• Non ci sono risultati.

17 Magnetic Resonance Imaging for Restenosis

N/A
N/A
Protected

Academic year: 2022

Condividi "17 Magnetic Resonance Imaging for Restenosis"

Copied!
9
0
0

Testo completo

(1)

Magnetic Resonance Imaging for Restenosis

Robert Jan M. van Geuns, MD , P h D and Timo Baks, MD

C

ONTENTS

I

NTRODUCTION

S

AFETY OF

S

TENTS

S

TENT

R

ELATED

A

RTEFACTS

MR F

LOW

M

EASUREMENTS

MR C

OMPATIBLE

(T

RANSPARENT

) S

TENTS

S

UMMARY

R

EFERENCES

17

INTRODUCTION

Magnetic resonance imaging (MRI) of the coronary arteries is a difficult subject.

Owing to the motion of the heart during contraction and respiration artefact free imag- ing needs extremely fast techniques that MRI does not fully qualify yet. However, con- stant progress is made (1) and with electrocardiography triggering and respiratory motion correction techniques, submilliter resolution can be achieved with reasonable acquisition times (2). Only one study focussed on the use of magnetic resonance (MR) coronary angiography techniques for the detection of restenosis after coronary angioplasty (3). In this study, in which 118 patients after percutaneous transluminal coronary angioplasty were investigated, sensitivity for restenosis detection was 73%

(11/15) and specificity was 49%. Overall accuracy was 53% for MRI, which was sig- nificantly (p = 0.014) lower compared with electron beam computed tomography.

Unfortunately, MR coronary angiography of restenosis is currently hindered by the obligatory use of coronary artery stents to decrease the incidence of restenosis. Still MRI for the detection of restenosis is being performed using different alternative tech- niques and this subject will be reviewed in this chapter.

SAFETY OF STENTS

Coronary artery stents have demonstrated their efficacy as additional treatment in angioplasty, initially for treatment of abrupt vessel closure or dissection, but later also for reducing restenosis as demonstrated in the STRESS and BENESTENT studies (4,5).

From: Contemporary Cardiology: Essentials of Restenosis: For the Interventional Cardiologist Edited by: H. J. Duckers, E. G. Nabel, and P. W. Serruys © Humana Press Inc., Totowa, NJ

277

(2)

Since then stents have been used in more than 80% of percutaneous coronary interven- tion (PCI) procedures in 2000 and presently this approaches 100%. As most stents are produced of 316 low-carbon Stainless steel, safety in MRI scanners has always been an issue. First, the magnetic field deploys a force on the stent depending on the static magnetic field strength (usually 1 or 1.5 T, in the future 3 T) and the applied gradients during the examination (6), the materials used (degree of ferromagnetism), the mass of the implant used, and the location and orientation relative to the magnetic field. These induced forces may in theory lead to displacement of the stent. Nevertheless these forces are much smaller then those used during implantation or those that the stents experience during motion of the heart at the time of contraction (7). The second issue is heating because stents are closed loops of conductive low-ferromagnetism materials and rapid changing magnetic forces can induce electrical currents that can cause local heating. Thermal injuries during MR examination associated with electrical monitoring devices have been described (8). Radiofrequency (RF) pulses used are a second source of heating. Several studies have evaluated the heating effects in MR scanners, in one study on a nitinol-based guidewire, temperature rose by a maximum of 48°C in a phan- tom (9,10). Heating of MR tracking catheters has also been reported (11), although to a lesser degree. Studies on coronary artery stents on the other hand showed a maximum rise of 0.3 °C in one study (12) and no significant heating effects in another (7,13,14).

This is explained by the length of the conducting material used and by flow phantoms that allow some cooling by continuous refreshment of the blood within the stent.

MRI after stent implantation is considered to be safe once endothelialization has occurred (15), because endothelialization presumably opposes possible dislodgement (16,17). Actual patient information on the Internet by the American Heart Association advises to postpone MR imaging till 4 wk after the procedure. On the contrary, the American Heart Association Diagnostic and Interventional Catheterization Committee does not preclude MRI in the presence of coronary stents when it is clinically indicated (18). Additionally, Gerber et al. (19) have demonstrated in a retrospective clinical trial the safety of MRI within 8 wk (median of 18 d) after stent implantation in 111 patients;

no stent thrombosis or cardiac death occurred within 30 d. Recent studies in celltrans- plantation mostly treated with PCI including stent implantation, using repetitive MRI measurements (TOPCARE-AMI and BOOST, mean 4.7 and 3.5 d postPCI, respectively) showed no adverse events (20,21). Supported by these last studies no restrictions on MRI after coronary stent implantation are applied in the institution.

