Rapid 2nd-Tier Test for Measurement of
3-OH-Propionic and Methylmalonic Acids on
Dried Blood Spots: Reducing the False-Positive
Rate for Propionylcarnitine during Expanded
Newborn Screening by Liquid Chromatography–
Tandem Mass Spectrometry
Giancarlo la Marca,
1*Sabrina Malvagia,
1Elisabetta Pasquini,
1Marzia Innocenti,
2Maria Alice Donati,
1and Enrico Zammarchi
1Background: The expansion of newborn screening
pro-grams has increased the number of newborns diag-nosed with inborn errors of metabolism in the pre-symptomatic phase, but it has also increased the number of costly, stress-producing false-positive results. Be-cause propionylcarnitine (C3) is one of the analytes most frequently responsible for false-positive results, we aimed to develop a rapid liquid chromatography– tandem mass spectrometry (LC-MS/MS) method to identify free methylmalonic (MMA) and 3-OH propi-onic (3OH-PA) acids in blood spots.
Methods: We studied newborn screening spots from 250
healthy controls; 124 from infants with abnormal C3, of whom only 5 (4%) were truly affected; 124 from infants with altered isolated methylmalonylcarnitine; and 4 from clinically diagnosed patients. Whole blood was eluted from a 3.2-mm dried blood spot by a CH3CN/H2O
7:3 and 5 mL/L formic. This extract was injected into a LC-MS/MS equipped with pneumatically assisted elec-trospray without derivatization. Total analysis time was 5 min per sample.
Results: The assays were linear up to 3300 nmol/L for
both metabolites. Intra- and interassay imprecision data were 3.6%– 8% and 3.1%– 6%, respectively, for MMA
and 5.2%–20% and 3.6%–17% for 3OH-PA. Limit of detection and limit of quantitation were 1.95 and 4.2mol/L, respectively, for MMA and 8 and 10 mol/L for 3OH-PA. The recoveries were 92.9%–106.1%. No deterioration was noted on the columns after 500 chro-matographic runs. If the new method had been used as a 2nd-tier test for the 124 samples, only the 5 true positives would have been recalled for additional sam-ples, and the positive predictive value would have been 100%.
Conclusions: This method has the potential to
mark-edly reduce false-positive results and the associated costs and anxiety. It may also be suitable for diagnosing and routinely monitoring blood spots for methylma-lonic aciduria and propionic acidemia.
© 2007 American Association for Clinical Chemistry
The introduction of an expanded liquid chromatography– tandem mass spectrometry (LC-MS/MS)3– based
new-born screening program has significantly increased early diagnoses of inborn errors of metabolism. False-positive screening results have also increased since the time when neonatal screening was limited to phenylketonuria and congenital hypothyroidism (1, 2 ). This increase in retest-ing has had important consequences, includretest-ing increased 1Metabolic Unit, Department of Paediatrics, Meyer Children’s Hospital
and2Pharmaceuticals Department, University of Florence, Florence, Italy.
* Address correspondence to this author at: Meyer Children’s Hospital, Via Luca Giordano 13, 50132 Florence, Italy. Fax 39-0555662541, e-mail: g.lamarca@meyer.it.
Received February 19, 2007; accepted April 24, 2007.
Previously published online at DOI: 10.1373/clinchem.2007.087775
3Nonstandard abbreviations: LC-MS/MS, liquid chromatography–tandem
mass spectrometry; C3, propionylcarnitine; C4, isobutyryl/butyrylcarnitine; LA, lactic acid (2-hydroxypropionic acid);⌴⌴〈, methylmalonic acid; SA, succinic acid; 3OH-PA, 3-OH-propionic acid; C4DC, methylmalonylcarnitine; PA, propi-onic acidemia; MRM, multiple reaction monitoring.
laboratory analyses, personnel costs for repeat tests, and considerable anxiety for parents (3– 6 ). The impact of a screening recall on a family is substantial (7, 8 ). Even after a normal retesting, a screening recall often results in increased anxiety over a child’s health, an altered parent-child relationship, and increased hospitalizations for un-related illnesses (2, 6, 9 ). Many authors have tried to assess what levels of parental stress are acceptable (10 ).
