N i.e., 1% of⌬Id/Idneeded to obtain 50 mV/N POSFET output. Hence the response of POSFET is mainly due to the piezoelectric behavior of PVDF-TrFE film.

In summary, the POSFET touch sensing devices have a linear response over wide range of forces. The response of POSFET is mainly due to piezoelectric behavior of P共VDF-TrFE兲 film and other effects such as piezoresistive, that show up in presence of stress, contribute only to a minor extent. The POSFET devices as an integral sensor and elec-tronic unit offer practical advantages such as reduction in number of wires which is essential for the target application, namely, fingertips of robotic hands.

This work was supported in part by European Commis-sion project ROBOSKIN 共Grant No. ICT-FP7-231500兲 and Italian MIUR project PRIN 2007 “Tactile Sensing System for Humanoid Robots using Piezo-polymer-FET devices.”

1M. R. Cutkosky, R. D. Howe, and W. Provancher, Springer Handbook of

Robotics共Springer, Berlin, Heidelberg, 2008兲.

2H. Takao, K. Sawada, and M. Ishida,IEEE Trans. Electron Devices 53,

1250共2006兲.

3T. Someya, T. Sekitani, S. Iba, Y. Kato, H. Kawaguchi, and T. Sakurai,

Proc. Natl. Acad. Sci. U.S.A. 101, 9966共2004兲.

4D. H. Kim, J. H. Ahn, H. S. Kim, K. J. Lee, T. H. Kim, C. J. Yu, R. G.

Nuzzo, and J. A. Rogers,IEEE Electron Device Lett. 29, 73共2008兲. 5H. Ishiwara,Curr. Appl. Phys. 9, S2共2009兲.

6R. S. Dahiya, M. Valle, and L. Lorenzelli,IEEE Trans. Ultrason.

Ferro-electr. Freq. Control 56, 387共2009兲.

7R. S. Johansson and I. Birznieks,Nat. Neurosci. 7, 170共2004兲. 8R. G. Swartz and J. D. Plummer,IEEE Trans. Electron Devices 26, 1921

共1979兲.

9R. S. Dahiya, G. Metta, and M. Valle, Proceedings of the Fifth IEEE

International Conference on Mechatronics共IEEE, New York, 2009兲, Vol.

1, p. 1.

10P. I. Hsu, M. Huang, Z. Xi, S. Wagner, Z. Suo, and J. C. Sturn,J. Appl.

Phys. 95, 705共2004兲.

11D. H. Kim, J. H. Ahn, W. M. Choi, H. S. Kim, T. H. Kim, J. Song, Y. Y.

Huang, Z. Liu, C. Lu, and J. A. Rogers,Science 320, 507共2008兲. 12S. Martinoia, N. Rosso, M. Grattarola, B. M. L. Lorenzelli, and M. Zen,

Biosens. Bioelectron. 16, 1043共2001兲.

13R. S. Dahiya, M. Valle, L. Lorenzelli, G. Metta, and S. Pedrotti,

Proceed-ings of the IEEE Sensors 2008共IEEE, New York, 2008兲, Vol. 1, p. 490.

14A. T. Bradley, R. C. Jaeger, J. C. Suhling, and K. J. O’Connor, IEEE

Trans. Electron Devices 48, 2009共2001兲.

15R. Vatedka, H. Takao, K. Sawada, and M. Ishida,Sens. Actuators, A 140,

89共2007兲. 0 1 2 3 4 5 0 0.05 0.1 0.15 0.2 0.25 Force (N) P OS FET O utput (V ) 0 5 10 15 20 25 −0.25 −0.2 −0.15 −0.1 −0.05 0 Force (N) ∆ I d /I d (%) POSFET Device 1 POSFET Device 2 POSFET Device 3

Linear region (with PDMS) Linear region (without PDMS) Sat. region (with PDMS) Sat. region (without PDMS)

(a) (b)

Avg. Slope = 50 mV/N

y = − 0.0088x+0.002

y = −0.0105x+0.0036 FIG. 4.devices to dynamic normal forces共a兲 Response of three POSFET

共sinusoidal, 20 Hz兲 in the range 0.15– 5N.共b兲 ⌬Id/Idas a result of

piezore-sistive effect in MOS devices. 034105-3 Dahiya et al. Appl. Phys. Lett. 95, 034105共2009兲

Riferimenti

Documenti correlati

This prospective, multi-center, study shows that an ablation protocol respecting strict criteria for contiguity and quality lesion results in high and comparable rate of acute

— Changes in the acid-base equilibrium and blood oxygen con- tent in pulmonary diseases with respiratory insufficiencv.. — Blood gas analysis and the assesment of

It is a short deviation from orthodoxy to imagine that such non-renormalizable operators can be generated by quantum gravity ef- fects (and hence be naturally suppressed by the

Even though no plateau or hump is obvious in the transfer characteristics of devices with counter doped channel it may be beneficial to the SS to delay the onset of diagonal

Briefly, the explored aspects in this section are: the features of autism spectrum disorder according to the latest re- searches; the network of care and services for people diagnosed

Numero di passeggeri a metro (sinistra) e comfort ratio (destra) in funzione della lunghezza media dei moduli Considerando che i tram multiarticolati composti da moduli sospesi

Certamente la tutela del preminente interesse del minore a un sano e armonico sviluppo psicofisico assicurato attraverso l’effettivo apporto personale ed economico

euro di fatturato nell’ultima stagione sportiva. Nel caso particolare dell’Italia sono, però, necessari alcuni correttivi, prima per uscire dalla crisi che attanaglia la maggior