STENT RELATED ARTEFACTS

Currently, the majority of coronary stents are nonferromagnetic and considered MRI safe as explained earlier, but stents remain the source of image artefacts owing to local magnetic field inhomogeneities and provoked eddy currents in the conductive material of the stent, hampering image interpretation. The size of the artefacts is dependent on the materials and imaging sequences used. Here in principle Spin-echo techniques and its derivatives show significant smaller artefacts than gradient echo-sequences (7).

Comparing 14 different stents lumen was visible in all except one colbalt stent, and in

10 of the 14 stents artificial lumen narrowing was less than 33%, using a contrast

enhanced magnetic resonance angiography sequences that can be considered standard

for noncoronary angiography (22). Best results were obtained with Nitinol (a

nickel–titanium alloy) and Tantalum stents, although within the Nitinol stents a variable

appearance was observed depending on the stent geometry. Artefacts are reduced by

(3)

using short echo-times wherein the shorter echo-times in gradient-echo sequences can compensate for the increased sensitivity of GE imaging to field inhomogeneities. Besides the materials and techniques used it was shown that the angle β to B0 was the main fac- tor that influenced the artefact size, with a larger angle increasing the artefact in a non- linear relationship (23–25). As artefacts hamper lumen visualization on most MRA and all MR coronary angiography techniques after current stent treatment, alternative analy- sis techniques and stent material and design are being investigated.

MR FLOW MEASUREMENTS

A very attractive tool in MRI is the ability to noninvasive measure flow with dedi- cated MR flow sensitive techniques. This has been used to quantify valve regurgitation and stenosis (26) and is frequently used in congenital heart disease (27). The same technique has been used to quantify flow in the coronary arteries and detection of flow- limiting stenosis in native coronary arteries (28–30) and bypass grafts (31–33) using vasodilatators. In these settings a reasonable sensitivity and specificity for graft func- tion has been obtained. Hundley et al. (34) used these flow techniques to measure flow reserve in patients with recurrent angina after balloon angioplasty. For a flow reserve of less than 2, 100% and 82% sensitive and 89% and 100% specific, detecting a luminal diameter narrowing of less than or equal to 70% and 50%, respectively was achieved (Table 1). The same approach was used by Saito et al. (35) for the detection of instent restenosis, with the modification that flow was measured just distal from the stent to avoid the imaging artefact related to stainless steel stents. They demonstrated that flow could accurately be measured just distal to the stent (Fig. 1). Nagel et al. (36) studied 38 patients after successful PCI with stent deployment with this approach and coronary flow velocity reserve with MRI similar to Doppler results (r = 0.87), with a mean rela- tive difference of 7.5%. Using a threshold of 1.2 for coronary flow velocity reserve, a sensitivity of 83% with a specificity of 94% was achieved for more than or equal to 75% stenoses. In both these studies imaging was limited to the proximal vessels and left anterior descending coronary arteries were nearly exclusively included, addition- ally, patients with suspected microvasculare disease were excluded.

MR COMPATIBLE (TRANSPARENT) STENTS

Metallic stents signal attenuation within the stent is caused by RF (RF shielding) of the metallic stent material. Induced eddy currents in the stent may also lead to a lower nominal RF excitation angle inside the stent (37,38). Only with large-diameter nitinol stents in the iliac arteries have sufficiently small artefacts been reported to allow diag- nostic follow-up, either by standard MRA (39) or by increasing the excitation angle of the MRA sequence (23). As the material of stent is of great influence on stent artefact,

Table 1

Sensitivity and Specificity of MR Coronary Flow Measurements for the Detection of Restenosis After PCI

Restenosis Year Treatment Patients Sign Sensitivity Specificity

Hundley 2000 Ballon 17 ≤50% 82 100

Saito 2001 Stent 10 – – –

Nagel 2003 Stent 38 ≥75% 83 94

(4)

a quest for metal alloys with minimal artefacts was started to develop a truly MR trans- parent stent. One alloy consisting of copper (75%), silver (8%), platinum (2%), gold (14%), and palladium (1%) was proposed to minimize susceptibility artefacts (40).