An expanded program for newborn screening by LC-MS/MS was established in Tuscany in 2001 (the first regional expanded newborn screening program in Italy). Our experience shows that some metabolites—propionyl-carnitine (C3), isobutyryl/butyrylmetabolites—propionyl-carnitine (C4), and ty-rosine—are more likely than others to cause false-positive results and increase recall rates. C3, in particular, causes a high number of false-positive results. In ⬃67 000 new-borns screened by our center between November 2004 and November 2006, the recall rate for altered C3 wasⱖ20 percent of all recalls. Diagnosis of propionic acidemia (PA) was confirmed in 2 patients and methylmalonic aciduria in 3 patients. The positive predictive value of C3 was 4%.
These data encouraged us to develop a strategy to reduce false-positive rates and improve the positive pre-dictive value of an initial abnormal result.
Materials and Methods
Chemical standard 2-hydroxypropionic acid [lactic acid (LA)], methylmalonic acid (MMA), and succinic acid (SA) were purchased from Sigma-Aldrich; 3-OH-propionic acid (3OH-PA) was from Tokyo Chemical Industries. A working calibrating solution containing 10 mol/L of each was prepared in CH3CN/H2O 7:3 containing 5 mL
formic acid per liter. MMA (methyl-D3) was from
Cam-bridge Isotopes Laboratories.
We tested 250 newborn screening spots from healthy controls, 119 spots with false-positive results due to
abnormal values of C3, and 124 spots with isolated methylmalonylcarnitine (C4DC) outside the reference in-terval (Table 1), although we have not presupposed any recall for altered isolated C4DC. In addition, we studied newborn screening spots from 9 patients with confirmed diagnoses: 4 patients with PA and 5 patients with meth-ylmalonic aciduria. Two patients with PA (PA3 and PA4) and 3 with methylmalonic aciduria (MMA1, MMA2, and MMA3) were identified by newborn screening; in 2 PA patients (PA1 and PA2) and 2 methylmalonic aciduria patients (MMA4 and MMA5), diagnosis was made clini-cally before the expanded newborn screening program was started. In the latter cases, stored dried blood spots from newborn screening were retrospectively analyzed. PA1 and PA2 had acute neonatal onset, whereas PA3 and PA4 were asymptomatic when newborn screening results were available. MMA1 was due to maternal cobalamin deficiency; in MMA2 no mutation was identified in the MUT4 (methylmalonyl coenzyme A mutase) gene, and
complementation studies are in progress. MMA3, MMA4, and MMA5 had mutase deficiencies; MMA2 and MMA3 had clinical symptoms when recalled for neonatal screen-ing; MMA4 had acute neonatal onset; and MMA5 had acute late onset.
Specimens of dried blood spots used as controls were collected from neonates born in Tuscany. Our local ethics committee approved the procedure. Blood collection for newborn screening purposes is made between 48 and 72 h of life. Blood taken by heel stick is spotted on filter paper (903, Whatman), dried, and sent by courier to the screen-ing center. Blood spot samples are stored at room tem-perature until analysis.
We punched a 3.2-mm filter paper disk containing ⬃3.4 L whole blood from each dried blood spot and
4Human gene: MUT, methylmalonyl coenzyme A mutase.
Table 1. MMA, 3OH-PA, C3, and C4DC in samples from screening patients, false-positives samples, C4DC-positive spots, and controls.
Sample Diagnosis year MMA,mol/L 3OH-PA,mol/L C3,mol/L C4DC,mol/L
PA1 2002 NDa 76 7.89 0.03 PA2 2002 ND 69.4 8.79 0.03 PA3 2006 ND 104 10.76 0.14 PA4 2005 ND 106.7 11.83 0.41 MMA1 2006 37.4 11.3 7.25 0.25 MMA2 2006 109.2 31.9 9.84 0.32 MMA3 2005 111.1 29.9 6.88 0.54 MMA4 1999b 83.5 29.4 0.86b 0.09b MMA5 2002b 190 27.9 4.4b 0.18b False-positive samples (n⫽ 124) ND ND 6.01–15.2 WR C4DC-positive samples (n⫽ 124) ND ND WRa 0.55–1.1 Controls (n⫽ 250) ND ND WR WR Normal values ND ND 0.2–5.65 0.04–0.54
aND, Not detectable; WR, within accepted interval. bNewborn screening spot retrospectively tested in 2005.
