Preconcentration of Arsenic in Water Samples Using the

Composition-Induced Phase Separation Method and Determination by ETAAS

M. Güçoğlu1 , M. Efeçınar2 and N. Şatıroğlu3

1Department of Chemistry, Hacettepe University, Ankara, TURKEY, melekalp@hacettepe.edu.tr

2Department of Chemistry, Hacettepe University, Ankara, TURKEY, melise@hacettepe.edu.tr

3Department of Chemistry, Hacettepe University, Ankara, TURKEY, nurayo@hacettepe.edu.tr

Abstract. A new phase transition microextraction method was developed for determination of trace amount of arsenic ions in water samples in this work. The method is based on the critical point of miscibility of solvents. In this method the mixed solution of sample and organic solvent is initially homogeneous but is separated into two phases by adding a secondary solvent (modifier). In acidic medium As(V) was complexed with ammonium molybdate, this complex was quantitatively extracted to the mixture of organic solvent (acetonitrile/methyl isobutyl ketone) before ETAAS determination. Total inorganic arsenic (III, V) was extracted similarly after oxidation of As(III) to As(V) with nitric acid. Concentration of As(III) was calculated by difference in the concentration between total arsenic and As(V). Optimization of the experimental conditions and instrumental parameters was investigated in detail. A detection limit of 0.05 µgL-1 with enrichment factor of 85 was achieved for only 5 mL of sample. The analytical curve was linear in the concentration range of 0.25-4.00 µgL-1. Relative standard deviation (RSD) for 10 replicate determinations of 2.0 µgL-1 of As(V) was 4,1%. The method was successfully applied to preconcentration and determination of arsenic in real water samples.

Key words: Arsenic, preconcentration, composition-induced phase separation, ETAAS

Introduction

Arsenic is a ubiquitous element in the environment originating from natural sources as well as human activities, and has been identified as a public health problem because it has serious toxic effects even at low exposure levels (Mazumder, 1997). Different forms of arsenic have different toxicity and inorganic arsenic is more toxic than organic arsenic and As(III) is more toxic than As(V) (Majid et al., 2006). There are many reports in literature about the higher risks of skin, lung, liver and kidney cancer. The World Health Organization’s (WHO) maximum permissible contaminant level of arsenic in drinking water is 10 µgL-1. Therefore, the development of a sensitive and simple method for the determination of trace arsenic in water samples is significant.

The development of separation and preconcentration procedure prior to trace element determination has been explored in considerable depth in recent decades. Several procedures have been used for the separation and preconcentration of contaminants from environmental

samples; liquid–liquid extraction, co-precipitation, solid phase extraction.

Liquid-liquid extraction is a unit operation for separating one or more components from a liquid mixture. In recent years, research efforts in separation science and related research fields have focused on miniaturization of sample preparation process.

Liquid-liquid microextraction is a modified solvent extraction method and in this method organic phase-aqueous phase ratio (µL:mL) is greatly reduced comparing with the other extraction methods. It needs low volume of the sample and has sufficient preconcentration factor.

Therefore, this method is very useful for the separation and preconcentration of metal ions after the formation of sparingly water-soluble complexes (Pena-Pereira et al., 2009).

A novel liquid extraction process, named phase transition extraction, which is based on using solvents that, together with the liquid to be extracted, form a solution with a critical point of miscibility was developed E3S Web of Conferences

DOI: 10.1051/

COwned by the authors, published by EDP Sciences, 2013 ,

/20130137002 1

conf e3s

37002 (2013)

This is an Open Access article distributed under the terms of the Creative Commons Attribution License 2 0 , which . permits unrestricted use, distributi and reproduction in any medium, provided the original work is properly cited.

on, Article available athttp://www.e3s-conferences.orgorhttp://dx.doi.org/10.1051/e3sconf/20130137002

(2)

E3S Web of Conferences

by Ullmann et al. (Ullmann et al., 1995). In this process, the phase transition of partially miscible mixtures is achieved by adding a modifier which is insoluble in native (water) solvent. The method carried out isothermally by changing the mixture composition instead of its temperature is called as composition-induced phase separation (CIPS) (Gupta et al., 1996) or solvent-induced phase transition extraction (SIPTE) (Liu et al., 2010). This extraction method has been applied for the separation of organic compounds from matrix in literature. In this work, first time CIPS methodology was successfully applied for the microextraction of arsenic in water samples as a prior step to their determination by ETAAS.