The examined prototypes of fully MR-compatible MRI stents allowed artefact-free visualization of the stent lumen with phase-contrast and contrast-enhanced T1-weighted angiography, as well as phase-contrast flow measurements in the stented area (41).

As copper can induce an inflammatory reaction causing in-stent restenosis, the large amount in this alloy is worrying, but biocompatible coatings of the MR stent may solve this problem. Another alloy proposed by van Dijk et al. (42) consists mainly of palla- dium and silver and is iron- and nickel-free. The mechanical properties of ABI alloy resemble Stainless steel. Using 2D and 3D gradient echo sequences artefacts of this ABI alloy stent was significantly smaller than Nitinol or Tantalum stent. So far only a simple helix has been built out of this material.

The BIOTRONIK absorbable metal stent (BIOTRONIK, Bulach, Switzerland) is a

magnesium alloy stent that undergoes degradation over 2- to 3-mo period after deploy-

ment. This stent is reported compatible with MR angiography. As the stent is absorbable

their use does not require prolonged antiplatelet therapy and neointimal proliferation,

restenosis, and chronic inflammation may be reduced. Recently, results from a first-

in-man BEST BTK trial, in which absorbable metal stents were implanted below the

knee were reported (EuroPCR2004). Between December 2003 and January 2004, 63

patients with critical limb ischemia were enrolled just before limb amputation

Fig. 1. (A) Definition of slice location for MRI blood flow measurements in the LAD artery on dou- ble oblique scout magnetic resonance imaging (MRI) images. (B) Magnitude and phase-difference images acquired with fast velocity-encoded cine MRI acquisition. (C) MRI blood flow velocity curves in the LAD artery before and after intravenous injection of dipyridamole. Reproduced from ref. 35.

(5)

(Rutherford 4 and 5). At 3 mo, 89% of the stented vessels remained patent and a ran- domized trial is planned by the company. One limitation is that the magnesium alloy is not visible on fluoroscopy, which is partly resolved by two markers on the balloon that guide the procedure. With this kind of stent, instent thrombus has been identified for the first time noninvasively (43) (Fig. 2).

Another interesting approach to visualize the stent lumen is to use the stent as a receiver coil itself (44). To achieve this stents need to be designed to act as active reso- nant structures at the Larmor frequency of the MR system. The stents thus acted as local RF signal amplifiers. The first generation used a coaxial cable to transport the signal to the scanner system. However, such a cable-based approach for high-resolution in-stent imaging is invasive and thus is limited to the time of the intervention. Follow- up studies would be difficult to achieve, because the approach requires electrical con- tact between the stent and the MR scanner, and hence repeated invasive access. In a more sophisticated way, Quick et al. (45) employed the principle of inductive coupling wherein the B(1) fields of the stents were coupled to that of an outside surface coil.

Thus, the signal from the stent antenna could be wireless and received by a surface coil connected to the MR system (Fig. 3). This concept has been tested in vitro and in vivo and a near 30 times improvement in signal intensity in the stent lumen has been demon- strated. Unfortunately, several limitations are to be noted. First, low flip angle sequences were necessary as transmit flip angles were also amplified and artefacts were intro- duced owing to inhomogeneity of the B1 field of the stent.

Fig. 2. In vivo magnetic resonance imaging of Gd-labeled fibrinogen clots. (A,D) Coronary MRA before (A) and after (D) thrombus delivery. On both scans, no apparent thrombus is visible (circle).

(B,E) Black-blood inversion recovery TFE scans before (B) and after (E) clot delivery (same view as A and D). After thrombus delivery (E), 3 bright areas are readily visible (arrows and circle), consis- tent with location of thrombus. No apparent thrombus was visible on prethrombus (B) images (arrow and circle). (C) X-ray angiogram confirming MR finding of thrombus in mid-LAD (circle). (F) Magnified view of C. LM indicates left main. Reproduced from ref. 43.