extracted it for 15 min with 200 L of a solution con-taining CH3CN/H2O 7:3 and 5 mL/L formic acid, plus
330 nmol/L labeled MMA as internal standard. Calibra-tors, containing internal standard at 330 nmol/L, were at concentrations of 0, 33, 165, 330, and 3300 nmol/L. We injected 2L into the LC column. For enriching studies, we evaluated linearity by analyzing supplemented 3.2-mm dried blood spots prepared at 0, 33, 165, 330, and 3300 nmol/L. The resulting calibration values were 0,
1.947, 9.735, 19.47, and 194.7 mol/L, respectively, com-pared with blood (3.4 L diluted 59-fold). The solution was shaken on a vortex-mix system for 15 min at room temperature, and 2L solution was injected into the mass spectrometer.
The hardware configuration includes an Applied Bio-systems/MDS Sciex API 4000™ Triple-Quad Mass Spec-trometer equipped with the TurboV-Spray®source with
the turbo gas temperature set at 425 °C. The source
Fig. 1. Extract ion chromatograms of control (A), C3 false positive (B), PA (C), and methylmalonic aciduria (D).
Table 2. MRM transitions, DP, CE, quantification strategies, and retention times.a MRM transitions
Measured compound
Chromatographic
retention time, min Polarity
Precursor ion
Fragment
ion DP, V CE, eV Quantification strategy
Lactic acid 1.08 ⫺ 89 59 ⫺50 ⫺14
3OH-PA 1.24 ⫺ 89 59 ⫺50 ⫺14 External calibration
SA 1.30 ⫺ 117.1 73 ⫺40 ⫺12
2
H3-MMA 2.37 ⫺ 120.1 76 ⫺40 ⫺12
2.39 ⫺ 117.1 73 ⫺40 ⫺12 Isotopic dilution
operates in negative ionization polarity at a potential of ⫺4500 V. We performed multiple reaction monitoring (MRM) measurements using declustering potential and collision energy values as automatically optimized by the software functionality for each of the analytes. A list of exploited transitions, collision energies, and declustering potentials is reported in Table 2. The choice of ionization conditions for each analyte was made to give maximum sensitivity in the experimental conditions.
We used an Agilent 1100 Quaternary Capillary-Pump for chromatography and a Gemini C6-phenyl, 3-m 100 ⫻ 2-mm (i.d.) column and a 4⫻ 2-mm precolumn cartridge (Phenomenex) for separation. The chromatographic run was performed at 200L/min with an isocratic profile of 40:60 between mobile phase of H2O (eluent A) and
CH3CN (eluent B), each containing 5 mL/L formic acid.
The next injection was performed after 5 min. Retention times are reported in Table 2.
Results
After extensive evaluation of several LC columns working either in normal phase or in reversed phase (data not shown), the best performance resulted from the C6-phenyl column: the eluent is identical to that used during newborn screening for acylcarnitines and amino acids. No substantial modifications were noted when 100% metha-nol was used both as extraction solvent and eluent B during a chromatographic run. In these conditions, the isobaric SA and MMA were completely resolved. The respective retention times were 1.30 and 2.39 min. In
the case of LA and 3OH-PA, the retention times were 1.08 and 1.24 min, respectively (Fig. 1).
All analytes are acids and display poor sensitivity when ionized in positive mode. This implies that their measurement should be done in negative ion mode for better sensitivity. As already mentioned, where the isoto-pically labeled form of MMA was available, measuments were made by isotopic dilution strategy. The re-lated labeled standard of 3OH-PA was not available, so we used the labeled MMA, which has chemical similari-ties, as internal standard.
The assays were linear up to 3300 nmol/L for both metabolites, but for the routine purpose of the protocol (screening), the MMA and 3OH-PA concentrations in samples from unaffected neonates were below detection limits. Intra- and interday imprecision data are reported in Table 3. For blood spots with added analyte, the limit of detection or limit of the blank (mean plus 3 SD of blank) for MMA was 1.95mol/L and the limit of quantification (mean plus 10 SD of blank) 4.2mol/L; for 3OH-PA the limit of detection was 8mol/L and the limit of quanti-tation 10 mol/L. Intra- and interday imprecision (CV) was in the range of 3.5%–7.8% and 3.1%– 6%, respectively, for MMA and 5.2%–19.6% and 3.6%–16.9% for 3OH-PA. Imprecision could not be determined for blood spots that had not been supplemented because the concentrations of MMA and 3OH-PA were lower than the limit of quanti-tation. The recoveries ranged from 92.9% to 106.1%. No deterioration was noted on the columns after 500 chro-matographic runs.