Materials and Methods

All chemicals used in this work were of analytical grade and all solutions were prepared with deionized water (18.1 MΩ cm) obtained from a Barnstead, Nanopure Diamond purification system. As(V) and As(III) stock standard solutions (1000 mgL-1) were prepared from Na2HAsO4.7H2O (Sigma-Aldrich) and As2O3 (Fisher Scientific). Ammonium heptamolybdate tetrahydrate ((NH4)6Mo7O24.4H2O) (Merck) was used as a complexing agent. Ammonium molybdate was prepared in H2SO4. H2SO4, HCl and HNO3 were used for adjusting pH of samples. In addition concentrated nitric acid was used as an oxidant agent for converting As(III) to As(V). The organic extractant was mixture solution of acetonitrile and methyl isobutyl ketone. Pd(NO3)2 (Sigma-Aldrich) was used as a chemical modifier in ETAAS.

Measurements were carried out on Perkin-Elmer AAnalyst 800 atomic absorption spectrophotometer equipped with THGA prolytic coated graphite furnace atomizer and arsenic electrodeless discharge lamp.

The test solution which is containing arsenic and complexing agent is first mixed with acetonitrile; then a drop of MIBK is added as the modifier to induce phase separation of the mixture. The separated organic phase with the analyte extracted can be directly injected to the ETAAS.

It is known that only As(V) forms complex with ammonium molybdate, but not As (III). In this method, after As(V) was complexed with ammonium molybdate, organic solvents were added into 5 mL of aqueous sample (final pH of this arsenic sample was 0.8-1.0) and added volume of organic solvents was in the µL level. The mixture was stirred using vortex for about 5 minutes.

Separation of the aqueous and organic phases was accomplished approximately in 30 seconds. After extraction, the 60 µL of organic phase enriched with arsenic was taken by a micropipette and determined by ETAAS. After simple oxidation with nitric acid, converting As(III) to As(V), total arsenic was extracted and determined similarly (Sounderajan et al., 2007). The concentration of As(III) was calculated by difference in the concentration between total arsenic and As(V).

Results and Discussion

In this method acetonitrile and MIBK were used as organic solvents. Acetonitrile is completely miscible with water but MIBK and water form a broad miscibility gap.

Therefore MIBK is used as a secondary solvent or modifier.

Several factors that influence the microextraction efficiency and ETAAS signal, such as pH, concentration of chelating agent, extraction time, organic solvents and extraction solvents, organic phase ratio, pyrolysis and atomization temperature were investigated.

Important parameters such as the linear range, calibration graph, precision, detection limit, recoveries, and enrichment factor were determined to evaluate the method performance. The calibration graph was linear in the range of 0.25-4.0 µgL-1 with correlation coefficient of 0.9983 under the optimum conditions of the recommended procedure. The detection limit was 0.05 µgL-1 and the relative standard deviation (n=10) at 2 µgL−1 As(V) concentration was 4.1%. The enhancement factor defined as the slope ratio of two calibration curves with microextraction and without microextraction was 85.

In order to verify the accuracy of the proposed procedure, the method was applied to the determination of 1.36 µgL−1 As in standard reference material, SLEW-3 Estuarine water. The presented procedure was applied to the speciation of arsenic in a tap water, sea water, drinking water and ground water. Various amounts of As(V) and As(III) were also spiked to these samples. A good agreement was obtained between the added and measured analyte amounts. The conventional LLE and CIPS methods were also applied to the standard solutions and obtained results were compared.

Conclusion

We have presented a fast, reliable, sensitive and accurate method, which benefits from a very low detection limit, for the determination of As(V) and total arsenic in aqueous solutions. Using the proposed method, arsenic can be determined with a very high sensitivity and relatively good reproducibility in real water samples such as tap water, sea water, ground water and drinking water.

References

Gupta R, Mauri R, Shinnar R. Liquid-Liquid Extraction Using the Composition-Induced Phase Separation Process, Ind. Eng. Chem. Res., 1996, 35, 236.

Liu G, Zhou N, Zhang M, Li S , Tian Q, Chen J, Chen B, Wu Y, Yao S. Hydrophobic solvent induced phase transition extraction to extract drugs from plasma for high performance liquid chromatography–mass spectrometric analysis, Journal of Chromatography A, 2010, 1217, 243.

Majid E, Harapoviç S, Liu Y, Male K.B, Luong J.H.T.

Electrochemical determination of arsenite using a

37002-p.2

(3)

ICHMET 2012

gold nanoparticle modified glassy carbon electrode and flow analysis, Anal. Chem., 2006, 78 , 762.