(6)

Additionally, the use of RF-intensive sequences potentially by this amplification can

lead to local concentration of RF energy (hot spots) in the vicinity of the stent, which

could result in heating of surrounding tissues. Decoupling, i.e., detuning of the resonant

structure during RF transmission, would decrease the potential risk of RF heating, and

can be achieved passively by incorporation of additional electronic elements (e.g.,

Fig. 3. (A) Photograph of a solenoidal stent prototype (length 35 mm, inner diameter 4 mm). The 14- turn solenoidal stent mesh was tuned with a chip capacitor (arrow) to the resonant frequency of the scanner. (B) Photograph of a balloon-expandable self-resonant stent prototype (length 28 mm, inner diameter before/after expansion 1.8/4 mm); stent in the folded state, (C) stent mounted on a 5F bal- loon catheter (balloon 40 × 4 mm2) after full inflation of balloon, (D) fully deployed stent, and (E) MR image acquired with the stent immersed in an NaCl phantom acquired with the following sequence: FLASH 2D sequence, TR/TE = 200/11 ms, flip = 2°, FOV = 60 × 60 mm2, matrix= 512 × 512, slice = 2 mm2, time= 1:44 min, in-plane resolution 117 × 117 m2. Reproduced from ref. 45.

(7)

crossed diodes) in the design. Second, optimal stent design for balloon-expandable delivery is more difficult and the redesigned solenoidal design with a zigzag pattern of the stent wire resulted in local signal cancellations (Fig. 3). Third, the stents should be coated with a polymer to guarantee electric isolation from the surrounding tissue and blood, otherwise electricity and therefore signal may be lost. One author performed a study on heating of coronary artery stents with the intention to induce cell necrosis and thereby restenosis (46). Afterwards with maximum energy power output they were able to raise the temperature of the stent to 80

°

C with necrosis on microscopy.

SUMMARY

In the past coronary artery stents where regarded as possible contraindications for MRI.

Additional safety studies have resolved this issue and MRI is now frequently performed even within the first days of implantation. Artefacts related to the stent hinder image inter- pretation but flow measurement just distal of a stent have proven to be able to detect sig- nificant instent restenosis. More recent development includes MR transparent stents and in the future, stent might serve as local imaging coils to increase signal to noise ratio.

REFERENCES

1. van Geuns RJ, Wielopolski PA, de Bruin HG, et al. Magnetic resonance imaging of the coronary arteries: techniques and results. Prog Cardiovasc Dis 1999;42:157–166.

2. Stuber M, Botnar RM, Danias PG, et al. Double-oblique free-breathing high resolution three-dimen- sional coronary magnetic resonance angiography. J Am Coll Cardiol 1999;34:524–531.

3. Ropers D, Regenfus M, Stilianakis N, et al. A direct comparison of noninvasive coronary angiography by electron beam tomography and navigator-echo-based magnetic resonance imaging for the detec- tion of restenosis following coronary angioplasty. Invest Radiol 2002;37:386–392.

4. Schatz RA, Baim DS, Leon M, et al. Clinical experience with the Palmaz-Schatz coronary stent.

Initial results of a multicenter study. Circulation 1991;83:148–161.

5. Serruys PW, de Jaegere P, Kiemeneij F, et al. A comparison of balloon-expandable-stent implantation with balloon angioplasty in patients with coronary artery disease. Benestent Study Group. N Engl J Med 1994;331:489–495.

6. Shellock FG. Biological effects and safety aspects of magnetic resonance imaging. Magn Reson Q 1989;5:243–261.

7. Hug J, Nagel E, Bornstedt A, Schnackenburg B, Oswald H, Fleck E. Coronary arterial stents: safety and artifacts during MR imaging. Radiology 2000;216:781–787.

8. Dempsey MF, Condon B. Thermal injuries associated with MRI. Clin Radiol 2001;56:457–465.

9. Konings MK, Bartels LW, Smits HF, Bakker CJ. Heating around intravascular guidewires by res- onating RF waves. J Magn Reson Imaging 2000;12:79–85.

10. Liu CY, Farahani K, Lu DS, Duckwiler G, Oppelt A. Safety of MRI-guided endovascular guidewire applications. J Magn Reson Imaging 2000;12:75–78.

11. Wildermuth S, Dumoulin CL, Pfammatter T, Maier SE, Hofmann E, Debatin JF. MR-guided percu- taneous angioplasty: assessment of tracking safety, catheter handling and functionality. Cardiovasc Intervent Radiol 1998;21:404–410.