Table 3. Intra- and interday imprecision. Metabolite Amount added, nmol/L Intraday CV (nⴝ 6), % Interday CV (nⴝ 6), % Mean result, mol/L Recovery(nⴝ 6) Blood 1 MMA 0 0 0.0 0.0 MMA 33 3.5 3.1 34.0 103.1 MMA 165 4.9 3.7 158.0 95.8 MMA 330 7.8 6.0 336.2 101.9 MMA 3300 3.8 4.8 3299.7 100.0 PA 0 0 0.0 0.0 PA 33 19.6 16.9 32.8 92.9 PA 165 6.6 4.8 169.6 100.3 PA 330 5.2 3.6 325.4 100.6 PA 3300 5.6 6.7 3300.2 100.3 Blood 2 MMA 0 0 0.0 0.0 MMA 33 7.8 6.1 42.6 123.0 MMA 165 9.4 8.0 175.0 106.1 MMA 330 1.4 1.8 308.9 93.6 MMA 3300 1.5 2.7 3301.5 100.0 PA 0 0 0.0 0.0 PA 33 9.3 17.1 30.7 93.2 PA 165 4.4 10.6 168.3 102.0 PA 330 7.2 10.5 328.9 96.7 PA 3300 2.5 3.3 3300.0 100.0
Between November 2004 and November 2006 (n ⫽ 67 586), the Newborn Screening Tuscany program, lim-ited to LC-MS/MS testing, had a detection rate of 1:1950 and a false-positive rate of 0.83%. Of 564 total recalls, 124 (22%) were for abnormal values of isolated C3 (normal values 0.2 to 5.65mol/L) and 1 or more abnormal ratios (C3/free carnitine 0.03– 0.13; C3/C4 1.1–12.5; C3/C16 0.11–1.16). Only 5 of the 124 were true positives: 3 methylmalonic acidurias and 2 PAs.
To assess whether 2nd-tier testing could be effective, we applied it to 250 blood spots reported as normal during newborn screening by LC-MS/MS, 124 spots re-called for abnormal values of C3, and 124 spots having C4DC outside the accepted intervals. In addition, we tested 9 truly positive newborn screening blood spots (5 methylmalonic acidurias and 4 PAs). No signal corre-sponding to free diagnostic acids was detected for con-trols, false positives, or C4DC-positive spots. Free 3OH-PA, free MMA, C3, and C4DC carnitine values from affected newborns are reported in Table 1.
This methodology succeeds in distinguishing true pos-itives from false pospos-itives and controls, as shown in Fig. 1. In each panel, the extracted ion chromatograms referring to the implicated metabolites are highlighted for both the control/false positive and the affected patient.
Discussion
LC-MS/MS is increasingly gaining acceptance in clinical laboratories. In addition to its benefits in terms of sensi-tivity and specificity, it enables multiple compounds to be analyzed at one time. Routine analysis of multiple com-ponents in one fast step has been implemented in clinical laboratories for neonatal screening, steroid profiling, and so on. The method we describe allows qualification of the disorder and reduces the false-positive rate due to abnor-mal values of C3 and/or C4DC carnitines during new-born screening. To our knowledge, this is the first time free MMA and 3OH-PA involved in propionic and meth-ylmalonic acidurias have been both monitored and quan-tified using a single blood spot sample.
Other methods that measure MMA and separate it from SA as n-butyl ester derivatives in plasma and urine by LC-MS/MS have been published by several authors (11–13 ). There are several advantages of our method. It reveals contemporaneously free 3OH-PA and MMA, clearly separated from each other and differentiated from LA and SA by a fast chromatographic run in the same spot used in newborn screening. In addition, sample preparation is minimal and quick, without a derivatiza-tion step. Moreover, our method permits follow-up stud-ies and a 1st rapid evaluation in suspected patients.