Mazumder N.G. Chronic arsenic toxicity in West Bengal, Curr. Sci., 1997, 72, 114.

Pena-Pereira F, Lavilla I, Bendicho C. Miniaturized preconcentration methods based on liquid–liquid extraction and their application in inorganic ultratrace analysis and speciation: A review, Spectrochimica Acta Part B, 2009, 64, 15.

Shemirani F, Baghdadi M, Ramezani M.

Preconcentration and determination of ultra trace amounts of arsenic (III) and arsenic (V) in tap water

and total arsenic in biological samples by cloud point extraction and electrothermal atomic absorption spectrometry, Talanta, 2005, 65, 887.

Sounderajan S, Udas A.C., Venkataramani B, Characterization of arsenic (V) and arsenic (III) in water samples using ammonium molybdate and estimation by graphite furnace atomic absorption spectroscopy, Journal of Hazardous Materials, 2007, 149, 238.

Ullmann A, Ludmer Z, Shinnar R. Phase transition extraction using solvent mixtures with a critical point of miscibility, AIChE J., 1995, 41, 488.

37002-p.3

Riferimenti

Documenti correlati

I rilievi sul numero e stadio fenologico dei polloni hanno evidenziato un’importante differenza tra le tesi riguardanti il trattamento precoce e il doppio

In this thesis, I carried out a detailed study combining (1) paleomagnetic analysis of the central Iran in the late Jurassic-early Cretaceous to reconstruct the history of

L’analisi, pur rilevando una grande vivacità del settore, in parte soste- nuta dalla Regione mediante un vasto ed articolato programma di attività di conservazione del

a 2-period cycle appears close to the fixed point and it is stable along the diagonal. By taking into account the

The model takes as input a set of viral load measurements, gender, age, risk group, resistance mutations, number of past treatments recorded, the lower quantile of the distance

Example 16 is interesting not only because it is one of the few cases in which the pronoun has been rendered explicit in the Italian subtitles, but also because of the fact that it

Based on the background-suppressed spectrum and the expected statistical improve- ment, it seems to be possible to reduce the uncertainty of the low-energy fraction of the π + → e +

Scintillator paddles of approximately 50 × 8.4 cm were used to detect betas and silicon surface barrier detectors biased to −28 kV were used to detect protons.. Details of

Noté 0.0/5: Achetez Dinosauri. Chi cerca trova.. Presso una sala della Promotrice delle Belle Arti sarà organizzata un'area interattiva e didattica nella quale i più piccoli

In this study, determination of fluorine in water samples was established via the rotational fine structured molecular absorption of calcium mono-fluoride (CaF)

To obtain optimal conditions of removing surface active substances by adsorption on Amberlite XAD-16 resin the effects of the pH of the acetate buffer, the amount of

This work provides a simple and fast procedure for determination of Ti(IV), which allows for the analysis of environmental water samples with a complicated matrix containing

In this study we present the results of arsenic concentration in water samples from public water supply system of city Zrenjanin and three Northern municipalities Elemiš,

A novel solidified floating organic drop microextraction method for preconcentration and determination of copper ions by flow injection flame atomic

'Candidatus Moeniiplasma glomeromycotorum', an endobacterium of arbuscular mycorrhizal fungi / Naito, Mizue; Desirò, Alessandro; Gonzalez, Jonathan B; Tao, Gang; Morton, Joseph

Figure 3 – Set-ups for structural health monitoring and non-destructive testing (dimensions in [mm]): a) static test; b) dynamic test; c) example of speckled region of interest for

For the water quality evaluation problem, this paper will develop a new evaluation method based on efficacy coefficient method combining with the coefficient of variation (CV)

Progressive freeze concentration is a method of separating solvent and solution into a single ice layer on the wall where coolant is applied, whereas suspension freeze

If the volume of EDTA required in the first titration is low due to the fact that the water is not very hard, increase the volume of the water sample so that in succeeding

Abstract: Dispersive liquid-liquid microextraction (DLLME) combined with graphite furnace atomic absorption spectrometry (GFAAS) was proposed for the determination of trace amounts of

Although full recognition of the drug resistance phenomenon demands the identification of bacterial species for the comparison with susceptibility testing guidelines, this

The computation results successfully explained the time series of river discharge observed at stations located in the river reach, as well as the longitudinal profile of the

The aim of the work detailed here was to develop and validate a method by which to make determinations for dibutyl phthalate in water samples, using solid phase extraction and