12. Shellock FG, Shellock VJ. Metallic stents: evaluation of MR imaging safety. AJR Am J Roentgenol 1999;173:543–547.

13. Friedrich MG, Strohm O, Kivelitz D, et al. Behaviour of implantable coronary stents during mag- netic resonance imaging. Int J Cardiovasc Intervent 1999;2:217–222.

14. Strohm O, Kivelitz D, Gross W, et al. Safety of implantable coronary stents during 1H-magnetic res- onance imaging at 1.0 and 1.5 T. J Cardiovasc Magn Reson 1999;1:239–245.

15. Roubin GS, Robinson KA, King SB 3rd, et al. Early and late results of intracoronary arterial stenting after coronary angioplasty in dogs. Circulation 1987;76:891–897.

16. Shellock F. Pocket Guide to MR Procedures and Metallic Objects: Update 2001. Lippincott Williams

& Wilkins, Philadelphia, PA, 2001.

17. Jost C, Kumar VV. Are Current Cardiovascular Stents MRI Safe? J Invasive Cardiol 1998;

10:477–479.

(8)

18. Levine GN, Kern MJ, Berger PB, et al. Management of patients undergoing percutaneous coronary revascularization. Ann Intern Med 2003;139:123–136.

19. Gerber TC, Fasseas P, Lennon RJ, et al. Clinical safety of magnetic resonance imaging early after coronary artery stent placement. J Am Coll Cardiol 2003;42:1295–1298.

20. Britten MB, Abolmaali ND, Assmus B, et al. Infarct remodeling after intracoronary progenitor cell treatment in patients with acute myocardial infarction (TOPCARE-AMI): mechanistic insights from serial contrast-enhanced magnetic resonance imaging. Circulation 2003;108:2212–2218.

21. Wollert KC, Meyer GP, Lotz J, et al. Intracoronary autologous bone-marrow cell transfer after myocardial infarction: the BOOST randomised controlled clinical trial. Lancet 2004;364:141–148.

22. Lenhart M, Volk M, Manke C, et al. Stent appearance at contrast-enhanced MR angiography: in vitro examination with 14 stents. Radiology 2000;217:173–178.

23. Meyer JM, Buecker A, Schuermann K, Ruebben A, Guenther RW. MR evaluation of stent patency:

in vitro test of 22 metallic stents and the possibility of determining their patency by MR angio- graphy. Invest Radiol 2000;35:739–746.

24. Bartels LW, Smits HF, Bakker CJ, Viergever MA. MR imaging of vascular stents: effects of suscep- tibility, flow, and radiofrequency eddy currents. J Vasc Interv Radiol 2001;12:365–371.

25. Meyer JM, Buecker A, Spuentrup E, et al. Improved in-stent magnetic resonance angiography with high flip angle excitation. Invest Radiol 2001;36:677–681.

26. Caruthers SD, Lin SJ, Brown P, et al. Practical value of cardiac magnetic resonance imaging for clin- ical quantification of aortic valve stenosis: comparison with echocardiography. Circulation 2003;108:2236–2243.

27. Simpson IA, Chung KJ, Glass RF, Sahn DJ, Sherman FS, Hesselink J. Cine magnetic resonance imaging for evaluation of anatomy and flow relations in infants and children with coarctation of the aorta. Circulation 1988;78:142–148.

28. Clarke GD, Eckels R, Chaney C, et al. Measurement of absolute epicardial coronary artery flow and flow reserve with breath-hold cine phase-contrast magnetic resonance imaging. Circulation 1995;91:

2627–2634.

29. Hundley WG, Lange RA, Clarke GD, et al. Assessment of coronary arterial flow and flow reserve in humans with magnetic resonance imaging. Circulation 1996;93:1502–1508.

30. Davis CP, Liu PF, Hauser M, et al. Coronary flow and coronary flow reserve measurements in humans with breath-held magnetic resonance phase contrast velocity mapping. Magn Reson Med 1997;

37:537–544.

31. Hoogendoorn LI, Pattynama PM, Buis B, et al. Noninvasive evaluation of aortocoronary bypass grafts with magnetic resonance flow mapping. Am J Cardiol 1995;75:845–848.

32. Galjee MA, van Rossum AC, Doesburg T, Hofman MB, Falke TH, Visser CA. Quantification of coronary artery bypass graft flow by magnetic resonance phase velocity mapping. Magn Reson Imaging 1996;14:485–493.