MS is a promising tool in clinical analysis, both for research (mainly through proteomic and metabolomic approaches) and in routine procedures (through the quantization of targeted metabolites and markers). In the past, mass spectrometry performance was noted for sen-sitivity, resolution, and selectivity factors. Today, the
clinical laboratory demands those factors and also short-and long-term robustness. An analytical methodology cannot be exploited in the clinical domain if it is lacking robustness and requires labor-consuming sample prepa-ration (14, 15 ).
Our methodology is capable of monitoring and quan-tifying MMA and 3OH-PA during newborn screening as a 2nd-tier test. It can follow up on and diagnose PA and methylmalonic acidurias. The method is precise and ro-bust and therefore suitable for implementation in routine clinical screening programs and quantification environ-ments. In our opinion, the application of this method in newborn screening analysis will reduce or even eliminate false-positive results for C3 and prevent a great deal of unnecessary anxiety for parents and associated problems (2 ). New and better strategies should be developed to provide testing options that, applied to newborn screen-ing programs, reduce recall rate, overall expenses, and parental anxiety.
Grant/funding support: This work was partially supported by grants from the Family Association AMMEC.
Financial disclosures: None declared.
References
1. Kwon C, Farrell PM. The magnitude and challenge of false-positive newborn screening test results. Arch Pediatr Adolesc Med 2000; 154:714 – 8.
2. Tarini BA, Christakis DA, Welch HG. State newborn screening in the tandem mass spectrometry era: more tests, more false-positive results. Pediatrics 2006;118:448 –56.
3. Tymstra T. False positive results in screening tests: experiences of parents of children screened for congenital hypothyroidism. Fam Pract 1986;3:92– 6.
4. Baroni MA, Anderson YE, Mischler E. Cystic fibrosis newborn screening: impact of early screening results on parenting stress. Pediatr Nurs 1997;23:143–51.
5. Schulze A, Lindner M, Kohlmuller D, Olgemoller K, Mayatepek E, Hoffmann GF. Expanded newborn screening for inborn errors of metabolism by electrospray ionization-tandem mass spectrome-try: results, outcome, and implications. Pediatrics 2003;111: 1399 – 406.
1. Gurian EA, Kinnamon DD, Henry JJ, Waisbren SE. Expanded newborn screening for biochemical disorders: the effect of a false-positive result. Pediatrics 2006;117:1915–21.
7. Sorenson JR, Levy HL, Mangione TW, Sepe SJ. Parental re-sponse to repeat testing of infants with ‘false-positive’ results in a newborn screening program. Pediatrics 1984;73:183–7. 8. Moran J, Quirk K, Duff AJ, Brownlee KG. Newborn screening for
CF in a regional paediatric centre: the psychosocial effects of false-positive IRT results on parents. J Cyst Fibros 2007;6: 250 – 4.
9. Fyro K. Neonatal screening: life-stress scores in families given a false-positive result. Acta Paediatr Scand 1988;77:232– 8. 10. Waisbren SE, Albers S, Amato S, Ampola M, Brewster TG,
Demmer L, et al. Effect of expanded newborn screening for biochemical genetic disorders on child outcomes and parental stress. JAMA 2003;290:2564 –72.
11. Magera MJ, Helgeson JK, Matern D, Rinaldo P. Methylmalonic acid measured in plasma and urine by stable-isotope dilution and
electrospray tandem mass spectrometry. Clin Chem 2000;46: 1804 –10.
12. Kushnir MM, Komaromy-Hiller G, Shushan B, Urry FM, Roberts WL. Analysis of dicarboxylic acids by tandem mass spectrom-etry: high-throughput quantitative measurement of methylma-lonic acid in serum, plasma, and urine. Clin Chem 2001;47: 1993–2002.
13. Schmedes A, Brandslund I. Analysis of methylmalonic acid in plasma by liquid chromatography-tandem mass spectrometry. Clin Chem 2006;52:754 –7.
14. Chace DH. Mass spectrometry in the clinical laboratory. Chem Rev 2001;101:445–77.
15. Dooley KC. Tandem mass spectrometry in the clinical chemistry laboratory. Clin Biochem 2003;36:471– 81.