33. Langerak SE, Kunz P, de Roos A, Vliegen HW, van Der Wall EE. Evaluation of coronary artery bypass grafts by magnetic resonance imaging. J Magn Reson Imaging 1999;10:434–441.

34. Hundley WG, Hillis LD, Hamilton CA, et al. Assessment of coronary arterial restenosis with phase- contrast magnetic resonance imaging measurements of coronary flow reserve. Circulation 2000;

101:2375–2381.

35. Saito Y, Sakuma H, Shibata M, et al. Assessment of coronary flow velocity reserve using fast veloc- ity-encoded cine MRI for noninvasive detection of restenosis after coronary stent implantation.

J Cardiovasc Magn Reson 2001;3:209–214.

36. Nagel E, Thouet T, Klein C, et al. Noninvasive determination of coronary blood flow velocity with cardiovascular magnetic resonance in patients after stent deployment. Circulation 2003;107:

1738–1743.

37. Maintz D, Kugel H, Schellhammer F, Landwehr P. In vitro evaluation of intravascular stent artifacts in three-dimensional MR angiography. Invest Radiol 2001;36:218–224.

38. Bartels LW, Bakker CJ, Viergever MA. Improved lumen visualization in metallic vascular implants by reducing RF artifacts. Magn Reson Med 2002;47:171–180.

39. Juergens KU, Tombach B, Reimer P, Vestring T, Heindel W. Three-dimensional contrast-enhanced MR angiography of endovascular covered stents in patients with peripheral arterial occlusive disease.

AJR Am J Roentgenol 2001;176:1299–1303.

40. Buecker A, Spuentrup E, Ruebben A, Gunther RW. Artifact-free in-stent lumen visualization by stan- dard magnetic resonance angiography using a new metallic magnetic resonance imaging stent.

Circulation 2002;105:1772–1775.

(9)

41. Buecker A, Spuentrup E, Ruebben A, et al. New metallic MR stents for artifact-free coronary MR angiography: feasibility study in a swine model. Invest Radiol 2004;39:250–253.

42. van Dijk LC, van Holten J, van Dijk BP, Matheijssen NA, Pattynama PM. A precious metal alloy for con- struction of MR imaging-compatible balloon-expandable vascular stents. Radiology 2001;219:284–287.

43. Botnar RM, Perez AS, Witte S, et al. In vivo molecular imaging of acute and subacute thrombosis using a fibrin-binding magnetic resonance imaging contrast agent. Circulation 2004;109:2023–2029.

44. Quick HH, Ladd ME, Nanz D, Mikolajczyk KP, Debatin JF. Vascular stents as RF antennas for intravascular MR guidance and imaging. Magn Reson Med 1999;42:738–745.

45. Quick HH, Kuehl H, Kaiser G, Bosk S, Debatin JF, Ladd ME. Inductively coupled stent antennas in MRI. Magn Reson Med 2002;48:781–790.

46. Floren MG, Gunther RW, Schmitz-Rode T. Noninvasive inductive stent heating: alternative approach to prevent instent restenosis? Invest Radiol 2004;39:264–270.

Riferimenti

Documenti correlati

– immediately after the 90° pulse, a negative magnetic field gradient is applied; the spins start precessing at a position-dependant speed, dephas- ing in an “ordinate” manner with

Target Volume Defi nition: Tissue Oxygenation The BOLD contrast can contribute to a further im- provement of stereotactic planning while areas of high oxygenation and vascularity can

Börnert, Novel prospective respiratory motion correction approach for free-breathing coronary MR angiography using a patient-adapted affine motion model, Magn Reson Med 50,

Recent advances in magnetic resonance imaging and magnetic resonance spectroscopy imaging have the potential to offer better target delineation in multiple tumor sites and for

Magnetic resonance fluoroscopic images of the pelvis at rest (a), during squeeze (b), and simulated defecation (c) in a 52-year-old asymptomatic patient after filling the rectum

The sensitivity of MRI to breast carcinoma, particularly in dense breast tissue, has led to the emerging role of MRI in breast cancer screening for women identified to be at

60 Magnetic resonance imaging for extent of disease assessment may be most valuable for women with a strong family history of breast cancer and invasive lobular histology in the

Monitoring the effects of neoadjuvant therapy is impor- tant to assess effectiveness of chemotherapy after initial doses, thereby determining if a specific treatment should be